Impact of Skin Pigmentation on Pulse Oximetry Blood Oxygenation and Wearable Pulse Rate Accuracy: Systematic Review and Meta-Analysis

被引:1
|
作者
Singh, Sanidhya [1 ]
Bennett, Miles Romney [2 ]
Chen, Chen [2 ]
Shin, Sooyoon [2 ]
Ghanbari, Hamid [1 ]
Nelson, Benjamin W. [2 ,3 ]
机构
[1] Univ Michigan, Med Sch, 1301 Catherine St, Ann Arbor, MI 48109 USA
[2] Verily Life Sci LLC, South San Francisco, CA USA
[3] Harvard Med Sch, Beth Israel Deaconess Med Ctr, Dept Psychiat, Div Digital Psychiat, Boston, MA USA
关键词
photoplethysmography; pulse oximetry; arterial blood gas; skin tone; skin pigmentation; bias; digital technology; RELIABILITY; SATURATION; EXERCISE;
D O I
10.2196/62769
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
Background:Photoplethysmography (PPG) is a technology routinely used in clinical practice to assess blood oxygenation (SpO(2)) and pulse rate (PR). Skin pigmentation may influence accuracy, leading to health outcomes disparities. Objective:This systematic review and meta-analysis primarily aimed to evaluate the accuracy of PPG-derived SpO(2) and PR by skin pigmentation. Secondarily, we aimed to evaluate statistical biases and the clinical relevance of PPG-derived SpO(2) and PR according to skin pigmentation. Methods:We identified 23 pulse oximetry studies (n=59,684; 197,353 paired SpO(2)-arterial blood observations) and 4 wearable PR studies (n=176; 140,771 paired PPG-electrocardiography observations). We evaluated accuracy according to skin pigmentation group by comparing SpO(2) accuracy root-mean-square values to the regulatory threshold of 3% and PR 95% limits of agreement values to +5 or -5 beats per minute (bpm), according to the standards of the American National Standards Institute, Association for the Advancement of Medical Instrumentation, and the International Electrotechnical Commission. We evaluated biases and clinical relevance using mean bias and 95% CI. Results:For SpO(2), accuracy root-mean-square values were 3.96%, 4.71%, and 4.15%, and pooled mean biases were 0.70% (95% CI 0.17%-1.22%), 0.27% (95% CI -0.64% to 1.19%), and 1.27% (95% CI 0.58%-1.95%) for light, medium, and dark pigmentation, respectively. For PR, 95% limits of agreement values were from -16.02 to 13.54, from -18.62 to 16.84, and from -33.69 to 32.54, and pooled mean biases were -1.24 (95% CI -5.31 to 2.83) bpm, -0.89 (95% CI -3.70 to 1.93) bpm, and -0.57 (95% CI -9.44 to 8.29) bpm for light, medium, and dark pigmentation, respectively. Conclusions:SpO(2) and PR measurements may be inaccurate across all skin pigmentation groups, breaching U.S. Food and Drug Administration guidance and industry standard thresholds. Pulse oximeters significantly overestimate SpO(2) for both light and dark skin pigmentation, but this overestimation may not be clinically relevant. PRs obtained from wearables exhibit no statistically or clinically significant bias based on skin pigmentation.
引用
收藏
页数:21
相关论文
共 50 条
  • [1] The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis
    Shi, Chunhu
    Goodall, Mark
    Dumville, Jo
    Hill, James
    Norman, Gill
    Hamer, Oliver
    Clegg, Andrew
    Watkins, Caroline Leigh
    Georgiou, George
    Hodkinson, Alexander
    Lightbody, Catherine Elizabeth
    Dark, Paul
    Cullum, Nicky
    BMC MEDICINE, 2022, 20 (01)
  • [2] The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation: a systematic review and meta-analysis
    Chunhu Shi
    Mark Goodall
    Jo Dumville
    James Hill
    Gill Norman
    Oliver Hamer
    Andrew Clegg
    Caroline Leigh Watkins
    George Georgiou
    Alexander Hodkinson
    Catherine Elizabeth Lightbody
    Paul Dark
    Nicky Cullum
    BMC Medicine, 20
  • [3] Skin Pigmentation Influence on Pulse Oximetry Accuracy: A Systematic Review and Bibliometric Analysis
    Cabanas, Ana M.
    Fuentes-Guajardo, Macarena
    Latorre, Katina
    Leon, Dayneri
    Martin-Escudero, Pilar
    SENSORS, 2022, 22 (09)
  • [4] The Effect of Skin Pigmentation on the Accuracy of Pulse Oximetry in Infants with Hypoxemia
    Foglia, Elizabeth E.
    Whyte, Robin K.
    Chaudhary, Aasma
    Mott, Antonio
    Chen, Jodi
    Propert, Kathleen J.
    Schmidt, Barbara
    JOURNAL OF PEDIATRICS, 2017, 182 : 375 - 377
  • [5] Effect of skin pigmentation on pulse oximetry accuracy in the emergency department
    Adler, JN
    Hughes, LA
    Vivilecchia, R
    Camargo, CA
    ACADEMIC EMERGENCY MEDICINE, 1998, 5 (10) : 965 - 970
  • [6] Accuracy of pulse CO-oximetry to evaluate blood carboxyhemoglobin level: a systematic review and meta-analysis of diagnostic test accuracy studies
    Papin, Mathilde
    Latour, Chloe
    Leclere, Brice
    Javaudin, Francois
    EUROPEAN JOURNAL OF EMERGENCY MEDICINE, 2023, 30 (04) : 233 - 243
  • [7] Skin Pigmentation Effects on Pulse Oximetry Accuracy Need a Prospective Study
    Blike, George T.
    McGrath, Susan P.
    McGovern, Krystal
    JOURNAL OF PATIENT SAFETY, 2023, 19 (04) : 221 - 222
  • [8] A review of the effect of skin pigmentation on pulse oximeter accuracy
    Al-Halawani, Raghda
    Charlton, Peter H.
    Qassem, Meha
    Kyriacou, Panayiotis A.
    PHYSIOLOGICAL MEASUREMENT, 2023, 44 (05)
  • [9] Pulse Oximetry and Perfusion Index Screening for Congenital Heart Defects: A Systematic Review and Meta-analysis
    Jiang, S. L.
    Zhan, Y. J.
    Yan, P.
    Yue, Y.
    Tang, J.
    AMERICAN JOURNAL OF PERINATOLOGY, 2023, 40 (15) : 1611 - 1617
  • [10] The diagnostic accuracy of carbon monoxide pulse oximetry in adults with suspected acute carbon monoxide poisoning: a systematic review and meta-analysis
    Ramponi, Giacomo
    Gianni, Francesca
    Karlafti, Eleni
    Piazza, Isabelle
    Albertoni, Francesco
    Colombo, Giorgio
    Casazza, Giovanni
    Garegnani, Anna
    Casella, Rosa
    Costantino, Giorgio
    FRONTIERS IN MEDICINE, 2023, 10