Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis

被引:0
|
作者
Lee, Hyun Hee [1 ]
Hong, Hyoung-Taek [2 ]
Kim, Jong-Keun [3 ]
Koh, Yong-Gon [4 ]
Park, Kwan Kyu [5 ]
Kang, Kyoung-Tak [2 ,6 ]
机构
[1] Int St Marys Hosp, Dept Orthopaed Surg, Incheon 22711, South Korea
[2] Skyve R&D LAB, Seoul 07217, South Korea
[3] Heung K Hosp, Dept Orthopaed Surg, Shihung 14999, South Korea
[4] Yonsei Sarang Hosp, Joint Reconstruct Ctr, Dept Orthopaed Surg, Seoul 06702, South Korea
[5] Yonsei Univ, Coll Med, Severance Hosp, Dept Orthoped Surg, Seoul 03722, South Korea
[6] Yonsei Univ, Dept Mech Engn, Seoul 03722, South Korea
来源
BIOENGINEERING-BASEL | 2025年 / 12卷 / 02期
关键词
tibial stem geometry; design of experiments; finite element analysis;
D O I
10.3390/bioengineering12020172
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The stability of the tibial component in Total Knee Arthroplasty (TKA) is critical to preventing aseptic loosening, a major cause of implant failure. However, existing tibial stem designs often lead to stress shielding and bone resorption, highlighting the need for further optimization. This study addresses these challenges by employing the Design of Experiments (DOE) methodology, specifically utilizing a full factorial design approach combined with finite element analysis (FEA), to optimize the geometry of the tibial stem. The material properties of the cortical and cancellous bone, as well as the tibial tray, were assigned based on values from the literature, representing their elastic moduli and Poisson's ratios. For boundary conditions, the distal end of the tibia was fully constrained to simulate realistic load transfer, while compressive loads representative of walking and daily activities were applied to the tibial base. Key design parameters, including stem diameter, length, mediolateral ratio (M/L ratio), and wing angle, were systematically analyzed. The results identified stem diameter and length as the most influential factors in improving biomechanical performance, while the wing angle showed minimal impact. The optimized design, featuring a stem diameter of 12 mm, length of 40 mm, M/L ratio of 0.61, and a wing angle of 60 degrees, demonstrated significant reductions in stress shielding and aseptic loosening compared to conventional models. These findings provide valuable insights into enhancing the long-term success of TKA implants by balancing implant stability and minimizing bone resection.
引用
收藏
页数:16
相关论文
共 50 条
  • [31] Is it necessary to cement the tibial stem in cemented total knee arthroplasty?
    Bert, JM
    McShane, M
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (356) : 73 - 78
  • [32] Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study
    Bilal Farouk El-Zayat
    Thomas J. Heyse
    Nelson Fanciullacci
    Luc Labey
    Susanne Fuchs-Winkelmann
    Bernardo Innocenti
    Archives of Orthopaedic and Trauma Surgery, 2016, 136 : 1741 - 1752
  • [33] Fixation techniques and stem dimensions in hinged total knee arthroplasty: a finite element study
    El-Zayat, Bilal Farouk
    Heyse, Thomas J.
    Fanciullacci, Nelson
    Labey, Luc
    Fuchs-Winkelmann, Susanne
    Innocenti, Bernardo
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2016, 136 (12) : 1741 - 1752
  • [34] Prediction of Wear on Tibial Inserts Made of UHMWPE, PEEK, and CFR-PEEK in Total Knee Arthroplasty Using Finite-Element Analysis
    Koh, Yong-Gon
    Lee, Jin-Ah
    Kang, Kyoung-Tak
    LUBRICANTS, 2019, 7 (04)
  • [35] FINITE ELEMENT MODEL OF SCREWS FOR CEMENT AUGMENTATION IN TOTAL KNEE ARTHROPLASTY WITH TIBIAL UNCONTAINED BONE DEFECT
    Yuenyongviwat, V.
    Kwanyuang, A.
    AGING CLINICAL AND EXPERIMENTAL RESEARCH, 2024, 36 : S598 - S598
  • [36] PREDICTION OF FATIGUE FAILURE OF A TOTAL KNEE REPLACEMENT TIBIAL PLATEAU USING FINITE-ELEMENT ANALYSIS
    PAGANELLI, JV
    SKINNER, HB
    MOTE, CD
    ORTHOPEDICS, 1988, 11 (08) : 1161 - 1168
  • [37] Current analysis of tibial coverage in total knee arthroplasty
    Westrich, GH
    Agulnick, MA
    Laskin, RS
    Haas, SB
    Sculco, TP
    KNEE, 1997, 4 (02): : 87 - 91
  • [38] Optimisation of the posterior stabilised tibial post for greater femoral rollback after total knee arthroplasty-a finite element analysis
    Chandran, Nagarajan
    Amirouche, Farid
    Gonzalez, Mark H.
    Hilton, Kevin M.
    Barmada, Riad
    Goldstein, Wayne
    INTERNATIONAL ORTHOPAEDICS, 2009, 33 (03) : 687 - 693
  • [39] The effect of tibial stem design on component micromotion in knee arthroplasty
    Stern, SH
    Wills, RD
    Gilbert, JL
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1997, (345) : 44 - 52
  • [40] Increased Rates of Tibial Aseptic Loosening in Primary Cemented Total Knee Arthroplasty With a Short Native Tibial Stem Design
    Garceau, Simon P.
    Pivec, Robert
    Teo, Greg
    Chisari, Emanuele
    Enns, Paul A.
    Weinblatt, Aaron, I
    Aggarwal, Vinay K.
    Austin, Matthew S.
    Long, William J.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2022, 30 (07) : E640 - E648