Optimization of Tibial Stem Geometry in Total Knee Arthroplasty Using Design of Experiments: A Finite Element Analysis

被引:0
|
作者
Lee, Hyun Hee [1 ]
Hong, Hyoung-Taek [2 ]
Kim, Jong-Keun [3 ]
Koh, Yong-Gon [4 ]
Park, Kwan Kyu [5 ]
Kang, Kyoung-Tak [2 ,6 ]
机构
[1] Int St Marys Hosp, Dept Orthopaed Surg, Incheon 22711, South Korea
[2] Skyve R&D LAB, Seoul 07217, South Korea
[3] Heung K Hosp, Dept Orthopaed Surg, Shihung 14999, South Korea
[4] Yonsei Sarang Hosp, Joint Reconstruct Ctr, Dept Orthopaed Surg, Seoul 06702, South Korea
[5] Yonsei Univ, Coll Med, Severance Hosp, Dept Orthoped Surg, Seoul 03722, South Korea
[6] Yonsei Univ, Dept Mech Engn, Seoul 03722, South Korea
来源
BIOENGINEERING-BASEL | 2025年 / 12卷 / 02期
关键词
tibial stem geometry; design of experiments; finite element analysis;
D O I
10.3390/bioengineering12020172
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
The stability of the tibial component in Total Knee Arthroplasty (TKA) is critical to preventing aseptic loosening, a major cause of implant failure. However, existing tibial stem designs often lead to stress shielding and bone resorption, highlighting the need for further optimization. This study addresses these challenges by employing the Design of Experiments (DOE) methodology, specifically utilizing a full factorial design approach combined with finite element analysis (FEA), to optimize the geometry of the tibial stem. The material properties of the cortical and cancellous bone, as well as the tibial tray, were assigned based on values from the literature, representing their elastic moduli and Poisson's ratios. For boundary conditions, the distal end of the tibia was fully constrained to simulate realistic load transfer, while compressive loads representative of walking and daily activities were applied to the tibial base. Key design parameters, including stem diameter, length, mediolateral ratio (M/L ratio), and wing angle, were systematically analyzed. The results identified stem diameter and length as the most influential factors in improving biomechanical performance, while the wing angle showed minimal impact. The optimized design, featuring a stem diameter of 12 mm, length of 40 mm, M/L ratio of 0.61, and a wing angle of 60 degrees, demonstrated significant reductions in stress shielding and aseptic loosening compared to conventional models. These findings provide valuable insights into enhancing the long-term success of TKA implants by balancing implant stability and minimizing bone resection.
引用
收藏
页数:16
相关论文
共 50 条
  • [1] Finite element analysis of the tibial bone graft in cementless total knee arthroplasty
    Koji Totoribe
    Etsuo Chosa
    Go Yamako
    Hiroaki Hamada
    Koki Ouchi
    Shutaro Yamashita
    Gang Deng
    Journal of Orthopaedic Surgery and Research, 13
  • [2] Finite element analysis of the tibial bone graft in cementless total knee arthroplasty
    Totoribe, Koji
    Chosa, Etsuo
    Yamako, Go
    Hamada, Hiroaki
    Ouchi, Koki
    Yamashita, Shutaro
    Deng, Gang
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2018, 13
  • [3] Analysis of the shape of the tibial tray in total knee arthroplasty using a three dimension finite element model
    Miyoshi, S
    Takahashi, T
    Ohtani, M
    Yamamoto, H
    Kameyama, K
    CLINICAL BIOMECHANICS, 2002, 17 (07) : 521 - 525
  • [4] Finite element analysis of the tibial component alignment in a transverse plane in total knee arthroplasty
    Popescu, Roman
    Cristea, Stefan
    Oleksik, Valentin
    Pascu, Adrian Marius
    Haritinian, Emil George
    JOURNAL OF APPLIED BIOMEDICINE, 2021, 19 (04) : 234 - 239
  • [5] Finite element analysis for the biomechanical effect of tibial insert materials in total knee arthroplasty
    Kang, Kyoung-Tak
    Son, Juhyun
    Kwon, Sae Kwang
    Kwon, Oh-Ryong
    Park, Joon-Hee
    Koh, Yong-Gon
    COMPOSITE STRUCTURES, 2018, 201 : 141 - 150
  • [6] Finite -Element analysis of a lateral femoro-tibial impact on the total knee arthroplasty
    Arab, Ali Zine El-Abidine
    Merdji, Ali
    Benaissa, Ali
    Roy, Sandipan
    Bouiadjra, Bel-Abbes Bachir
    Layadi, Khaled
    Ouddane, Abdelhakim
    Mukdadi, Osama M.
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 192 (192)
  • [7] The Effect of Malrotation of Tibial Component of Total Knee Arthroplasty on Tibial Insert during High Flexion Using a Finite Element Analysis
    Osano, Kei
    Nagamine, Ryuji
    Todo, Mitsugu
    Kawasaki, Makoto
    SCIENTIFIC WORLD JOURNAL, 2014,
  • [8] AN ANALYSIS OF TIBIAL COMPONENT DESIGN IN TOTAL KNEE ARTHROPLASTY
    MURASE, K
    CROWNINSHIELD, RD
    PEDERSEN, DR
    CHANG, TS
    JOURNAL OF BIOMECHANICS, 1983, 16 (01) : 13 - &
  • [9] Finite element analysis in the optimization of posterior-stabilized total knee arthroplasty
    Dagneaux, Louis
    Canovas, Francois
    Jourdan, Franck
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2024, 110 (01)
  • [10] Finite element analysis of the effect of tibial osteotomy on the stress of polyethylene liner in total knee arthroplasty
    Sun, Lihui
    Han, Yu
    Jing, Zheng
    Li, Dongbo
    Liu, Jianguo
    Li, Dongsong
    JOURNAL OF ORTHOPAEDIC SURGERY, 2024, 32 (02)