Segmentation methods and dosimetric evaluation of 3D-printed immobilization devices in head and neck radiotherapy

被引:0
|
作者
Yin, Yunpeng [1 ,2 ]
Zhang, Weisha [2 ]
Zou, Lian [2 ]
Liu, Xiangxiang [2 ]
Yu, Luxin [2 ]
Wang, Ming [1 ]
机构
[1] Chengdu Univ Technol, Coll Nucl Technol & Automat Engn, Chengdu 610059, Peoples R China
[2] Univ Elect Sci & Technol China, Sichuan Prov Peoples Hosp, Canc Ctr, Chengdu 610072, Peoples R China
关键词
Radiation therapy; 3D printing; Image segmentation; MODULATED RADIATION-THERAPY; IMRT; CANCER; IMPACT; ARC;
D O I
10.1186/s12885-025-13669-0
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background Treatment planning systems (TPS) often exclude immobilization devices from optimization and calculation, potentially leading to inaccurate dose estimates. This study employed deep learning methods to automatically segment 3D-printed head and neck immobilization devices and evaluate their dosimetric impact in head and neck VMAT. Methods Computed tomography (CT) positioning images from 49 patients were used to train the Mask2Former model to segment 3D-printed headrests and MFIFs. Based on the results, four body structure sets were generated for each patient to evaluate the impact on dose distribution in volumetric modulated arc therapy (VMAT) plans: S (without immobilization devices), S_(MF) (with MFIFs), S_3D (with 3D-printed headrests), and S_(3D+MF) (with both). VMAT plans (P, P_(MF), P_(3D), and P_(3D+MF)) were created for each structure set. Dose-volume histogram (DVH) data and dose distribution of the four plans were compared to assess the impact of the 3D-printed headrests and MFIFs on target and normal tissue doses. Gafchromic EBT3 film measurements were used for patient-specific verification to validate dose calculation accuracy. Results The Mask2Former model achieved a mean average precision (mAP) of 0.898 and 0.895, with a Dice index of 0.956 and 0.939 for the 3D-printed headrest on the validation and test sets, respectively. For the MFIF, the Dice index was 0.980 and 0.981 on the validation and test sets, respectively. Compared to P, P_MF reduced the V-100% for PGTVnx, PGTVnd, PGTVrpn, PTV1, and PTV2 by 5.99%, 6.51%, 5.93%, 2.24%, and 1.86%, respectively(P <= 0.004). P_3D reduced the same targets by 1.78%, 2.56%, 1.75%, 1.16%, and 1.48%(P < 0.001), with a 31.3% increase in skin dose (P < 0.001). P_(3D+MF) reduced the V-100% by 9.15%, 10.18%, 9.16%, 3.36%, and 3.28% (P < 0.001), respectively, while increasing the skin dose by 31.6% (P < 0.001). EBT3 film measurements showed that the P_(3D+MF) dose distribution was more aligned with actual measurements, achieving a mean gamma pass rate of 92.14% under the 3%/3 mm criteria. Conclusions This study highlights the potential of Mask2Former in 3D-printed headrest and MFIF segmentation automation, providing a novel approach to enhance personalized radiation therapy plan accuracy. The attenuation effects of 3D-printed headrests and MFIFs reduce V-100% and D-mean for PTVs in head and neck cancer patients, while the buildup effects of 3D-printed headrests increases the skin dose (31.3%). Challenges such as segmentation inaccuracies for small targets and artifacts from metal fasteners in MFIFs highlight the need for model optimization and validation on larger, more diverse datasets.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] A Dosimetric Evaluation of 3D-Printed Thermoplastic Polyurethane (TPU) for Use as a Custom Breast Bolus
    Bittinger, K.
    Graeper, G.
    MEDICAL PHYSICS, 2022, 49 (06) : E942 - E943
  • [32] Evaluation of Fused Deposition Modeling Materials for 3D-Printed Container of Dosimetric Polymer Gel
    Lee, Minsik
    Noh, Seonyeong
    Shin, Jun-Bong
    Kwak, Jungwon
    Jeong, Chiyoung
    GELS, 2024, 10 (02)
  • [33] Development and clinical implementation of a digital workflow utilizing 3D-printed oral stents for patients with head and neck cancer receiving radiotherapy
    Tino, Rance
    Roach, Millicent A.
    Fuentes, Gabriela D.
    Agrawal, Anshuman
    Zaid, Mohamed
    Cooper, Dylan J.
    Bajaj, Nimit
    Lin, Ruitao
    Xiao, Lianchun
    Mayo, Lauren L.
    Wiederhold, Lee R.
    Shah, Shalin J.
    Tate, Molly K.
    Chronowski, Gregory M.
    Reddy, Jay P.
    Mezera, Megan
    Mann, Justin M.
    Augspurger, Mark
    Ottum, Adegbenga O.
    Chambers, Mark S.
    Koay, Eugene J.
    ORAL ONCOLOGY, 2024, 157
  • [34] Optimization methods for the design of advanced 3d-printed metal seismic dissipation devices
    Andreacola, Francesca Romana
    Brando, Giuseppe
    ce/papers, 2023, 6 (3-4) : 888 - 892
  • [35] Dosimetric verification of IMRT head and neck plans using a 3D printed heterogeneous phantom
    Olasolo Alonso, J.
    Collado Chamorro, P.
    Sanz Freire, C. J.
    Diaz Pascual, V.
    Vazquez Galinanes, A.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S993 - S994
  • [36] 3D-Printed Microfluidic Devices for Materials Science
    Alizadehgiashi, Moien
    Gevorkian, Albert
    Tebbe, Moritz
    Seo, Minseok
    Prince, Elisabeth
    Kumacheva, Eugenia
    ADVANCED MATERIALS TECHNOLOGIES, 2018, 3 (07):
  • [37] 3D-Printed alternating current electroluminescent devices
    Brubaker, Cole D.
    Newcome, Kailey N.
    Jennings, G. Kane
    Adams, Douglas E.
    JOURNAL OF MATERIALS CHEMISTRY C, 2019, 7 (19) : 5573 - 5578
  • [38] Biocompatibility of 3D-Printed Methacrylate for Hearing Devices
    Alifui-Segbaya, Frank
    George, Roy
    INVENTIONS, 2018, 3 (03)
  • [39] 3D-Printed Wearable Electrochemical Energy Devices
    Zhang, Shuai
    Liu, Yuqing
    Hao, Junnan
    Wallace, Gordon G.
    Beirne, Stephen
    Chen, Jun
    ADVANCED FUNCTIONAL MATERIALS, 2022, 32 (03)
  • [40] DOSIMETRIC EVALUATION OF AUTOMATIC SEGMENTATION FOR ADAPTIVE IMRT FOR HEAD-AND-NECK CANCER
    Tsuji, Stuart Y.
    Hwang, Andrew
    Weinberg, Vivian
    Yom, Sue S.
    Quivey, Jeanne M.
    Xia, Ping
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 77 (03): : 707 - 714