Design of a medical ultrasound imaging system based on a ring array

被引:0
|
作者
Mo, Xiaohai [1 ,2 ]
Su, Chang [1 ,2 ]
Kong, Chao [1 ]
Jiang, Lekang [1 ,2 ]
Lin, Weijun [1 ,2 ]
机构
[1] Institute of Acoustics, Chinese Academy of Sciences, Beijing,100190, China
[2] University of Chinese Academy of Sciences, Beijing,100049, China
关键词
Ultrasonic imaging;
D O I
10.19650/j.cnki.cjsi.J2412587
中图分类号
学科分类号
摘要
In order to address the limitations of traditional handheld ultrasound scanning systems, such as narrow field of view, and non-standardization, this article presents the design and implementation of comprehensive ultrasound imaging system based on a ring array. This system encompasses a 512-element ultrasonic annular array, a high-performance 256 / 1 024-channel acquisition circuit, an efficient PCIe-based data transmission module, and comprehensive software for acquisition control and rapid imaging. The multi-channel acquisition circuit comprising eight 32 / 128 ultrasonic transceiver control modules enables full-matrix data acquisition and high-speed transmission of 512 annular array elements within 12 seconds, achieving a remarkable data transmission efficiency of 1. 2 Gb / s. The system boasts a broadband frequency range of 0. 16 ~ 15 MHz and adjustable A / D sampling rates ranging from 12. 5 ~ 100 MHz, which meets the diverse requirements of clinical and experimental applications. The acquisition control software facilitates multi-mode ultrasound data acquisition, enabling the rapid imaging monitoring and secure storage of full-matrix data for subsequent precise image reconstruction. Phantom and in-vivo experiments show that the system can obtain an imaging range exceeding 10 cm× 10 cm, resolve 1 mm abnormalities accurately, and provide detailed visualization of interfaces within biological tissues such as fat, muscle, and bone. These results demonstrate the system′ s capability of performing circumferential scanning, producing comprehensive and standardized medical ultrasound images with a wide field of view. © 2024 Science Press. All rights reserved.
引用
收藏
页码:198 / 207
相关论文
共 50 条
  • [21] Coherence-based phase aberration correction and beamforming for ring-array ultrasound imaging
    Lan, Zhengfeng
    Rong, Chao
    Han, Changshan
    Qu, Xiaolei
    Li, Jingsong
    Lin, Hongxiang
    MECHANICAL SYSTEMS AND SIGNAL PROCESSING, 2024, 220
  • [22] Volatility Clustering in Medical Ultrasound Imaging and System Identification Based Deconvolution
    Kostoglou, Kyriaki
    Wallinger, Christian
    Huemer, Mario
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1459 - 1462
  • [23] Rapid and Stable Photoacoustic Imaging System Based on Fiber Ultrasound Detector Array
    Wang, Anqi
    Yang, Liuyang
    Xu, Dongchen
    Chen, Geng
    Liu, Yueqi
    Zhang, Zhi
    Zhang, Shaoling
    Li, Hao
    Yan, Zhijun
    Chen, Gina Jinna
    Shum, Perry Ping
    Sun, Qizhen
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2025, 43 (05) : 2340 - 2348
  • [24] NONLINEAR CONTRAST IMAGING WITH AN ARRAY-BASED MICRO-ULTRASOUND SYSTEM
    Needles, A.
    Arditi, M.
    Rognin, N. G.
    Mehi, J.
    Coulthard, T.
    Bilan-Tracey, C.
    Gaud, E.
    Frinking, P.
    Hirson, D.
    Foster, F. S.
    ULTRASOUND IN MEDICINE AND BIOLOGY, 2010, 36 (12): : 2097 - 2106
  • [25] CARDIAC IMAGING USING A PHASED-ARRAY ULTRASOUND SYSTEM .1. SYSTEM-DESIGN
    VONRAMM, OT
    THURSTONE, FL
    CIRCULATION, 1976, 53 (02) : 258 - 262
  • [26] Design of a calibratable optical antenna system based on ring array light source
    Zhou, Lei
    Cao, Yabin
    Zhong, Liang
    Liu, Jianing
    Yang, Huajun
    Jiang, Ping
    JOURNAL OF OPTICS, 2024, 26 (12)
  • [27] Design of polymer-based PMUT array for multi-frequency ultrasound imaging
    Gao, Hang
    Gijsenbergh, Pieter
    Halbach, Alexandre
    Serrahima, Clea Parcerisas
    Torri, Guilherme Brondani
    Jeong, Yongbin
    Billen, Margo
    Cheyns, David
    Haouari, Rachid
    Rottenberg, Xavier
    Rochus, Veronique
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1092 - 1095
  • [28] DESIGN OF MODULATED EXCITATION WAVEFORM BASED ON OFDM SIGNALS FOR MEDICAL ULTRASOUND IMAGING
    Chi, Cheng
    Li, Zhaohui
    2014 12TH INTERNATIONAL CONFERENCE ON SIGNAL PROCESSING (ICSP), 2014, : 102 - 107
  • [29] Design of a Foveated Imaging System Based on Liquid Crystal Microlens Array
    Xie, Xingwang
    Han, Xinjie
    Dai, Wanwan
    Xin, Zhaowei
    Wei, Dong
    Chen, Mingce
    Zhang, Xinyu
    Wang, Haiwei
    Xie, Changsheng
    TENTH INTERNATIONAL CONFERENCE ON INFORMATION OPTICS AND PHOTONICS, 2018, 10964
  • [30] DESIGN AND FABRICATION OF A 20 MHz ANNULAR ARRAY FOR MEDICAL ULTRASOUND
    Hu, Zhenhua
    Peng, Jue
    JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2012, 5 (01)