Self-calibration method and software error correction for three-dimensional coordinate measuring machines using artefact measurements

被引:41
|
作者
Kruth, J.P. [1 ]
Vanherck, P. [1 ]
De Jonge, L. [1 ]
机构
[1] Katholieke Universiteit Leuven, Leuven, Belgium
关键词
Calibration - Computer software - Error correction - Geometry - Identification (control systems) - Mathematical models - Online systems - Three dimensional;
D O I
10.1016/0263-2241(94)90024-8
中图分类号
学科分类号
摘要
The proposed self-calibration method uses an artefact, a ball plate, for error identification. Unlike other methods this ball plate need not be calibrated, i.e. the distances between the different balls on the plate need not be known accurately. After measurement of the ball plate in different positions within the measuring volume, the systematic geometrical errors, like scale errors, axes' pitch, yaw and non-orthogonality, of the geometric error model of the measuring machine can be calculated. This geometrical error model is used afterwards by the software error correction to correct on-line every single measurement carried by the measuring machine. This results in a higher accuracy of the measurement. As an example, the self-calibration method has been carried out on a CNC 3-D coordinate measuring machine and the resulting software error correction has been evaluated using step-gauge measurements. The results of the evaluation show that the systematic geometrical errors are greatly reduced.
引用
收藏
页码:157 / 167
相关论文
共 50 条
  • [21] Self-calibration method for installation angle of three-dimensional force sensor in active compliance assembly system
    Wenmin Chu
    Gen Li
    Xiang Huang
    The International Journal of Advanced Manufacturing Technology, 2024, 130 : 1969 - 1985
  • [22] Camera self-calibration with varying intrinsic parameters by an unknown three-dimensional scene
    Nabil El Akkad
    Mostafa Merras
    Abderrahim Saaidi
    Khalid Satori
    The Visual Computer, 2014, 30 : 519 - 530
  • [23] Camera self-calibration with varying intrinsic parameters by an unknown three-dimensional scene
    El Akkad, Nabil
    Merras, Mostafa
    Saaidi, Abderrahim
    Satori, Khalid
    VISUAL COMPUTER, 2014, 30 (05): : 519 - 530
  • [24] Integrated self-calibration of single axis motion for three-dimensional reconstruction of roots
    Kumar, Pankaj
    Miklavcic, Stanley J.
    IET COMPUTER VISION, 2015, 9 (06) : 850 - 856
  • [25] Optimization of three-dimensional angiographic data obtained by self-calibration of multiview imaging
    Noel, Peter B.
    Hoffmann, Kenneth R.
    Kasodekar, Snehal
    Walczak, Alan M.
    Schafer, Sebastian
    MEDICAL PHYSICS, 2006, 33 (10) : 3901 - 3911
  • [26] Three-Dimensional Morphological Measurement Method for a Fruit Tree Canopy Based on Kinect Sensor Self-Calibration
    Yang, Haihui
    Wang, Xiaochan
    Sun, Guoxiang
    AGRONOMY-BASEL, 2019, 9 (11):
  • [27] Practical experimental design and uncertainty evaluation method for dimensional and form measurements using coordinate measuring machines
    Sato, Osamu
    Takatsuji, Toshiyuki
    Matsuzaki, Kazuya
    Watanabe, Mari
    Kajima, Mariko
    Miura, Yuka
    Nakanishi, Shoichi
    MEASUREMENT, 2024, 227
  • [28] 15-term self-calibration methods for the error-correction of on-wafer measurements
    Heuermann, H
    Schiek, B
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1997, 46 (05) : 1105 - 1110
  • [29] Camera Self-Calibration with Varying Intrinsic Parameters by an Unknown Three-Dimensional Scene
    Satouri, B.
    Tairi, H.
    El Abderrahmani, A.
    Satori, K.
    INTERNATIONAL JOURNAL OF ADVANCED COMPUTER SCIENCE AND APPLICATIONS, 2016, 7 (09) : 77 - 87