HIGH-RESOLUTION ALGORITHM FOR FAN BEAM-CT SYSTEM.

被引:0
|
作者
Horiba, Isao [1 ]
Yanaka, Shigenobu [1 ]
Iwata, Akira [1 ]
Suzumura, Nobuo [1 ]
机构
[1] Hitachi Medical Co, Kashiwa, Jpn, Hitachi Medical Co, Kashiwa, Jpn
关键词
IMAGE PROCESSING - Reconstruction;
D O I
暂无
中图分类号
学科分类号
摘要
This paper theoretically analyzes the frequency component of the projection data, obtained by the third-generation CT system using a wide-angle fan beam. It is shown that the projection data contain the effective frequency component above the Nyquist frequency determined by the spacing between detectors. Then a new algorithm is proposed which utilizes that effective frequency component. The usefulness of the algorithm is examined by an experiment using a phantom and the human body. It is shown as a result that the new algorithm can realize spatial resolution corresponding to approximately twice the Nyquist frequency determined by the detector spacing. Using this algorithm, the spatial resolution of the CT image can be improved without decreasing the detector spacing.
引用
收藏
页码:19 / 29
相关论文
共 50 条
  • [41] HIGH-RESOLUTION CT OF THORACOLUMBAR FRACTURES
    BRANTZAWADZKI, M
    JEFFREY, RB
    MINAGI, H
    PITTS, LH
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1982, 3 (01) : 69 - 74
  • [42] HIGH-RESOLUTION CT SCANNING OF NECK
    HILAL, SK
    SHEPP, LA
    STEIN, JA
    JOSEPH, PM
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (03) : 369 - 369
  • [43] Perspective - High-resolution CT of the lungs
    Kazerooni, EA
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 177 (03) : 501 - 519
  • [44] REVIEW OF HIGH-RESOLUTION CT OF THE LUNG
    CORCORAN, HL
    RENNER, WR
    MILSTEIN, MJ
    RADIOGRAPHICS, 1992, 12 (05) : 917 - 939
  • [45] HIGH-RESOLUTION CT IN PULMONARY TUBERCULOSIS
    JERAJANI, H
    ANBARASU, A
    JANKHARIA, BG
    RADIOLOGY, 1995, 197 : 525 - 525
  • [46] HIGH-RESOLUTION CT OF CENTRILOBULAR LESIONS
    GREAVES, SM
    HART, EM
    BROWN, K
    YOUNG, DA
    BATRA, P
    ABERLE, DR
    RADIOLOGY, 1995, 197 : 529 - 529
  • [47] Concepts for High-Resolution CT of the Breast
    Kalender, Willi A.
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 421 - 427
  • [48] Expiratory high-resolution CT scan
    Arakawa, H
    Webb, WR
    RADIOLOGIC CLINICS OF NORTH AMERICA, 1998, 36 (01) : 189 - +
  • [49] HIGH-RESOLUTION CT OF BRONCHIOLOALVEOLAR CARCINOMA
    ADLER, B
    PADLEY, S
    MILLER, RR
    MULLER, NL
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1992, 159 (02) : 275 - 277
  • [50] HIGH-RESOLUTION CT OF MECKEL CAVE
    CHUI, M
    TUCKER, W
    HUDSON, A
    BAYER, N
    NEURORADIOLOGY, 1985, 27 (05) : 403 - 409