Determination of arylamine compounds in waste water using solid-phase extraction and reversed-phase high performance liquid chromatography

被引:0
|
作者
Zhao, Shu-Li [1 ]
Wei, Fu-Sheng [1 ]
Zou, Han-Fa [2 ]
Xu, Xiao-Bai [3 ]
机构
[1] China Natl. Environ. Monitoring Ctr., Beisihuan Dong Lu, Beijing, 100029, China
[2] Dalian Institute of Chemical Physics, Chinese Academy of Sciences, Dalian, 116012, China
[3] Res. Ctr. for Eco-Environ. Sciences, Chinese Academy of Sciences, Beijing, 100085, China
关键词
High performance liquid chromatography - Methanol - Separation - Solvent extraction - Ultraviolet detectors - Wastewater treatment;
D O I
暂无
中图分类号
学科分类号
摘要
A simple, rapid, and sensitive method for determining 10 arylamine compounds in wastewater sample using solid-phase extraction (SPE) and high performance liquid chromatography (HPLC) with UV detection is developed. An SPE procedure is described for extracting and preconcentrating the arylamines on Sep-Pak C18 cartridges with methanol as solvent. A mobile phase gradient of acetonitrile-water was used for the separation and determination of the arylamines. The detection limits of the compounds were in the range of 0.1-1.5 μg/L. Recoveries were 84-107% for 10 arylamines at concentration levels of 20 μg/L with relative standard deviations of less than 8% (n = 4).
引用
收藏
页码:717 / 724
相关论文