Microstrip gas chamber with true two-dimensional and pixel readout

被引:0
|
作者
Angelini, F. [1 ]
Bellazzini, R. [1 ]
Bosisio, L. [1 ]
Brez, A. [1 ]
Massai, M.M. [1 ]
Perret, A. [1 ]
Spandre, G. [1 ]
Torquati, M.R. [1 ]
机构
[1] INFN-Pisa, Pisa, Italy
关键词
Charging effect - Gain loss - Microstrip gas chamber - Pixel readout;
D O I
暂无
中图分类号
学科分类号
摘要
A true two-dimensional μstrip gas chamber has been constructed and successfully tested. This new detector has an effective substrate thickness of less than 2 μm. An ion implanted oxide layer of 1.8 μm thickness provides the necessary insulation between the front and back plane and permits collection on the back electrodes of a large fraction of the induced charge. The back electrode signal is used to measure the coordinate along the anode strips (X-Y readout) or to provide true space points (pixel readout). Very good imaging capabilities have been obtained in both cases. A flux of 107 particles/mm2 s has been measured without significant gain loss. No charging effect has been observed after three days continuously running at a flux of 104 particles/mm2 s, while a 15% gain loss, probably due to ageing effects, has been measured after collection on the strips of a charge corresponding to the more than six years of running at the design luminosity of LHC, at 50 cm from the beam axis.
引用
收藏
页码:229 / 235
相关论文
共 50 条
  • [31] Readout circuits for two-dimensional microbolometer arrays
    Gu, WY
    Pi, DF
    ADVANCES IN MICROELECTRONIC DEVICE TECHNOLOGY, 2001, 4600 : 198 - 202
  • [32] A triple GEM detector with two-dimensional readout
    Ziegler, M
    Sievers, P
    Straumann, U
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2001, 471 (1-2): : 260 - 263
  • [33] A MULTISTEP AVALANCHE CHAMBER WITH OPTICAL READOUT FOR USE AS A TWO-DIMENSIONAL VUV PHOTON OR PARTICLE TRACK DETECTOR
    GOOCH, TK
    GILMORE, RS
    JEFFERY, DRN
    KWAN, WL
    LLEWELLYN, TJ
    MCARTHUR, IC
    MALOS, J
    TAPPER, RJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1985, 241 (2-3): : 363 - 374
  • [34] Method to determine the true modulation ratio for comprehensive two-dimensional gas chromatography
    Pinkerton, David K.
    Parsons, Brendon A.
    Synovec, Robert E.
    JOURNAL OF CHROMATOGRAPHY A, 2016, 1476 : 114 - 123
  • [35] The study of the two-dimensional position sensitive gas electron multiplier based on strips readout
    Dong Jing
    Lue Xin-Yu
    Liu Ben
    Liu Rong-Guang
    Ma Xiao-Yan
    Wang Lan
    Chen Yuan-Bo
    Ouyang Qun
    Xie Yi-Gang
    ACTA PHYSICA SINICA, 2010, 59 (09) : 6029 - 6035
  • [36] Packaging of two-dimensional smart pixel arrays
    Kabal, DN
    Boisset, GC
    Rolston, DR
    Plant, DV
    IEEE/LEOS 1996 SUMMER TOPICAL MEETINGS - ADVANCED APPLICATIONS OF LASERS IN MATERIALS AND PROCESSING, DIGEST, 1996, : B53 - B54
  • [37] Two-dimensional and quasi-three-dimensional dosimetry of hadron and photon beams with the Magic Cube and the pixel ionization chamber
    Cirio, R.
    Garelli, E.
    Schulte, R.
    Amerio, S.
    Boriano, A.
    Bourhaleb, F.
    Coutrakon, G.
    Donetti, M.
    Giordanengo, S.
    Koss, P.
    Madon, E.
    Marchetto, F.
    Nastasi, U.
    Peroni, C.
    Santuari, D.
    Sardo, A.
    Scielzo, G.
    Stasi, M.
    Trevisiol, E.
    Phys. Med. Biol., 16 (3713-3724):
  • [38] Two-dimensional and quasi-three-dimensional dosimetry of hadron and photon beams with the Magic Cube and the Pixel Ionization Chamber
    Cirio, R
    Garelli, E
    Schulte, R
    Amerio, S
    Boriano, A
    Bourhaleb, F
    Coutrakon, G
    Donetti, M
    Giordanengo, S
    Koss, P
    Madon, E
    Marchetto, F
    Nastasi, U
    Peroni, C
    Santuari, D
    Sardo, A
    Scielzo, G
    Stasi, M
    Trevisiol, E
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (16): : 3713 - 3724
  • [39] Time Projection Chamber with Triple GEM and Pixel Readout
    Kaminski, Jochen
    Killenberg, Martin
    Bamberger, Andreas
    Blank, Hubert
    Brezina, Christoph
    Desch, Klaus
    Krautscheid, Thorsten
    Ockenfels, Walter
    Renz, Uwe
    Ummenhofer, Martin
    Wienemann, Peter
    Zimmermann, Simone
    Zwerger, Andreas
    2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 2201 - +
  • [40] A Time Projection Chamber with triple GEM and pixel readout
    Brezina, C.
    Desch, K.
    Kaminski, J.
    Killenberg, M.
    Kloeckner, F.
    Krautscheid, T.
    Renz, U.
    Schumacher, M.
    JOURNAL OF INSTRUMENTATION, 2009, 4