Automatic detection of atrial fibrillation and flutter by wave rectification method
被引:9
|作者:
Christov, I.
[1
]
Bortolan, G.
[1
]
Daskalov, I.
[1
]
机构:
[1] Centre of Biomedical Engineering, Bulgarian Academy of Sciences, Acad. G. Bonchev Str., bl., 105 1113 Sofia, Bulgaria
来源:
关键词:
Automatic testing - Electrocardiography - Measurement errors - Patient monitoring - Wave effects;
D O I:
10.1080/03091900110065942
中图分类号:
学科分类号:
摘要:
Rapid detection of atrial flutter or fibrillation is needed in intensive care or home ECG monitoring with alarm generation, and in portable monitors with warning function, etc. Detection and assessment of these atrial abnormalities is necessary in computerized morphological analysis as well, to decide whether parameter measurements should be rejected, restricted to QRS and/or T wave only, or limited to those leads where atrial flutter or fibrillation waves are less expressed. A method for the detection and measurement of atrial flutter and fibrillation in the T-P segments of the ECG is proposed. An atrial flutter/fibrillation parameter (AFF) is defined as the mean value of the differentiated filtered and rectified signal in these segments. The AFF has been measured in 329 patients from an annotated atrial flutter-fibrillation database. A threshold of AFF=0.35 % with respect to the maximum signal excursion was chosen by a heuristic algorithm, to separate patients with atrial arrhythmia. The accuracy of the method was 91.8 %. The positive and negative detection errors of the AFF classification are discussed.
引用