Development of a membrane oxygenator driven by a pneumatic pulse pump

被引:0
|
作者
Hashimoto, Shigehiro
Tominaga, Seiichi
Yoshida, Minoru
Sasaki, Yasunori
机构
关键词
Biomechanics - Biomedical engineering - Blood - Equipment testing - Flow control - Fluid mechanics - Pneumatic control equipment - Reciprocating pumps;
D O I
暂无
中图分类号
学科分类号
摘要
A new membrane oxygenator combined with a reciprocating pump was designed to shorten the extracorporeal circulation circuit and to increase the efficiency of oxygenation. At a polycarbonate cylinder, 1 700 silicone tubes were bundled and packed among elastic silicone filler. In both inflow and outflow sides of the cylinder, ball valves were equipped. The total available area of the membrane was 8.8×10-2 m2. The membrane was pneumatically driven with a reciprocating pump in performance tests, in which 100% oxygen circulated in the gas side. The mock test shows that the partial pressure of oxygen increases from 1.57×104 Pa at the inlet to 4.72×104 Pa at the outlet of the oxygenator when 2.0×10-7 m3 s-1 saline solution is pumped. The performance has been maintained in animal tests.
引用
收藏
页码:491 / 495
相关论文
共 50 条
  • [1] A MEMBRANE COMBINED OXYGENATOR AND PUMP - PRINCIPLES
    FRANTZ, SL
    CHOPRA, P
    GOLDENBERG, AL
    BROWN, L
    MILLER, FM
    DENNIS, C
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1968, 14 (APR): : 233 - +
  • [2] Design and validation of a dielectric elastomer membrane actuator driven pneumatic pump
    Linnebach, Philipp
    Rizzello, Gianluca
    Seelecke, Stefan
    SMART MATERIALS AND STRUCTURES, 2020, 29 (07)
  • [3] PROLONGED VENOARTERIAL BYPASS WITH MEMBRANE PUMP OXYGENATOR
    CHOUDHRY, KU
    BAINS, M
    BEATTIE, EJ
    RYAN, GM
    GOLDINER, PL
    RAY, C
    HOWLAND, WS
    NEW YORK STATE JOURNAL OF MEDICINE, 1972, 72 (22) : 2804 - +
  • [4] DEVELOPMENT OF A SIMPLIFIED MEMBRANE OXYGENATOR
    CRESCENZI, AA
    HOFSTRA, PC
    DIBENEDETTO, A
    SZE, KC
    FOSTER, BH
    GLASS, P
    CLAFF, CL
    COOPER, P
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1959, 5 : 148 - &
  • [5] Wrist flexible heart pulse sensor integrated with a soft pump and a pneumatic balloon membrane
    Yamaguchi, Takafumi
    Yamamoto, Daisuke
    Arie, Takayuki
    Akita, Seiji
    Takei, Kuniharu
    RSC ADVANCES, 2020, 10 (29) : 17353 - 17358
  • [6] COMBINED MEMBRANE PUMP-OXYGENATOR - DESIGN AND TESTING
    LEWIS, FR
    TYLKE, JA
    WINCHELL, HS
    TRANSACTIONS AMERICAN SOCIETY FOR ARTIFICIAL INTERNAL ORGANS, 1974, A 20 : 253 - 261
  • [7] A NEW MEMBRANE PUMP-OXYGENATOR FOR PROLONGED SUPPORT
    LANDE, AJ
    PARKER, B
    KASTER, R
    RASSMAN, W
    TANAKA, S
    LILLEHEI, CW
    JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1968, 204 (06): : 529 - &
  • [8] CONSIDERATIONS IN THE DEVELOPMENT AND USE OF A PUMP-OXYGENATOR
    BAHNSON, HT
    BULLETIN OF THE JOHNS HOPKINS HOSPITAL, 1958, 102 (04): : 213 - 213
  • [9] Development of a heat-driven pulse pump for spacecraft applications
    Benner, SM
    Martins, MS
    IECEC 96 - PROCEEDINGS OF THE 31ST INTERSOCIETY ENERGY CONVERSION ENGINEERING CONFERENCE, VOLS 1-4, 1996, : 1416 - 1420
  • [10] Progress with the development of the intravenous membrane oxygenator
    Hattler, BG
    Federspiel, WJ
    PERFUSION-UK, 1999, 14 (04): : 311 - 315