Characterization of an external electro-optic sampling probe: Influence of probe height on distortion of measured voltage pulses

被引:0
|
作者
Seitz, S. [1 ]
Bieler, M. [1 ]
Hein, G. [1 ]
Pierz, K. [1 ]
Siegner, U. [1 ]
Schmückle, F.J. [2 ]
Heinrich, W. [2 ]
机构
[1] Physikalisch-Technische Bundesanstalt, Bundesallee 100, D-38116 Braunschweig, Germany
[2] Ferdinand-Braun-Institut für Hochfrequenztechnik, Gustav-Kirchhoff-Strasse 4, D-12489 Berlin, Germany
来源
Journal of Applied Physics | 2006年 / 100卷 / 11期
关键词
We experimentally and theoretically investigate the distortion that an external electro-optic sampling tip imprints on a measured voltage pulse that propagates along a coplanar stripline. By using the electro-optic effect of the GaAs substrate on which the stripline is fabricated we obtain the undisturbed voltage pulse; which serves as a reference pulse. This allows us to quantitatively compare the voltage pulse measured with the external electro-optic sampling tip to the undisturbed voltage pulse. The sensitivity of the sampling measurement and the distortion of the voltage pulse are determined for different heights of the tip which is lifted off the stripline. The experimental data agree very well with numerical finite difference calculations. We find that small changes of the tip height already influence the measurements strongly. From the experiments we infer an optimum working height of the sampling tip of only 0.5 μm. Our investigations help to improve the accuracy of electro-optic sampling measurements. © 2006 American Institute of Physics;
D O I
暂无
中图分类号
学科分类号
摘要
Journal article (JA)
引用
收藏
相关论文
共 50 条
  • [1] Characterization of an external electro-optic sampling probe:: Influence of probe height on distortion of measured voltage pulses
    Seitz, S.
    Bieler, M.
    Hein, G.
    Pierz, K.
    Siegner, U.
    Schmueckle, F. J.
    Heinrich, W.
    JOURNAL OF APPLIED PHYSICS, 2006, 100 (11)
  • [2] Correction of picosecond voltage pulses measured with external electro-optic sampling tips
    Seitz, S.
    Bieler, M.
    Hein, G.
    Pierz, K.
    Siegner, U.
    Schmueckle, F. J.
    Heinrich, W.
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2007, 18 (05) : 1353 - 1360
  • [3] Distortion of terahertz pulses in electro-optic sampling
    Bakker, HJ
    Cho, GC
    Kurz, H
    Wu, Q
    Zhang, XC
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 1998, 15 (06) : 1795 - 1801
  • [4] Tiny Gallium Arsenide probe for electro-optic sampling
    Li, HL
    Zhang, XT
    Chen, ZG
    Jia, G
    Zhou, ZX
    OPTICAL DESIGN AND TESTING, 2002, 4927 : 286 - 289
  • [5] SUBPICOSECOND SAMPLING USING A NONCONTACT ELECTRO-OPTIC PROBE
    NAGATSUMA, T
    SHIBATA, T
    SANO, E
    IWATA, A
    JOURNAL OF APPLIED PHYSICS, 1989, 66 (09) : 4001 - 4009
  • [6] ELECTRO-OPTIC SAMPLING OF ULTRASHORT HIGH-VOLTAGE PULSES
    STEARNS, DG
    JOURNAL OF APPLIED PHYSICS, 1989, 65 (03) : 1308 - 1316
  • [7] Broad-bandwidth THz pulse characterization through electro-optic sampling with narrow-bandwidth probe pulses
    van Tilborg, J.
    Bakker, D. J.
    Matlis, N. H.
    Leemans, W. P.
    2012 CONFERENCE ON LASERS AND ELECTRO-OPTICS (CLEO), 2012,
  • [8] External electro-optic measuring system with high spatial resolution and high voltage sensitivity by using an electro-optic solid immersion probe
    Chen, ZG
    Jia, G
    Yi, MB
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2001, 34 (20) : 3078 - 3082
  • [9] Influence of wavelength of probe laser on terahertz free-space electro-optic sampling detection
    Du Hai-Wei
    Peng Xiao-Yu
    JOURNAL OF INFRARED AND MILLIMETER WAVES, 2017, 36 (03) : 266 - +
  • [10] Highly precise and accurate terahertz polarization measurements based on electro-optic sampling with polarization modulation of probe pulses
    Nemoto, Natsuki
    Higuchi, Takuya
    Kanda, Natsuki
    Konishi, Kuniaki
    Kuwata-Gonokami, Makoto
    OPTICS EXPRESS, 2014, 22 (15): : 17915 - 17929