Fretting-corrosion at the Implant-Abutment Interface Simulating Clinically Relevant Conditions

被引:2
|
作者
Sun, Yani [1 ]
Shukla, Apurwa [2 ]
Ramachandran, Remya Ampadi [3 ]
Kanniyappan, Hemalatha [4 ]
Yang, Bin [2 ]
Harlow, Rand [2 ]
Campbell, Stephen D. [2 ]
Thalji, Ghadeer [2 ]
Mathew, Mathew [2 ,3 ,4 ]
机构
[1] Univ Illinois, Dept Civil Mat & Environm Engn, Chicago, IL USA
[2] Univ Illinois, Coll Dent, Dept Restorat Dent, Chicago, IL 60607 USA
[3] Univ Illinois, Dept Biomed Engn, Chicago, IL USA
[4] Univ Illinois, Sch Med Rockford, Dept Biomed Sci, Rockford, IL USA
关键词
Fretting-corrosion; Implant-abutment interface; Dental erosion; Parafunctional habits; TITANIUM; BEHAVIOR; INFLAMMATION; ZIRCONIA; MICROGAP;
D O I
10.1016/j.dental.2024.08.007
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: Implant treatment is provided to individuals with normal, idealized masticatory forces and also to patients with parafunctional habits such as grinding, clenching, and bruxing. Dental erosion is a common increasing condition and is reported to affect 32 % of adults, increasing with age. This oral environment is conducive to tribocorrosion and the potential loss of materials from the implant surfaces and interfaces with prosthetic components. Although several fretting-corrosion studies have been reported, until now, no study has simulated clinically relevant micromotion. Therefore, our aim is to investigate fretting-corrosion using our new micro-fretting corrosion system, simulating clinical conditions with 5 mu m motion at the implant-abutment interface under various occlusal loads and acidic exposures. Methods: We simulated four conditions in an oral environment by varying the contact load (83 N and 233 N) and pH levels (3 and 6.5). The commonly used dental implant material, Grade IV titanium, and abutment material Zirconia (ZrO2)/ Grade IV titanium were selected as testing couple materials. Artificial saliva was employed to represent an oral environment. In addition, a standard tribocorrosion protocol was followed, and the pin was controlled to oscillate on the disk with an amplitude of 5 mu m during the mastication stage. After the testing, 3D profilometry and scanning electron microscopy (SEM) with energy dispersive spectroscopy (EDS) were utilized to analyze the worn surfaces. Inductively coupled plasma mass spectrometry (ICP-MS) was also used to measure the metal ion release. Results: Energy ratios were below 0.2, indicating a fretting regime of partial slip for all groups. Open-circuit potential (OCP) and electrochemical impedance spectroscopy (EIS) were analyzed to compare the electrochemical behavior among groups. As a result, corrosive damage was observed to be more in the Ti4- Ti4 groups than in Zr-Ti4 ones, whereas more mechanical damage was found in the Zr-Ti4 groups than in the Ti4-Ti4 groups. Possible mechanisms were proposed in the discussion to explain these findings. Significance: The results observed from this study might be helpful to clinicians with implant selection. For example, for patients with bruxism, a titanium implant paired with a titanium abutment may be preferable, while patients with GERD may benefit more from a titanium implant paired with a zirconia abutment.
引用
收藏
页码:1823 / 1831
页数:9
相关论文
共 50 条
  • [21] Micromotion and dynamic fatigue properties of the denial implant-abutment interface
    Gratton, DG
    Aquilino, SA
    Stanford, CM
    JOURNAL OF PROSTHETIC DENTISTRY, 2001, 85 (01): : 47 - 52
  • [22] INVESTIGATION OF AXIAL DISPLACEMENT OF SUPRASTRUCTURES AT THE CONICAL IMPLANT-ABUTMENT INTERFACE
    Mishin, Dmitry
    Kalinovsky, Sergei
    Ilyasov, Vyacheslav
    Dedova, Veronika
    STRUCTURAL INTEGRITY AND LIFE-INTEGRITET I VEK KONSTRUKCIJA, 2023, 23 (01): : 65 - 69
  • [23] Influence of connection geometry on dynamic micromotion at the implant-abutment interface
    Meng, Joe C.
    Everts, Joshua E.
    Qian, Fang
    Gratton, David G.
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2007, 20 (06) : 623 - 625
  • [24] Effect of the implant-abutment interface on peri-implant tissues: A systematic review
    Bishti, Shaza
    Strub, Joerg R.
    Att, Wael
    ACTA ODONTOLOGICA SCANDINAVICA, 2014, 72 (01) : 13 - 25
  • [25] The Role of Chlorhexidine on Endotoxin Penetration to the Implant-Abutment Interface (IAI)
    Koutouzis, Theofilos
    Gadalla, Hana
    Kettler, Zachary
    Elbarasi, Amelsaad
    Nonhoff, Joerg
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2015, 17 (03) : 476 - 482
  • [26] Effects of two fretting damage modes on the dental implant-abutment interface and the generation of metal wear debris: An in vitro study
    Gao, Jing
    Min, Jie
    Chen, Xin
    Yu, Ping
    Tan, Xin
    Zhang, Qianqian
    Yu, Haiyang
    FATIGUE & FRACTURE OF ENGINEERING MATERIALS & STRUCTURES, 2021, 44 (03) : 847 - 858
  • [27] Hip implant modular junction: The role of CoCrMo alloy microstructure on fretting-corrosion
    Manthe, Jacob
    Cheng, Kai Yuan
    Bijukumar, Divya
    Barba, Mark
    Pourzal, Robin
    Neto, Mozart
    Mathew, Mathew T.
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2022, 134
  • [28] In vitro evaluation of bacterial leakage along the implant-abutment interface of different implant systems
    Steinebrunner, L
    Wolfart, S
    Bössmann, K
    Kern, M
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2005, 20 (06) : 875 - 881
  • [29] Bacterial leakage along the implant-abutment interface of premachined or cast components
    do Nascimento, C.
    Barbosa, R. E. S.
    Issa, J. P. M.
    Watanabe, E.
    Ito, I. Y.
    Junior, R. F. Albuquerque
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2008, 37 (02) : 177 - 180
  • [30] Evaluation of the marginal fit at implant-abutment interface by optical coherence tomography
    Kikuchi, Keisuke
    Akiba, Norihisa
    Sadr, Alireza
    Sumi, Yasunori
    Tagami, Junji
    Minakuchi, Shunsuke
    JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (05)