Fretting-corrosion at the Implant-Abutment Interface Simulating Clinically Relevant Conditions

被引:2
|
作者
Sun, Yani [1 ]
Shukla, Apurwa [2 ]
Ramachandran, Remya Ampadi [3 ]
Kanniyappan, Hemalatha [4 ]
Yang, Bin [2 ]
Harlow, Rand [2 ]
Campbell, Stephen D. [2 ]
Thalji, Ghadeer [2 ]
Mathew, Mathew [2 ,3 ,4 ]
机构
[1] Univ Illinois, Dept Civil Mat & Environm Engn, Chicago, IL USA
[2] Univ Illinois, Coll Dent, Dept Restorat Dent, Chicago, IL 60607 USA
[3] Univ Illinois, Dept Biomed Engn, Chicago, IL USA
[4] Univ Illinois, Sch Med Rockford, Dept Biomed Sci, Rockford, IL USA
关键词
Fretting-corrosion; Implant-abutment interface; Dental erosion; Parafunctional habits; TITANIUM; BEHAVIOR; INFLAMMATION; ZIRCONIA; MICROGAP;
D O I
10.1016/j.dental.2024.08.007
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: Implant treatment is provided to individuals with normal, idealized masticatory forces and also to patients with parafunctional habits such as grinding, clenching, and bruxing. Dental erosion is a common increasing condition and is reported to affect 32 % of adults, increasing with age. This oral environment is conducive to tribocorrosion and the potential loss of materials from the implant surfaces and interfaces with prosthetic components. Although several fretting-corrosion studies have been reported, until now, no study has simulated clinically relevant micromotion. Therefore, our aim is to investigate fretting-corrosion using our new micro-fretting corrosion system, simulating clinical conditions with 5 mu m motion at the implant-abutment interface under various occlusal loads and acidic exposures. Methods: We simulated four conditions in an oral environment by varying the contact load (83 N and 233 N) and pH levels (3 and 6.5). The commonly used dental implant material, Grade IV titanium, and abutment material Zirconia (ZrO2)/ Grade IV titanium were selected as testing couple materials. Artificial saliva was employed to represent an oral environment. In addition, a standard tribocorrosion protocol was followed, and the pin was controlled to oscillate on the disk with an amplitude of 5 mu m during the mastication stage. After the testing, 3D profilometry and scanning electron microscopy (SEM) with energy dispersive spectroscopy (EDS) were utilized to analyze the worn surfaces. Inductively coupled plasma mass spectrometry (ICP-MS) was also used to measure the metal ion release. Results: Energy ratios were below 0.2, indicating a fretting regime of partial slip for all groups. Open-circuit potential (OCP) and electrochemical impedance spectroscopy (EIS) were analyzed to compare the electrochemical behavior among groups. As a result, corrosive damage was observed to be more in the Ti4- Ti4 groups than in Zr-Ti4 ones, whereas more mechanical damage was found in the Zr-Ti4 groups than in the Ti4-Ti4 groups. Possible mechanisms were proposed in the discussion to explain these findings. Significance: The results observed from this study might be helpful to clinicians with implant selection. For example, for patients with bruxism, a titanium implant paired with a titanium abutment may be preferable, while patients with GERD may benefit more from a titanium implant paired with a zirconia abutment.
引用
收藏
页码:1823 / 1831
页数:9
相关论文
共 50 条
  • [1] The implant-abutment interface of alumina and zirconia abutments
    Yuezueguellue, Bulem
    Avci, Mehmet
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2008, 10 (02) : 113 - 121
  • [2] Persistent acute inflammation at the implant-abutment interface
    Broggini, N
    McManus, LM
    Hermann, JS
    Medina, RU
    Oates, TW
    Schenk, RK
    Buser, D
    Mellonig, JT
    Cochran, DL
    JOURNAL OF DENTAL RESEARCH, 2003, 82 (03) : 232 - 237
  • [3] Radiographical Evaluation of the Gap at the Implant-Abutment Interface
    Papavassiliou, Harris
    Kourtis, Stefanos
    Katerelou, Julia
    Chronopoulos, Vasillios
    JOURNAL OF ESTHETIC AND RESTORATIVE DENTISTRY, 2010, 22 (04) : 235 - 250
  • [4] Parameters Determining Micromotion at the Implant-Abutment Interface
    Karl, Matthias
    Taylor, Thomas D.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2014, 29 (06) : 1338 - 1347
  • [5] Peri-implant inflammation defined by the implant-abutment interface
    Broggini, N
    McManus, LM
    Hermann, JS
    Medina, R
    Schenk, RK
    Buser, D
    Cochran, DL
    JOURNAL OF DENTAL RESEARCH, 2006, 85 (05) : 473 - 478
  • [6] Fretting-corrosion behavior on dental implant connection in human saliva
    Corne, Pascale
    De March, Pascal
    Cleymand, Franck
    Geringer, Jean
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2019, 94 : 86 - 92
  • [7] Cross-sectional analysis of the implant-abutment interface
    Coelho, A. L.
    Suzuki, M.
    Dibart, S.
    Da Silva, N.
    Coelho, P. G.
    JOURNAL OF ORAL REHABILITATION, 2007, 34 (07): : 508 - 516
  • [8] Microbial leakage and marginal fit of the implant-abutment interface
    Jansen, VK
    Conrads, G
    Richter, EJ
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 1997, 12 (04) : 527 - 540
  • [9] The fracture strength by a torsion test at the implant-abutment interface
    Fumihiko Watanabe
    Kazuhiko Hiroyasu
    Kazuhiko Ueda
    International Journal of Implant Dentistry, 1 (1)
  • [10] Effect of Cyclic Loading on Micromotion at the Implant-Abutment Interface
    Karl, Matthias
    Taylor, Thomas D.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2016, 31 (06) : 1292 - 1302