A single-stage purification method for the precise determination of zirconium isotopic composition in geological samples by double spike MC-ICP-MS

被引:0
|
作者
Feng, Lanping [1 ]
Wang, Chen [1 ]
Han, Yan [1 ]
Zhang, Wen [1 ]
Guo, Jing-Liang [1 ]
Zhou, Lian [1 ]
Hu, Zhaochu [1 ]
机构
[1] State Key Laboratory of Geological Processes and Mineral Resources, School of Earth Sciences, China University of Geosciences, Wuhan,430074, China
基金
中国国家自然科学基金;
关键词
Geological reference material - Geological samples - HNO 3 - Isotope ratio - Isotopic composition - Multi-collector inductively coupled plasma mass spectrometry - Precise determinations - Purification method - Single stage - Stable isotopic fractionation;
D O I
10.1039/d4ja00332b
中图分类号
学科分类号
摘要
Interest in studying zirconium (Zr) stable isotopic fractionation in both low and high-temperature environments has increased significantly in recent years. Multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) is the preferred technique for stable Zr isotope analysis due to its low sample consumption and high precision. However, the multi-column separation steps required to remove interferences and meet the stringent matrix requirements for Zr isotopes during MC-ICP-MS testing typically make the sample preparation process time-consuming. Here, we present a new, simple, and efficient single-pass chromatographic exchange technique for the chemical isolation of Zr from complex sample matrices. This new separation protocol is based on DGA resin and consists of only three elution steps and two types of acid eluents (15 mL 7 mol L−1 HNO3 and 5 mL 3 mol L−1 HNO3 + 0.1 mol L−1 HF), which afford straightforward separation of Zr with high yield, good purity, and low blank levels. Zr isotope ratios were measured using Neptune Plus MC-ICP-MS with the 91Zr-96Zr double spike in combination with standard-sample bracketing for mass bias correction. The validity of the proposed method is demonstrated by analyzing eight geological reference materials (including BCR-2, which has an extremely high Mo content of ∼248 μg g−1) with satisfactory results that are in good agreement with published values. Based on repeated analyses of pure standard solutions and geological reference materials, the long-term measurement reproducibility is better than ±0.048‰ for δ94Zr. Therefore, the proposed method can be used for the rapid and efficient determination of stable Zr isotope ratios in geological samples. © 2024 The Royal Society of Chemistry.
引用
收藏
页码:3035 / 3047
相关论文
共 50 条
  • [41] Determination of the Isotopic Composition of Ytterbium by MC-ICP-MS Using an Optimized Regression Model
    Lin, Ran
    Lin, Jie
    Zong, Keqing
    Yang, Ao
    Chen, Kang
    Liu, Yongsheng
    Hu, Zhaochu
    ANALYTICAL CHEMISTRY, 2022, 94 (20) : 7200 - 7209
  • [42] Single-Stage Extraction Technique for Ce Stable Isotopes and Measurement by MC-ICP-MS
    Liu, Fang
    Zhang, Zhaofeng
    Li, Xin
    An, Yajun
    Liu, Yufei
    Chen, Kaiyun
    Bao, Zhian
    Li, Chunhui
    ANALYTICAL CHEMISTRY, 2021, 93 (37) : 12524 - 12531
  • [43] Precise measurements of d???????88/86Sr for twenty geological reference materials by double-spike MC-ICP-MS
    Chen, Xu-Qi
    Zeng, Zhen
    Yu, Hui-Min
    Sun, Nan
    Huang, Fang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2022, 479
  • [44] High precision cadmium isotope analysis of geological reference materials by double spike MC-ICP-MS
    Peng, Huan
    He, Dong
    Guo, Rui
    Liu, Xing
    Belshaw, Nicholas S.
    Zheng, Hongtao
    Hu, Shenghong
    Zhu, Zhenli
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2021, 36 (02) : 390 - 398
  • [45] An efficient Cd two-stage column system for high-precision determination of Cd isotopic compositions by double spike MC-ICP-MS
    Zhang, Zheyuan
    Li, Tao
    Li, Baichan
    Lin, Yibo
    Li, Gaojun
    Chen, Tianyu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (04) : 1142 - 1151
  • [46] One column procedure for Hf purification in geological samples using anion exchange chromatography and its isotopic analyses by MC-ICP-MS
    Yang Y H
    Zhang H F
    Liu Y
    Xie L W
    Qi C S
    Tu X L
    ACTA PETROLOGICA SINICA, 2007, 23 (02) : 227 - 232
  • [47] Chromatographic purification of Ca and Mg from biological and geological samples for isotope analysis by MC-ICP-MS
    Bao, Zhian
    Zong, Chunlei
    Chen, Kaiyun
    Lv, Nan
    Yuan, Honglin
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2020, 448
  • [48] A high performance method for the accurate and precise determination of silicon isotopic compositions in bulk silicate rock samples using laser ablation MC-ICP-MS
    Zhang, Chenxi
    Zhao, He
    Zhang, Wen
    Luo, Tao
    Li, Ming
    Zong, Keqing
    Liu, Yongsheng
    Hu, Zhaochu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (09) : 1887 - 1896
  • [49] A simplified dual-column system for Li purification and isotopic analysis of geological materials by MC-ICP-MS
    Hu, Wenjie
    Chen, Lu
    Geng, Xianlei
    Liang, Zhengwei
    Tian, Shihong
    Talanta, 2025, 283
  • [50] A fast double-stack column chemical separation of Zn and Mo from geological samples for isotopic analysis by MC-ICP-MS
    Han, Yan
    Zhou, Lian
    Shi, Minghui
    Hu, Yating
    Zhang, Ge
    Hou, Xin
    Feng, Lanping
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2024, 39 (01) : 269 - 280