Absorbed Dose Measurements for Teletheraphy Prediction of Superficial Dose Using Halcyon Linear Accelerator

被引:0
|
作者
Njinga, Raymond Limen [1 ]
Samuel, Adeneye [2 ]
Ajimo, Akinyode [1 ]
Olowookere, Christopher [1 ]
机构
[1] Department of Physics, University of Medical Sciences Ondo, Ondo State, Nigeria
[2] Department of Radio Biology, Radio Diagnosis and Radiotherapy, College of Medicine University of Lagos, NSIA-LUTH Cancer Centre: Idiaraba, Lagos, Nigeria
关键词
Dosimeters - Dosimetry - Ionization chambers - Linear accelerators;
D O I
10.22038/ijmp.2023.64557.2107
中图分类号
学科分类号
摘要
Introduction: Measurement of entrance dose and dose at different depth is essential to avoid over dose and under dose of patients. The aim of this study is to verify the variation in the absorbed dose using a water equivalent material. Material and Methods: The plastic phantom was arranged on the couch of the halcyon linear accelerator by Varian with the farmer ionization chamber inserted and connected to the electrometer. The image of the setup was taken using the High-Quality Single 1280x1280x16 higher on the service mode, to check the alignment with the isocenter. The beam quality TPR20, 10 (Tissue phantom ratio) was done to check the beam quality of the machine at a field size of 10 cm x 10 cm. The calibration was done using SAD type set-up at a depth of 5 cm. This process was repeated for ten consecutive weeks and the values recorded. Results: The results of the beam output for the teletherapy machine were satisfactory and accepted in comparison with the commissioned measurement of 0.62. The beam quality TPR20, 10 (Tissue phantom ratio) was reasonable with respect to the beam quality of the machine at a field size of 10 cm x 10 cm. Conclusion: The results of the beam quality and the absorbed dose rate showed a good consistency over the period of ten weeks with the commissioned measurement value. © (2024), (Mashhad University of Medical Sciences). All rights reserved.
引用
收藏
页码:265 / 271
相关论文
共 50 条
  • [41] TLD calibration and absorbed dose measurement in a radiation-induced liver injury model under a linear accelerator
    Xiao, Hui-Hui
    Liu, Ling-Ling
    Li, Wen-Yi
    Li, Bing-Bing
    Cui, Xiang-Li
    Li, Jie
    Li, Tao-Sheng
    Hu, Zong-Tao
    NUCLEAR SCIENCE AND TECHNIQUES, 2023, 34 (04)
  • [42] Determination of absorbed dose for photon and electron beams from a linear accelerator by comparing various protocols on different phantoms
    Banjade, DP
    Tajuddin, AA
    Shukri, A
    APPLIED RADIATION AND ISOTOPES, 2001, 55 (02) : 235 - 243
  • [43] TLD calibration and absorbed dose measurement in a radiation-induced liver injury model under a linear accelerator
    Hui-Hui Xiao
    Ling-Ling Liu
    Wen-Yi Li
    Bing-Bing Li
    Xiang-Li Cui
    Jie Li
    Tao-Sheng Li
    Zong-Tao Hu
    Nuclear Science and Techniques, 2023, 34
  • [44] Linear energy transfer dependence of a normoxic polymer gel dosimeter investigated using proton beam absorbed dose measurements
    Gustavsson, H
    Bjäck, SAJ
    Medin, J
    Grusell, E
    Olsson, LE
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17): : 3847 - 3855
  • [45] In vivo dosimetry for head and neck carcinoma: determination of target absorbed dose from entrance and exit absorbed dose measurements
    Farhat, L.
    Besbes, M.
    Bridier, A.
    Daoud, J.
    EUROPEAN PHYSICAL JOURNAL-APPLIED PHYSICS, 2011, 54 (01):
  • [46] Absorbed dose to water reference dosimetry using solid phantoms in the context of absorbed-dose protocols
    Seuntjens, J
    Olivares, M
    Evans, M
    Podgorsak, E
    MEDICAL PHYSICS, 2005, 32 (09) : 2945 - 2953
  • [47] Comparison of skin dose in IMRT and VMAT with TrueBeam and Halcyon linear accelerator for whole breast irradiation (vol 47, pg 443, 2024)
    Seok, Jae Hyun
    Ahn, So Hyun
    Ahn, Woo Sang
    Choi, Dong Hyeok
    Shin, Seong Soo
    Choi, Wonsik
    Jung, In-hye
    Lee, Rena
    Kim, Jin Sung
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, 47 (02) : 453 - 453
  • [48] Measurements of superficial dose distributions in radiation therapy using translucent cryogel dosimeters
    Eyadeh, M. M.
    Wierzbicki, M.
    Diamond, K. R.
    9TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY, 2017, 847
  • [49] RADIATION ABSORBED DOSE ESTIMATE FOR RB-82 USING INVIVO MEASUREMENTS IN MAN
    RYAN, J
    HARPER, P
    STARK, V
    PETERSON, E
    LATHROP, K
    JOURNAL OF NUCLEAR MEDICINE, 1984, 25 (05) : P94 - P94
  • [50] ABSORBED DOSE MEASUREMENTS USING PARALLEL PLATE POLYSTYRENE IONIZATION CHAMBERS IN POLYSTYRENE PHANTOMS
    HOLT, JG
    BUFFA, A
    PERRY, DJ
    MA, IC
    MCDONALD, JC
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (11/1): : 2031 - 2038