Examination of Magnetic Shield for Capsule Endoscope Magnetic Guidance Magnet

被引:0
|
作者
Ozawa Y. [1 ]
Morikawa N. [1 ]
Tashiro K. [1 ]
Wakiwaka H. [1 ]
Mizuno T. [1 ]
Ohmiya N. [2 ]
机构
[1] Tashiro Lab., School of Engineering, Shinshu University, 4-17-1, Wakasato, Nagano
[2] Fujita Medical University, 1-98, Tarakugokubo, Kutsukake-cho, Toyoake
关键词
magnetic dipole moment; magnetic shield; magnetic shield ratio;
D O I
10.1541/ieejfms.142.383
中图分类号
学科分类号
摘要
In this report, the magnetic dipole moment was used to determine the need for a magnetic shield for capsule endoscopy guiding device. In addition, the index of the number of layers of MSC was examined from the evaluation of the magnetic shield characteristics of the open pail can, which is attracting attention as MSC, and the magnetic dipole moment. As a result, it became clear that the magnetic guiding device needs to be magnetically shielded at 11 Am2 and above. In addition, when the shielding performance of the open pail can was evaluated, it was shown by analysis that the magnetic flux density could be reduced by up to 1.53 mT and the relative permeability was 1000. Finally, it was found that when the magnetic dipole moment of the magnetic inducer is less than 41 Am2, only the open pail can is required, and when it is 41 Am2 or more, the MSC needs to have two or more layers. © 2022 The Institute of Electrical Engineers of Japan.
引用
收藏
页码:383 / 389
页数:6
相关论文
共 50 条
  • [1] Examination of Magnetic Force Calculation Formula on the Central Axis of the Magnetic Guiding Magnet for Capsule Endoscope
    Morikawa N.
    Tashiro K.
    Wakiwaka H.
    Mizuno T.
    Ohmiya N.
    IEEJ Transactions on Fundamentals and Materials, 2024, 144 (03) : 111 - 118
  • [2] Magnetic actuation and guidance mechanism for active capsule endoscope
    Wang, Xiaona
    Meng, Max Q. -H.
    2007 IEEE/ASME INTERNATIONAL CONFERENCE ON ADVANCED INTELLIGENT MECHATRONICS, VOLS 1-3, 2007, : 689 - +
  • [3] A Locomotion Mechanism with External Magnetic Guidance for Active Capsule Endoscope
    Wang, Xiaona
    Meng, Max Q. -H.
    Chen, Xijun
    2010 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2010, : 4375 - 4378
  • [4] A Novel Control Method of Magnetic Navigation Capsule Endoscope for Gastrointestinal Examination
    Sun, Hongbo
    Liu, Jianhua
    Wang, Lei
    Niu, Chaoqun
    Wang, Qiuliang
    IEEE TRANSACTIONS ON MAGNETICS, 2022, 58 (01)
  • [5] Examination of simple analytical model for designing magnetic shield for storage of magnetic guiding magnet
    Ozawa Y.
    Tashiro K.
    Wakiwaka H.
    Mizuno T.
    Ohmiya N.
    Tashiro, Kunihisa (Tashiro@shinshu-u.ac.jp), 1600, Institute of Electrical Engineers of Japan (141): : 333 - 338
  • [6] A NEW TOOL OF MAGNETIC-ASSISTED CAPSULE ENDOSCOPE FOR UPPER GASTROINTESTINAL EXAMINATION
    Hsieh, Yi-Lun
    Chen, Chen-Hsun
    Wu, Meng-Chieh
    Shen, Chun-Sheng
    Hsu, Wen-Hung
    Wu, Deng-Chyang
    Wu, Jeng-Yih
    GASTROENTEROLOGY, 2021, 160 (06) : S752 - S753
  • [7] On localizing a Capsule Endoscope using Magnetic Sensors
    Moussakhani, Babak
    Ramstad, Tor
    Flam, John T.
    Balasingham, Ilangko
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 4058 - 4062
  • [8] BIVARIABLE MAGNETIC MOMENT CONTROL OF A CAPSULE ENDOSCOPE
    Chi, Minglu
    Liu, Rongsheng
    Chang, Cheng
    Wang, Yuanli
    Zheng, Huadong
    Qian, Xiaoyan
    Ren, Qinchao
    Ren, Ruihua
    INTERNATIONAL JOURNAL OF ROBOTICS & AUTOMATION, 2022, 37 (06): : 512 - 519
  • [9] Guided magnetic actuator for active capsule endoscope
    Wang, Xiaona
    Meng, Max Q. -H.
    2007 2ND IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS, VOLS 1-3, 2007, : 924 - +
  • [10] Fabrication of magnetic actuator for use in a capsule endoscope
    Sendoh, M
    Ishiyama, K
    Arai, KI
    IEEE TRANSACTIONS ON MAGNETICS, 2003, 39 (05) : 3232 - 3234