Study of open-type magnetically shielded room for MRI

被引:0
|
作者
Kamata K. [1 ]
Tamari Y. [1 ]
Temma S. [1 ]
Goto K. [1 ]
Yunokuchi K. [2 ]
机构
[1] National Institute of Technology, Kagoshima College, 1460-1, Shinko, Hayato, Kirishima-shi, Kagoshima
[2] Kagoshima University, 1-21-40, Korimoto, Kagoshima-shi, Kagoshima
关键词
3D magnetic field analysis; Amenity improvement; Lattice shape; Magnetic material plate; Open-type magnetically shielded room; Square cylinder made by magnetic material;
D O I
10.1541/ieejeiss.138.799
中图分类号
学科分类号
摘要
The magnetic resonance imaging (MRI) is usually installed in a magnetically shielded room (MSR), in order to reduce the leakage flux density to less than 0.5 mT outside the room. The small space enclosed by the wall causes patients a great deal of stress and discomfort because they are isolated from the operators and from the scenery outside. In order to improve amenity for patients in hospitals, open type of MSRs for MRI, which use canceling coils, combined with square cylinders made by magnetic and conductive material, and magnetic material plates instead of magnetic walls. The magnetic shielding effects of open type of magnetically shielded room combined with square cylinders model made by magnetic, conductive material, magnetic material plates model and lattice shape model made of magnetic material for MRI are estimated by 3-D magnetic field analysis. In this report, we clarified simple lattice of magnetic plate with gap to improve shielding performance using 3D magnetic field analysis. © 2018 The Institute of Electrical Engineers of Japan.
引用
收藏
页码:799 / 804
页数:5
相关论文
共 50 条
  • [41] Numerical Modeling and Material Characterization for Multilayer Magnetically Shielded Room Design
    Canova, A.
    Freschi, F.
    Giaccone, L.
    Repetto, M.
    IEEE TRANSACTIONS ON MAGNETICS, 2018, 54 (03)
  • [42] Optimization of the design of open-type ultraviolet irradiators
    Naumenko D.E.
    Levchenko V.A.
    Vasiliev A.I.
    Applied Physics, 2024, (03): : 99 - 103
  • [43] Open-type magnetocardiograph with cylindrical magnetic shield
    Seki, Y
    Kandori, A
    Suzuki, D
    Ohnuma, M
    APPLIED PHYSICS LETTERS, 2005, 86 (24) : 1 - 3
  • [44] Evaluation of a high-performance magnetically shielded room for biomagnetic measurement
    Harakawa, K
    Kajiwara, G
    Kazami, K
    Ogata, H
    Kado, H
    IEEE TRANSACTIONS ON MAGNETICS, 1996, 32 (06) : 5256 - 5260
  • [45] Note: Magnetic noise from the inner wall of a magnetically shielded room
    Khan, Sheraz
    Cohen, David
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (05):
  • [46] Simplified modelling of an open-type reciprocating compressor
    Winandy, E
    Saavedra, OC
    Lebrun, J
    INTERNATIONAL JOURNAL OF THERMAL SCIENCES, 2002, 41 (02) : 183 - 192
  • [48] Open-Type or Hermetically Sealed Turbocompressors.
    Hartmann, Klaus
    1600, (01):
  • [49] Study on the spatial distribution of the chloride ion supply in the superstructure of an open-type wharf
    Tanaka, Y.
    Kawabata, Y.
    Kato, E.
    LIFE-CYCLE ANALYSIS AND ASSESSMENT IN CIVIL ENGINEERING: TOWARDS AN INTEGRATED VISION, 2019, : 2659 - 2664
  • [50] OPEN-TYPE CRYOSTAT FOR X-RAY-DIFFRACTION STUDY OF SMALL SAMPLES
    BOIKO, AA
    SUVOROV, EV
    ZAVODSKAYA LABORATORIYA, 1974, 40 (03): : 283 - 284