Blood Biomarker Detection Using Integrated Microfluidics with Optical Label-Free Biosensor

被引:0
|
作者
Li, Chiung-Hsi [1 ]
Chang, Chen-Yuan [1 ]
Chen, Yan-Ru [1 ]
Huang, Cheng-Sheng [1 ]
机构
[1] Natl Yang Ming Chiao Tung Univ, Dept Mech Engn, Hsinchu 30010, Taiwan
关键词
optical biosensor; guided-mode resonance; lab-on-a-chip; optofluidics; PLASMA SEPARATION; CHIP; MICROCHIP; DEVICES; CELLS;
D O I
10.3390/s24206756
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this study, we developed an optofluidic chip consisting of a guided-mode resonance (GMR) sensor incorporated into a microfluidic chip to achieve simultaneous blood plasma separation and label-free albumin detection. A sedimentation chamber is integrated into the microfluidic chip to achieve plasma separation through differences in density. After a blood sample is loaded into the optofluidic chip in two stages with controlled flow rates, the blood cells are kept in the sedimentation chamber, enabling only the plasma to reach the GMR sensor for albumin detection. This GMR sensor, fabricated using plastic replica molding, achieved a bulk sensitivity of 175.66 nm/RIU. With surface-bound antibodies, the GMR sensor exhibited a limit of detection of 0.16 mu g/mL for recombinant albumin in buffer solution. Overall, our findings demonstrate the potential of our integrated chip for use in clinical samples for biomarker detection in point-of-care applications.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Label-Free Biosensor
    Zhang, Pengfei
    Wang, Rui
    BIOSENSORS-BASEL, 2023, 13 (05):
  • [22] Polymer dual ring resonators for label-free optical biosensing using microfluidics
    Salleh, Muhammad H. M.
    Glidle, Andrew
    Sorel, Marc
    Reboud, Julien
    Cooper, Jonathan M.
    CHEMICAL COMMUNICATIONS, 2013, 49 (30) : 3095 - 3097
  • [23] Label-free detection of exosomes using a surface plasmon resonance biosensor
    Abu Ali Ibn Sina
    Ramanathan Vaidyanathan
    Alain Wuethrich
    Laura G. Carrascosa
    Matt Trau
    Analytical and Bioanalytical Chemistry, 2019, 411 : 1311 - 1318
  • [24] Label-free detection of cytokines using optical microcavities
    Armani, Andrea M.
    Fraser, Scott E.
    SINGLE MOLECULE SPECTROSCOPY AND IMAGING, 2008, 6862
  • [25] Label-free detection of nosocomial bacteria using a nanophotonic interferometric biosensor
    Maldonado, Jesus
    Estevez, M. -Carmen
    Fernandez-Gavela, Adrian
    Jose Gonzalez-Lopez, Juan
    Belen Gonzalez-Guerrero, Ana
    Lechuga, Laura M.
    ANALYST, 2020, 145 (02) : 497 - 506
  • [26] Label-free detection of exosomes using a surface plasmon resonance biosensor
    Ibn Sina, Abu Ali
    Vaidyanathan, Ramanathan
    Wuethrich, Alain
    Carrascosa, Laura G.
    Trau, Matt
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2019, 411 (07) : 1311 - 1318
  • [27] A label-free nanostructured plasmonic biosensor based on Blu-ray discs with integrated microfluidics for sensitive biodetection
    Lopez-Munoz, Gerardo A.
    Estevez, M. -Carmen
    Cristina Pelaez-Gutierrez, E.
    Homs-Corbera, Antoni
    Carmen Garcia-Hernandez, M.
    Ignacio Imbaud, J.
    Lechuga, Laura M.
    BIOSENSORS & BIOELECTRONICS, 2017, 96 : 260 - 267
  • [28] Label-free tapered optical fiber plasmonic biosensor
    Liyanage, Thakshila
    Lai, Meimei
    Slaughter, Gymama
    ANALYTICA CHIMICA ACTA, 2021, 1169
  • [29] Multiplexed label-free optical biosensor for medical diagnostics
    Bottazzi, Barbara
    Fornasari, Lucia
    Frangolho, Ana
    Giudicatti, Silvia
    Mantovani, Alberto
    Marabelli, Franco
    Marchesini, Gerardo
    Pellacani, Paola
    Therisod, Rita
    Valsesia, Andrea
    JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (01)
  • [30] Integration of Capillary Ring Resonator Biosensor with PDMS Microfluidics for Label-Free Biosensing
    Farahi, Farnoosh
    White, Ian
    26TH SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE: SBEC 2010, 2010, 32 : 305 - 308