Research on angular vibration calibration devices

被引:1
|
作者
Xue, Jingfeng [1 ,2 ]
Peng, Jun [2 ]
Li, Xinliang [2 ]
Zhao, Weiqian [1 ]
机构
[1] School of Optoelectronics, Beijing Institute of Technology, Beijing,100081, China
[2] Optic Fiber Sensors Center, Beijing Changcheng Institute of Measurement and Metrology, Beijing,100095, China
关键词
Inertial navigation systems - Interferometers - Bearings (machine parts) - Laser interferometry;
D O I
10.7527/S1000-6893.2014.0110
中图分类号
学科分类号
摘要
In order to improve the evaluating capabilities of dynamic properties for inertial sensors, by loading standard sine wave angular vibration and measuring precisely the angular vibration process, the angular vibration sensor is calibrated. Based on the different frequency specialties, the calibration device is divided to high-frequency and low-frequency two parts. In low-frequency calibration device, an air bearing and brushless direct current torque motor is used. In high-frequency calibration device, a precise air bearing of light hollow cuptype and electromagnetic ariving structure of frame type is designed. Using precise angular encoder measure low-frequency angular vibration and using diffraction grating heterodyne laser interferometer measure high-frequency angular vibration, respectively. A primary angular vibration calibration device with frequency range of 0.25 Hz to 550 Hz, angular acceleration range of 0.1 rad/s2 to 1 760 rad/s2 and angular velocity wave-distortion below 2% is successfully developed. Compared with the angular vibration calibration device in the Physikalisch-Technische Bundesanstalt, this equipment could provide more load bearing, and could be widely used in dynamic properties tests and evaluating of inertial sensors. ©, 2015, AAAS Press of Chinese Society of Aeronautics and Astronautics. All right reserved.
引用
收藏
页码:962 / 969
相关论文
共 50 条
  • [41] Nonlinear vibration of micromachined angular vibration resonator in low pressure
    Chen, Zhiyong
    Zhou, Bin
    Zhang, Rong
    Yi Qi Yi Biao Xue Bao/Chinese Journal of Scientific Instrument, 2010, 31 (11): : 2481 - 2487
  • [42] PREDICTION OF ANGULAR VIBRATION OF AIRCRAFT STRUCTURES
    LEE, J
    WHALEY, PW
    JOURNAL OF SOUND AND VIBRATION, 1976, 49 (04) : 541 - 549
  • [43] ANGULAR FORCES AND NORMAL VIBRATION IN NICKEL
    KHAROO, HL
    GUPTA, OP
    HEMKAR, MP
    PHYSICAL REVIEW B, 1979, 19 (06): : 2986 - 2991
  • [44] Angular vibration survey of various aircraft
    Merritt, P
    Donaldson, J
    O'Brien, D
    Coleman, K
    Pyles, G
    LASER SYSTEMS TECHNOLOGY, 2003, 5087 : 35 - 45
  • [45] DEVICES FOR CALIBRATION OF INTRACARDIAC MICROPHONES
    DIESTEL, HG
    VIETOR, KW
    ACUSTICA, 1967, 19 (06): : 340 - &
  • [46] CORRECTNESS OF CALIBRATION OF MEASURING DEVICES
    VOLODARSKII, VY
    MEASUREMENT TECHNIQUES, 1974, 17 (10) : 1493 - 1496
  • [47] EXPERIENCES WITH CRYOPUMPS IN CALIBRATION DEVICES
    GROSSE, G
    MESSER, G
    VACUUM, 1970, 20 (09) : 373 - &
  • [48] Calibration of electromagnetic tracking devices
    Kindratenko V.
    Virtual Reality, 1999, 4 (2) : 139 - 150
  • [49] A calibration system for PTU devices
    Saxholm, Sari
    Heinonen, Martti
    MEASUREMENT, 2010, 43 (10) : 1583 - 1588
  • [50] Photoelectric angular displacement sensor and its calibration
    Yu, XY
    Zhao, RL
    Li, X
    Zhao, J
    Chen, DY
    Zhao, X