Results of an intercomparison for electric field strength measurements within the German calibration service

被引:1
|
作者
Pape R. [1 ]
Karsten U. [2 ]
Lindner F.-M. [3 ]
Rittmann F. [4 ]
Von Freeden J. [5 ]
Kleine-Ostmann T. [1 ]
Schrader T. [1 ]
机构
[1] Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, Braunschweig
[2] TESEQ GmbH, Landsberger Str. 255, Berlin
[3] Kalibrierzentrum der Bundeswehr, Sielower Landstr. 66, Cottbus
[4] Steep GmbH, Lise-Meitner-Strasse 6, Ottobrunn
[5] Narda Safety Test Solutions GmbH, Sandwiesenstr. 7, Pfullingen
关键词
Electric fields - Probes - Electromagnetic waves;
D O I
10.5194/ars-15-243-2017
中图分类号
学科分类号
摘要
In this paper we discuss the results of an intercomparison for electric field strength measurements within the German Calibration Service (Deutscher Kalibrierdienst-DKD). The comparison has been carried out on the field strength value required to reach a display reading of 20Vm-1 of the field probes for frequencies between 100MHz and 18 GHz. Five laboratories joined the intercomparison including the Physikalisch-Technische Bundesanstalt (PTB), the German National Metrology Institute that keeps the primary standard for electric field strength. As measurement artefacts both a small 1-Axis probe usually used as transfer sensor at PTB and a larger 3-Axis commercial field probe have been used. While the results agree well for the small field probe and when the larger commercial 3-Axis field probe is oriented in the direction of the magnetic field, larger deviations occur, when the larger 3-Axis field probe is oriented into the direction of the Poynting vector of the calibration field. © Author(s) 2017.
引用
收藏
页码:243 / 248
页数:5
相关论文
共 50 条
  • [31] EARLY RESULTS FROM ISEE-1 ELECTRIC-FIELD MEASUREMENTS
    HEPPNER, JP
    MAYNARD, NC
    AGGSON, TL
    SPACE SCIENCE REVIEWS, 1978, 22 (06) : 777 - 789
  • [32] Electric field strength and extrapolation in radiated emission measurements at an open area test site
    Kang, TW
    Kim, HT
    IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 2001, 148 (06) : 246 - 252
  • [33] Electrooptical measurements of the electric field strength in water with near-electrode conductive layers
    S. M. Korobeynikov
    A. V. Melekhov
    V. G. Posukh
    Doklady Physics, 2010, 55 : 391 - 393
  • [34] Measurements of benzene and toluene in ambient air using proton-transfer-reaction mass spectrometry: calibration, humidity dependence, and field intercomparison
    Warneke, C
    van der Veen, C
    Luxembourg, S
    de Gouw, JA
    Kok, A
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2001, 207 (03) : 167 - 182
  • [35] Levels of electric field strength within the immediate vicinity of FM radio stations in Accra, Ghana
    Azah, C. K.
    Amoako, J. K.
    Fletcher, J. J.
    RADIATION PROTECTION DOSIMETRY, 2013, 156 (04) : 395 - 400
  • [36] ELECTRIC FIELD WITHIN LIGHTNING PROTECTION VOLUME IN PRESENCE OF A DOWNWARD LEADER: SIMULATION RESULTS
    Shostak, Volodymyr
    Petrenko, Tetyana
    Janischewsky, Wasyl
    Rachide, Farhad
    2010 30TH INTERNATIONAL CONFERENCE ON LIGHTNING PROTECTION (ICLP), 2010,
  • [37] Key comparisons CCEM.RF-K-20 in the area of electric field strength measurements
    V. A. Tishchenko
    V. I. Tokatly
    V. I. Luk’yanov
    Measurement Techniques, 2007, 50 : 1014 - 1018
  • [38] Key comparisons CCEM.RF-K-20 in the area of electric field strength measurements
    Tishchenko, V. A.
    Tokatly, V. I.
    Luk'yanov, V. I.
    MEASUREMENT TECHNIQUES, 2007, 50 (09) : 1014 - 1018
  • [39] DETERMINATION OF THE POSITION OF AN UNDERGROUND PIPELINE USING THE RESULTS OF MEASUREMENTS OF ITS MAGNETIC FIELD STRENGTH
    Maksimenko, V. G.
    Naryshkin, V. I.
    Gatilov, A. V.
    MEASUREMENT TECHNIQUES, 2009, 52 (11) : 1200 - 1206
  • [40] Determination of the position of an underground pipeline using the results of measurements of its magnetic field strength
    V. G. Maksimenko
    V. I. Naryshkin
    A. V. Gatilov
    Measurement Techniques, 2009, 52 : 1200 - 1206