Three-dimensional vector velocity measurement in optical doppler tomography

被引:0
|
作者
Meng, Jie [1 ]
Ding, Zhihua [1 ]
Zhu, Ying [1 ]
机构
[1] State Key Laboratory of Modern Optical Instrumentation, Zhejiang University, Hangzhou, Zhejiang 310027, China
来源
Guangxue Xuebao/Acta Optica Sinica | 2009年 / 29卷 / 11期
关键词
Flow velocity - Vectors - Medical imaging - Plates (structural components) - Velocity measurement - Tomography - Velocity distribution;
D O I
10.3788/AOS20092911.3168
中图分类号
学科分类号
摘要
Optical Doppler tomography (ODT) is a high-resolved and non-invasive biomedical imaging technique, which can obtain tissue structure and blood flow velocity at the same time. A novel phase-resolved ODT method to measure three-dimensional vector velocity is presented. After inserting the designed narrowband phase plate between the collimator and focusing lens of the sample arm, three different phase retardations are formed. By calculating the Doppler shift and the Doppler bandwidth relative to the three phase retardations, three-dimensional vector velocity distribution of the solution inside the capillary is obtained. Several experiments of polystyrene solution with different angle setup and variational flow velocity are conducted, and the results demonstrate the feasibility of this method to measure three-dimensional vector velocity.
引用
收藏
页码:3168 / 3172
相关论文
共 50 条
  • [31] Three-dimensional vector microwave tomography: theory and computational experiments
    Bulyshev, AE
    Souvorov, AE
    Semenov, SY
    Posukh, VG
    Sizov, YE
    INVERSE PROBLEMS, 2004, 20 (04) : 1239 - 1259
  • [32] Quantitative measurement of blood velocity in zebrafish with optical vector field tomography
    Fieramonti, Luca
    Foglia, Efrem A.
    Malavasi, Stefano
    D'Andrea, Cosimo
    Valentini, Gianluca
    Cotelli, Franco
    Bassi, Andrea
    JOURNAL OF BIOPHOTONICS, 2015, 8 (1-2) : 52 - 59
  • [33] Absolute velocity measurement using three-beam spectral-domain Doppler optical coherence tomography
    P. Sharma
    Y. Verma
    S. Kumar
    P. K. Gupta
    Applied Physics B, 2015, 120 : 539 - 543
  • [34] Absolute velocity measurement using three-beam spectral-domain Doppler optical coherence tomography
    Sharma, P.
    Verma, Y.
    Kumar, S.
    Gupta, P. K.
    APPLIED PHYSICS B-LASERS AND OPTICS, 2015, 120 (03): : 539 - 543
  • [35] Measurement and imaging of a velocity vector field based on a three transducers Doppler system
    Bruni, G
    Calzolai, M
    Capineri, L
    Fort, A
    Masotti, L
    Rocchi, S
    Scabia, M
    ACOUSTICAL IMAGING, VOL 22, 1996, 22 : 431 - 437
  • [36] Three-dimensional optical coherence tomography of the murine embryo
    Luo, W
    Ralston, T
    Marks, DL
    Boppart, SA
    FASEB JOURNAL, 2005, 19 (04): : A798 - A798
  • [37] Three-dimensional optical tomography of the newborn infant brain
    Hebden, JC
    2002 IEEE/LEOS ANNUAL MEETING CONFERENCE PROCEEDINGS, VOLS 1 AND 2, 2002, : 361 - 362
  • [38] Three-dimensional optical tomography of hemodynamics in the human head
    Bluestone, AY
    Abdoulaev, G
    Schmitz, CH
    Barbour, RL
    Hielscher, AH
    OPTICS EXPRESS, 2001, 9 (06): : 272 - 286
  • [39] Three-dimensional optical coherence tomography for biomedical diagnosis
    Chan, KP
    2005 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, 2005, : 42 - 43
  • [40] Three-dimensional optical polarization tomography of single molecules
    Prummer, M
    Sick, B
    Hecht, B
    Wild, UP
    JOURNAL OF CHEMICAL PHYSICS, 2003, 118 (21): : 9824 - 9829