Development and validation of 3D printed anthropomorphic head phantom with eccentric holes for medical LINAC quality assurance testing in stereotactic radiosurgery

被引:0
|
作者
Nabhan, A. K. Shahi [1 ]
Saxena, Kritik [1 ]
Puzhakkal, Niyas [2 ]
Mathew, Jose [1 ]
Lawrence, K. Deepak [1 ]
机构
[1] Natl Inst Technol, Dept Mech Engn, Calicut, India
[2] MVR Canc Ctr & Res Inst, Dept Radiat Oncol, Calicut, India
关键词
Stereotactic radiosurgery; Linear accelerator; Anthropomorphic phantom; 3D printing; Treatment planning system; Patient specific quality assurance; WATER EQUIVALENCE; DOSIMETER;
D O I
10.1016/j.medengphy.2024.104217
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Stereotactic Radiosurgery (SRS) for brain tumors using Medical Linear Accelerator (LINAC) demands high precision and accuracy. A specific Quality Assurance (QA) is essential for every patient undergoing SRS to protect nearby non-cancerous cells by ensuring that the X-ray beams are targeted according to tumor position. In this work, a water-filled generic anthropomorphic head phantom consisting of two removable parts with eccentric holes was developed using Additive Manufacturing (AM) process for performing QA in SRS. In the patient specific QA, the planned radiation dose using Treatment Planning System (TPS) was compared with the dose measured in the phantom. Also, the energy consistency of radiation beams was tested at 200 MU for different energy beams at the central and eccentric holes of the phantom using an ionization chamber. Experimentally examined results show that planned doses in TPS are reaching the target within a 5% deviation. The ratio of the dose delivered in the eccentric hole to the dose delivered to the central hole shows variations of less than 2% for the energy consistency test. The designed, low-cost water-filled anthropomorphic phantom is observed to improve positioning verification and accurate dosimetry of patient-specific QA in SRS treatment.
引用
收藏
页数:11
相关论文
共 46 条
  • [21] Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy
    Noriyuki Kadoya
    Kota Abe
    Hikaru Nemoto
    Kiyokazu Sato
    Yoshiro Ieko
    Kengo Ito
    Suguru Dobashi
    Ken Takeda
    Keiichi Jingu
    Radiological Physics and Technology, 2019, 12 : 351 - 356
  • [22] Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy
    Kadoya, Noriyuki
    Abe, Kota
    Nemoto, Hikaru
    Sato, Kiyokazu
    Ieko, Yoshiro
    Ito, Kengo
    Dobashi, Suguru
    Takeda, Ken
    Jingu, Keiichi
    RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) : 351 - 356
  • [23] Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy
    Yea, Ji Woon
    Park, Jae Won
    Kim, Sung Kyu
    Kim, Dong Youn
    Kim, Jae Gu
    Seo, Chan Young
    Jeong, Won Hyo
    Jeong, Man Youl
    Oh, Se An
    PLOS ONE, 2017, 12 (07):
  • [24] Quality assurance for a six degrees-of-freedom table using a 3D printed phantom
    Woods, Kyle
    Ayan, Ahmet S.
    Woollard, Jeffrey
    Gupta, Nilendu
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (01): : 115 - 124
  • [25] 3D Printed Cylindrical Texture Inserts for a Commonly Used CT Phantom for Radiomics Quality Assurance
    ul Hassan, M. Shafiq
    Zhang, G.
    Latifi, K.
    Ullah, G.
    Hunt, D.
    Gillies, R.
    Moros, E.
    MEDICAL PHYSICS, 2017, 44 (06) : 3217 - 3217
  • [26] A Novel 3D Printed Patient-Specific Phantom for Spine SBRT Quality Assurance: Comparison of 3D Printing Techniques
    Lee, S.
    Kim, M.
    Lee, M.
    Suh, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3631 - 3631
  • [27] Quality Assurance for Multi-Focal Single Isocenter Stereotactic Radiosurgery Through 3D Gel Dosimetry
    Saenz, D.
    Papanikolaou, K.
    Rasmussen, K.
    Kirby, N.
    Stathakis, S.
    Pappas, E.
    Li, Y.
    MEDICAL PHYSICS, 2018, 45 (06) : E607 - E607
  • [28] Optimisation of cone beam CT radiotherapy imaging protocols using a novel 3D printed head and neck anthropomorphic phantom
    Alzahrani, Meshal
    O’Hara, Christopher
    Bird, David
    Baldwin, Jack P C
    Naisbit, Mitchell
    Teh, Irvin
    Broadbent, David A
    Al-Qaisieh, Bashar
    Johnstone, Emily
    Speight, Richard
    Physics in Medicine and Biology, 2024, 69 (21):
  • [29] Quality Assurance of Small Animal Irradiation: Validation of a 3D-Printed Phantom for "Quasi In-Vivo" Dosimetry
    Koutsouvelis, N.
    Tournier, B. B.
    Garibotto, V.
    Jaccard, M.
    Dubouloz, A.
    Nouet, P.
    Rouzaud, M.
    Zilli, T.
    Dipasquale, G.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E476 - E476
  • [30] A transit portal dosimetry method for respiratory gating quality assurance with a dynamic 3D printed tumor phantom
    Tan, Hong Qi
    Koh, Calvin Wei Yang
    Tan, Lloyd Kuan Rui
    Lew, Kah Seng
    Chua, Clifford Ghee Ann
    Ang, Khong Wei
    Lee, James Cheow Lei
    Park, Sung Yong
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (05):