Single-photon elimination in liquid scintillation counting with pulse shape discrimination and coincidence

被引:0
|
作者
Zhang, Pengpeng [1 ]
Li, Deyuan [1 ]
Li, Hui [1 ]
Zhang, Hui [1 ,2 ]
Zhang, Xuanru [1 ]
Ma, Yan [1 ,2 ]
Wang, Linjing [1 ]
Li, Hua [1 ,2 ]
Yan, Xuewen [1 ,2 ]
机构
[1] China Inst Radiat Protect, Nucl & Radiat Frontier Technol Res Ctr, Taiyuan 030006, Peoples R China
[2] Shanxi Prov Key Lab Translat Nucl Med & Precis Pro, Taiyuan 030006, Peoples R China
关键词
Liquid scintillation counting; Single-photon events; Fall time; Detection efficiency; Single-photon rejection ratio; CHEMI-LUMINESCENCE; FE-55;
D O I
10.1016/j.radmeas.2024.107299
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Liquid scintillation counting is widely used in the rapid measurement of beta activity in environmental and biological samples. However, the single-photons generated by chemiluminescence and photoluminescence in liquid scintillation cocktails seriously affect the measurement accuracy of low-energy beta activity. A novel method based on the combination of the signal characteristic analysis and selective gate to eliminate the single-photon signal was developed. A preprocessing circuit made of a fast response time photomultiplier tube (PMT, Hamamatsu R9420), two charge-sensitive preamplifiers (CSP), two comparators, an analog switch and delay-line devices were designed and developed to verify the feasibility and effectiveness. The output signals from the last dynode were characterized in the pulse time and were used to discriminate the beta signals from the single-photon ones. The beta signals were "tagged" through pulse width detection, pulse width-amplitude transform and pulse-height discrimination with the first comparator, the first CSP and the second comparator. The "tagged" beta signal were applied to control the analog switch. The anode signals were specially delayed and then selected by the analog switch to achieve the single-photon signal elimination. Liquid scintillation cocktails containing C-14 or NaOH used as beta or single-photon sources were provided to verify the feasibility of the principle. The results showed that the typical fall time of the single-photon and beta signal was 16.05 ns and 43.17 ns. The single-photon rejection ratio is 2.76 x 10(-3) +/- 3.89 x 10(-5), and the detection efficiency is up to 93.02%+/- 0.59%.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Quantum optics - Beyond single-photon counting
    Sergienko, Alexander V.
    NATURE PHOTONICS, 2008, 2 (05) : 268 - 269
  • [42] Mid-infrared single-photon counting
    Temporao, G
    Tanzilli, S
    Zbinden, H
    Gisin, N
    Aellen, T
    Giovannini, M
    Faist, J
    OPTICS LETTERS, 2006, 31 (08) : 1094 - 1096
  • [43] Linear arrays of single-photon detectors for photon counting and timing
    Guerrieri, F.
    Tisa, S.
    Tosi, A.
    Bellisai, S.
    Markovic, B.
    Zappa, F.
    SENSORS, CAMERAS, AND SYSTEMS FOR INDUSTRIAL, SCIENTIFIC, AND CONSUMER APPLICATIONS XII, 2011, 7875
  • [44] Photon counting probabilities of the output field for a single-photon input
    Dabrowska, Anita magdalena
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA B-OPTICAL PHYSICS, 2023, 40 (05) : 1299 - 1310
  • [45] Infrared photon counting with superconducting nanowire single-photon detectors
    Hadfield, Robert H.
    Natarajan, Chandra M.
    Tanner, Michael G.
    2013 IEEE PHOTONICS CONFERENCE (IPC), 2013, : 368 - 369
  • [46] Photon-counting distribution for arrays of single-photon detectors
    Jonsson, Mattias
    Bjork, Gunnar
    PHYSICAL REVIEW A, 2020, 101 (01)
  • [47] Standardization of Ga-68 by coincidence measurements, liquid scintillation counting and 4πγ counting
    Roteta, Miguel
    Peyres, Virginia
    Rodriguez Barquero, Leonor
    Garcia-Torano, Eduardo
    Arenillas, Pablo
    Balpardo, Christian
    Rodrigues, Dario
    Llovera, Roberto
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (09) : 2006 - 2011
  • [48] Avalanche discrimination and high-speed counting in periodically gated single-photon avalanche diodes
    Restelli, Alessandro
    Bienfang, Joshua C.
    ADVANCED PHOTON COUNTING TECHNIQUES VI, 2012, 8375
  • [49] Liquid scintillation counting by digital pulse height analysis
    GarciaTorano, E
    LosArcos, JM
    LIQUID SCINTILLATION SPECTROMETRY 1994, 1996, : 25 - 29
  • [50] STANDARDIZATION OF C-14 BY LIQUID-SCINTILLATION COINCIDENCE COUNTING
    WU, XZ
    SONG, L
    WANG, ZY
    YANG, YD
    WANG, C
    APPLIED RADIATION AND ISOTOPES, 1987, 38 (10) : 891 - 897