Dosimetric Comparison between the HyperArc and Conventional VMAT in Cervical Spine Stereotactic Radiosurgery

被引:0
|
作者
Park, Jeehoon [1 ]
Park, Byungdo [1 ]
Kim, Jeongho [1 ]
机构
[1] Sungkyunkwan Univ, Samsung Changwon Hosp, Sch Med, Dept Radiat Oncol, Chang Won 630522, South Korea
基金
新加坡国家研究基金会;
关键词
HyperArc; cervical spine; stereotactic radiosurgery; stereotactic radiosurgery normal tissue objective; BODY RADIATION-THERAPY; RADIOTHERAPY; IMMOBILIZATION; METASTASES; MOTION; TARGET; HEAD;
D O I
10.3390/jcm13185497
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This research aims to evaluate the usability of the HyperArc (HA) technique in stereotactic radiosurgery for cervical spine metastasis by comparing the dosimetry of the target and organs at risk, specifically the spinal cord, between HA and VMAT and conventional volumetric modulated arc therapy (VMAT). Methods: A RANDO (R) phantom and QFix EncompassTM and support system were used to simulate three target types (A, B, and C) based on RTOG0631 guidelines. Treatment plans included one VMAT and two HyperArc techniques with different SRS NTO values (100 and 250). Dosimetric parameters such as conformity index (CI), homogeneity index (HI), R50, and spinal cord sparing were analyzed. Gamma analysis was performed using portal dosimetry to validate the dose delivery accuracy. Results: HyperArc plans demonstrated higher conformity, sharper dose fall-off, and improved quality assurance (QA) results compared to VMAT plans. HA with SRS NTO 250 showed even better results in terms of conformity, dose fall-off, and spinal cord dose reduction (V10 and Dmax) compared to HA with SRS NTO 100. Although the mean gamma passing rates were slightly lower, all plans achieved rates above 95%. Conclusion: The findings suggest that HA provides superior dosimetric benefits over VMAT and could be effectively utilized for cervical spine radiation therapy.
引用
收藏
页数:11
相关论文
共 50 条
  • [31] Dosimetric comparison between RapidArc and HyperArc techniques in salvage stereotactic body radiation therapy for recurrent nasopharyngeal carcinoma
    Ho, Hsiu-Wen
    Lee, Steve P.
    Lin, Hisu-Man
    Chen, Hsiao-Yun
    Huang, Chun-Chiao
    Wang, Shih-Chang
    Yang, Ching-Chieh
    Lin, Yu-Wei
    RADIATION ONCOLOGY, 2020, 15 (01)
  • [32] Dosimetric comparison between RapidArc and HyperArc techniques in salvage stereotactic body radiation therapy for recurrent nasopharyngeal carcinoma
    Hsiu-Wen Ho
    Steve P. Lee
    Hisu-Man Lin
    Hsiao-Yun Chen
    Chun-Chiao Huang
    Shih-Chang Wang
    Ching-Chieh Yang
    Yu-Wei Lin
    Radiation Oncology, 15
  • [33] Spine Stereotactic Radiosurgery Systems : Comparison Between Brainlab Elements and Eclipse RapidArc
    Deshazer, G.
    Galhardo, E.
    Bimali, M.
    Morrill, S.
    Penagaricano, J.
    Narayanasamy, G.
    MEDICAL PHYSICS, 2019, 46 (06) : E329 - E329
  • [34] Interinstitutional Dosimetric Comparison Between Stereotactic Radiosurgery and Volumetric Modulated Arc Therapy Radiosurgery for Difficult Intracranial Cases
    Thomas, E. M.
    Chao, S. T.
    Markert, J. M.
    Neyman, G.
    Popple, R. A.
    Barnett, G. H.
    Fiveash, J. B.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 96 (02): : E693 - E693
  • [35] Dosimetric investigation of rotational setup errors for spine stereotactic radiosurgery: a phantom study
    Sasaki, M.
    Nakamura, M.
    Mukumoto, N.
    Nakata, M.
    Higashimura, K.
    Hiraoka, M.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S257 - S258
  • [36] Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of meningioma
    David Kaul
    Harun Badakhshi
    Thierry Gevaert
    Diana Pasemann
    Volker Budach
    Constantin Tulaesca
    Arne Gruen
    Vikas Prasad
    Marc Levivier
    Markus Kufeld
    Acta Neurochirurgica, 2015, 157 : 559 - 564
  • [37] Dosimetric comparison of different treatment modalities for stereotactic radiosurgery of meningioma
    Kaul, D.
    Gevaert, T.
    Pasemann, D.
    Budach, V
    Tuleasca, C.
    Levivier, M.
    Kufeld, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 : S107 - S107
  • [38] IMAGE-GUIDED STEREOTACTIC SPINE RADIOSURGERY ON A CONVENTIONAL LINEAR ACCELERATOR
    Wang, Jia-Zhu
    Rice, Roger
    Mundt, Arno
    Sandhu, Ajay
    Murphy, Kevin
    MEDICAL DOSIMETRY, 2010, 35 (01) : 53 - 62
  • [39] A Dosimetric Comparison of Proton and VMAT for Pediatric Ewing Sarcoma of Pelvis and Spine
    Meniai, F.
    Vigan, M.
    Goudjil, F.
    Zefkili, S.
    Helfre, S.
    PEDIATRIC BLOOD & CANCER, 2019, 66 : S490 - S490
  • [40] A dosimetric comparison of Proton and VMAT for Pediatric Ewing sarcoma of pelvis and spine
    Meniai-Merzouki, F.
    Vigan, M.
    Goudjil, F.
    Helfre, S.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S870 - S870