Homogeneous B0 coil design method for open-access ultra-low field magnetic resonance imaging: A simulation study

被引:0
|
作者
Karasawa, Tomohiro [1 ]
Saikawa, Jiro [1 ]
Munaka, Tatsuya [1 ]
Kobayashi, Tetsuo [2 ]
机构
[1] Shimadzu Co Ltd, Technol Res Lab, 3-9-4 Hikaridai, Seika 6190237, Japan
[2] Kyoto Univ, Off Inst Advancement & Commun, Sakyo Ku, Kyoto 6068501, Japan
关键词
Ultra-low-field MRI; Electromagnetic coil; Multi-objective optimization;
D O I
10.1016/j.mri.2024.07.006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A multimodal brain function measurement system integrating functional magnetic resonance imaging (fMRI) and magnetoencephalography (MEG) is expected to be a tool that will provide new insights into neuroscience. To integrate fMRI and MEG, an ultra-low-field MRI (ULF-MRI) scanner that can generate a static magnetic field (B0) with an electromagnetic coil and turn off the B0 during MEG measurements is desirable. While electromagnetic B0 coil has the above advantages, it also has a trade-off between size and the broadness of the magnetic field homogeneity. In this study, we proposed a method for designing a B0 multi-stage circular coil arrangement that determines the number of coils required to maximize magnetic field homogeneity and minimize the total wiring length of the coils. The optimized multi-stage coil arrangement had an external shape of 600 mm in diameter and a maximum height of 600 mm, with an aperture of 600 mm in diameter and 300 mm in height. The magnetic field homogeneity was <100 ppm over a 210 mm diameter spherical volume (DSV). Compared to a previous two coil pairs arrangement with the same magnetic field homogeneity, the diameter was 1/1.9 times smaller, indicating that the newly designed B0 coil arrangement realized a smaller size and wider magnetic field homogeneity.
引用
下载
收藏
页码:128 / 135
页数:8
相关论文
共 50 条
  • [21] 3D-Spatial encoding with permanent magnets for ultra-low field magnetic resonance imaging
    Michael W. Vogel
    Ruben Pellicer Guridi
    Jiasheng Su
    Viktor Vegh
    David C. Reutens
    Scientific Reports, 9
  • [22] Adaptive suppression of power line interference in ultra-low field magnetic resonance imaging in an unshielded environment
    Huang, Xiaolei
    Dong, Hui
    Qiu, Yang
    Li, Bo
    Tao, Quan
    Zhang, Yi
    Krause, Hans-Joachim
    Offenhaeusser, Andreas
    Xie, Xiaoming
    JOURNAL OF MAGNETIC RESONANCE, 2018, 286 : 52 - 59
  • [23] Ultra-low field magnetic resonance imaging detection with gradient tensor compensation in urban unshielded environment
    Dong, Hui
    Qiu, Longqing
    Shi, Wen
    Chang, Baolin
    Qiu, Yang
    Xu, Lu
    Liu, Chao
    Zhang, Yi
    Krause, Hans-Joachim
    Offenhaeusser, Andreas
    Xie, Xiaoming
    APPLIED PHYSICS LETTERS, 2013, 102 (10)
  • [24] Signal-to-Noise Ratio Enhancement Method for Ultra-Low Field Magnetic Resonance Imaging Based on Low-Frequency Surface Metamaterials
    Kong, Xiaohan
    Zhang, Yana
    Wu, Jiamin
    He, Yucheng
    Xu, Zheng
    Diangong Jishu Xuebao/Transactions of China Electrotechnical Society, 2024, 39 (13): : 3917 - 3927
  • [25] A Comparative Study of Birdcage RF Coil Configurations for Ultra-High Field Magnetic Resonance Imaging
    Seo, Jeung-Hoon
    Han, Yeji
    Chung, Jun-Young
    SENSORS, 2022, 22 (05)
  • [26] Design and Testing of a Multi-Channel 1H/3He Double-Nuclear Transmit Coil Array for Magnetic Resonance Imaging at an Ultra-Low Field
    Xu, Jinzhang
    Dou, Yan
    Wang, Yi
    Chen, Qiaoyan
    Wang, Lulu
    Li, Nan
    Luo, Chao
    Du, Feng
    Zhang, Xiaoliang
    Li, Ye
    JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2019, 9 (01) : 119 - 125
  • [27] SQUIDs vs. Induction Coils for Ultra-Low Field Nuclear Magnetic Resonance: Experimental and Simulation Comparison
    Matlashov, Andrei N.
    Schultz, Larry J.
    Espy, Michelle A.
    Kraus, Robert H.
    Savukov, Igor M.
    Volegov, Petr L.
    Wurden, Caroline J.
    IEEE TRANSACTIONS ON APPLIED SUPERCONDUCTIVITY, 2011, 21 (03) : 465 - 468
  • [28] Multi-channel SQUID-based Ultra-Low Field Magnetic Resonance Imaging in Unshielded Environment
    Matlashov, Andrei
    Magnelind, Per
    Newman, Shaun
    Sandin, Henrik
    Urbaitis, Algis
    Volegov, Petr
    Espy, Michelle
    2015 15TH INTERNATIONAL SUPERCONDUCTIVE ELECTRONICS CONFERENCE (ISEC), 2015,
  • [29] Design and development of a planar B0-coil for patient respiratory motion correction in magnetic resonance imaging
    Awan, Shakil A.
    Mcginley, John V. M.
    Dickinson, Robert J.
    Angeli, Stelios I.
    Young, Ian R.
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2012, 41B (04) : 130 - 138
  • [30] Design and Construction of a Prototype High-Power B0 Insert Coil for Field-Cycled Imaging in Superconducting MRI Systems
    Alford, Jamu K.
    Scholl, Timothy J.
    Handler, William B.
    Chronik, Blaine A.
    CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2009, 35B (01) : 1 - 10