Stem to prevent periprosthetic fracture after notching in total knee arthroplasty

被引:0
|
作者
Wan, Qian [1 ]
Zhang, Aobo [1 ]
Liu, Yang [1 ]
Chen, Hao [1 ]
Zhao, Xue [2 ]
Han, Qing [1 ]
Wang, Jincheng [1 ]
机构
[1] Second Hosp Jilin Univ, Dept Orthoped, Changchun 130000, Peoples R China
[2] First Hosp Jilin Univ, Dept Endocrinol & Metab, Changchun, Peoples R China
基金
中国国家自然科学基金;
关键词
aseptic loosening; finite element analysis; micromotion; notching; periprosthetic fracture; total knee arthroplasty; SUPRACONDYLAR FRACTURE; FEMUR; RISK; FAILURE; DESIGN; INTACT; LENGTH; HIP;
D O I
10.1002/cnm.3842
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Improper osteotomy during total knee arthroplasty (TKA) can lead to anterior femoral notching, which increases the risk of periprosthetic fractures due to stress concentration. One potential solution is the addition of an intramedullary stem to the femoral component. However, the optimal stem length remains unclear. In this study, we aimed to determine the optimal stem length using finite element models. Finite element models of femurs were developed with unstemmed prostheses and prostheses with stem lengths of 50, 75, and 100 mm. Under squat loading conditions, the von Mises stress at the notch and stress distribution on four transversal sections of the femur were analyzed. Additionally, micromotion of the prosthesis-bone interface was evaluated to assess initial stability. The unstemmed prosthesis exhibited a von Mises stress of 191.8 MPa at the notch, which decreased to 43.1, 8.8, and 23.5 MPa for stem lengths of 50, 75, and 100 mm, respectively. The stress reduction on four selected femoral transversal sections compared with the unstemmed prosthesis was 40.0%, 84.4%, and 67.1% for stem lengths of 50, 75, and 100 mm, respectively. Micromotion analysis showed a maximum of 118.8 mu m for the unstemmed prosthesis, which decreased significantly with the application of stems, particularly at the anterior flange. Intramedullary stems effectively reduced stress concentration at the femoral notch. The 50-mm stem length provided the optimal combination of reduced notch stress, minimized stress-shielding effect, and decreased micromotion at the anterior flange. This study used finite element models to investigate the effects of intramedullary stem on reducing risk of periprosthetic fractures after anterior femoral notching in total knee arthroplasty. Results showed that stemmed prostheses with a 50-mm stem length reduced stress concentration and micromotion at the anterior flange, making it the optimal choice.image
引用
收藏
页数:13
相关论文
共 50 条
  • [21] Periprosthetic fractures of the femur after total knee arthroplasty
    McGraw P.
    Kumar A.
    Journal of Orthopaedics and Traumatology, 2010, 11 (3) : 135 - 141
  • [22] Periprosthetic fractures after total knee joint arthroplasty
    Mittlmeier, T
    Stöckle, U
    Perka, C
    Schaser, KD
    UNFALLCHIRURG, 2005, 108 (06): : 481 - 495
  • [23] Treatment of periprosthetic fractures after total knee arthroplasty
    Boulton, C. L.
    Rodriguez, E. K.
    CURRENT ORTHOPAEDIC PRACTICE, 2009, 20 (01): : 58 - 64
  • [24] Periprosthetic fracture between a constrained total knee arthroplasty and a long-stem total hip arthroplasty - Treatment with a novel device
    Lachiewicz, Paul F.
    JOURNAL OF ARTHROPLASTY, 2007, 22 (03): : 449 - 452
  • [25] Management and clinical outcomes of periprosthetic fractures after total knee arthroplasty with a stem extension
    Shin, Jae Hoon
    Chang, Moon Jong
    Kang, Seung-Baik
    Chang, Chong Bum
    Kim, Dong Hwi
    Lee, Seung Ah
    Mok, Su Jung
    MEDICINE, 2019, 98 (25) : e16088
  • [26] Periprosthetic fracture of the acetabulum after total hip arthroplasty
    Peterson, CA
    Lewallen, DG
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (08): : 1206 - 1213
  • [27] Usefulness of bilateral plate fixation for periprosthetic distal femur fracture after total knee arthroplasty
    Kim, Kwangkyoun
    INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS, 2020, 68 : 43 - 47
  • [28] Administration of teriparatide treatment for a challenging case of nonunion of periprosthetic fracture after total knee arthroplasty
    Kensuke Ochi
    Katsunori Ikari
    Akari Naomi
    Shigeki Momohara
    Archives of Osteoporosis, 2013, 8
  • [29] Administration of teriparatide treatment for a challenging case of nonunion of periprosthetic fracture after total knee arthroplasty
    Ochi, Kensuke
    Ikari, Katsunori
    Naomi, Akari
    Momohara, Shigeki
    ARCHIVES OF OSTEOPOROSIS, 2013, 8 (1-2)
  • [30] Infection and periprosthetic fracture are the leading causes of failure after aseptic revision total knee arthroplasty
    Juliette A. Meyer
    Mark Zhu
    Alana Cavadino
    Brendan Coleman
    Jacob T. Munro
    Simon W. Young
    Archives of Orthopaedic and Trauma Surgery, 2021, 141 : 1373 - 1383