Monte Carlo dosimetry for BEBIG Cobalt-60 brachytherapy source with gynecological applicator in the presence of inhomogeneities

被引:0
|
作者
Moradi, Farhad [1 ]
Sani, Siti Fairus Abdul [2 ]
Mahyuddin, Azman [2 ]
Ung, Ngie Min [3 ]
Bradley, David Andrew [4 ,5 ]
Almugren, Kholoud [6 ]
Rashid, Hairul Azhar Abdul [1 ]
机构
[1] Multimedia Univ, Fac Engn, Jalan Multimedia, Cyberjaya 63100, Malaysia
[2] Univ Malaya, Dept Phys, Kuala Lumpur, Malaysia
[3] Univ Malaya, Fac Med, Clin Oncol Unit, Kuala Lumpur 50603, Malaysia
[4] Sunway Univ, Appl Phys & Radiat Technol Grp, CCDCU, PJ 46150, Malaysia
[5] Univ Surrey, Dept Phys, Guildford, England
[6] Princess Nourah Bint Abdulrahman Univ, Dept Phys, Riyadh, Saudi Arabia
来源
关键词
brachytherapy; Monte Carlo simulation; dosimetry; HDR; BEBIG; Co-60; DOSE-RATE BRACHYTHERAPY; TECHNICAL NOTE;
D O I
10.2478/pjmpe-2024-0017
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Introduction: High-dose-rate (HDR) Co-60 brachytherapy necessitates accurate dose calculations to minimize normal tissue toxicity and late malignancy risks. Conventionally, the American Association of Physicists in Medicine (AAPM) Task Group 43 (TG-43) formalism, utilizing table-based dose superposition, has been employed for dose calculations, overlooking tissue inhomogeneity effects. Material and methods: This study focuses on characterizing the high-dose-rate BEBIG Co-60 brachytherapy source, in combination with a gynecological applicator, using Monte Carlo simulations. The investigation is based on the model of the modified BEBIG Co-60 source (Co0.A86) implemented at the University of Malaya Medical Center. Dosimetric properties are evaluated according to AAPM TG-43 formalism, with validation against existing published data. Results: Our investigation presents comprehensive MC dosimetric properties of the high-dose-rate BEBIG Co-60 brachytherapy source, highlighting its accuracy in dose calculations compared to established data. The study also examined the impact of the applicator on depth dose calculations within a Krieger phantom and explored the influence of various tissue inhomogeneities on the depth dose. Conclusions: Our findings revealed that the applicator had a relatively minimal effect on the delivered dose, with only marginal differences observed. Furthermore, we investigated the depth doses along the central axis of the applicator, within a segment characterized by various tissue inhomogeneities where dose differences of up to 12% were observed, with the lowest and highest doses recorded within bone and adipose tissues, respectively. This study underscores the valuable role of MC simulations in estimating doses at locations where physical measurements are unfeasible, such as the intra-uterine tube surface, as well as in scenarios featuring tissue inhomogeneities.
引用
收藏
页码:145 / 151
页数:7
相关论文
共 50 条
  • [21] Monte Carlo dosimetry of 60Co HDR brachytherapy sources
    Papagiannis, P
    Angelopoulos, A
    Pantelis, E
    Sakelliou, L
    Karaiskos, P
    Shimizu, Y
    MEDICAL PHYSICS, 2003, 30 (04) : 712 - 721
  • [22] Monte carlo-aided dosimetry of the new Bebig IsoSeed® 103Pd interstitial brachytherapy seed
    Daskalov, GM
    Williamson, JF
    MEDICAL PHYSICS, 2001, 28 (10) : 2154 - 2161
  • [23] Monte Carlo dosimetry of a new 90Y brachytherapy source
    Wu Junxiang
    You Shihu
    Huang Jing
    Long Fengxiang
    Wang Chengkai
    Wu Zhangwen
    Hou Qing
    Gou Chengjun
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2015, 7 (05) : 397 - 406
  • [24] BRACHYTHERAPY: MONTE CARLO SOURCE, MODELLING, DOSIMETRY AND TG43
    Papagiannis, P.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S97 - S97
  • [25] Intra-cavitary brachytherapy in endometrial carcinoma: experience with cobalt-60 source
    Rullan, Jose Antonio Dominguez
    Miguelanez, Maria Teresa Munoz
    Fernandez, Rafael Colmenares
    Gallego, Miguel Camara
    Sanchez, Mercedes Martin
    Vicente, Feliciano Garcia
    Garcia, Sonsoles Sancho
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2022, 14 (03) : 227 - 232
  • [26] Monte Carlo calculated TG-60 dosimetry parameters for the β- emitter 153Sm brachytherapy source
    Sadeghi, Mahdi
    Taghdiri, Fatemeh
    Hosseini, S. Hamed
    Tenreiro, Claudio
    MEDICAL PHYSICS, 2010, 37 (10) : 5370 - 5375
  • [27] STUDIES ON THE DOSIMETRY AND BACTERICIDAL EFFECTS OF GAMMA RADIATIONS FROM A COBALT-60 SOURCE
    GOLDBLITH, SA
    PROCTOR, BE
    DAVISON, S
    LANG, DA
    KAN, B
    BATES, CJ
    KAREL, M
    RADIOLOGY, 1953, 60 (05) : 732 - 736
  • [28] Monte Carlo calculation of shielded colpostat effect on rectum received dose in high dose rate brachytherapy with Cobalt-60 sources
    Mesbahi, A.
    Haghzadeh, A.
    Naseri, A. R.
    Shirazi, A. R.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2015, 13 (02): : 165 - 171
  • [29] An analytic approach to the dosimetry of a new BEBIG Co-60 high-dose-rate brachytherapy source
    Bhola, Subhalaxmi
    Selvam, T. Palani
    Sridhar, Sahoo
    Vishwakarma, Ramkrishna S.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (03) : 129 - 137
  • [30] Comparison of Dose Distribution in Different Types of Cylindrical Applicators Using High Dose Rate Cobalt-60 Source Brachytherapy in Gynecological Cancer
    Abtahi, M.
    Gholami, S.
    Maddah-Safaei, A.
    MEDICAL PHYSICS, 2021, 48 (06)