Assessment of the accuracy and reliability of clinical thermometers for body temperature measurements at the hospital: a laboratory study

被引:1
|
作者
Shinsuphan, Nikorn [1 ,3 ]
Phanthi, Aphinan [3 ]
Leeudomwong, Theera [2 ]
Sanponpute, Tassanai [1 ]
机构
[1] Natl Inst Metrol Thailand, Med Metrol Lab, Pathum Thani 12120, Thailand
[2] King Mongkuts Univ Technol North Bangkok, Dept Ind Phys & Med Instrumentat, Bangkok 10800, Thailand
[3] Chiang Mai Univ, Maharaj Nakorn Chiang Mai Hosp, Fac Med, Maintenance Sect,Med Instrument Subsect, Chiang Mai 50200, Thailand
关键词
contact and non-contact thermometers; accuracy; body temperature; uncertainty; calibration; TYMPANIC THERMOMETER; SKIN THERMOMETER; MERCURY; PERFORMANCE;
D O I
10.1088/1361-6501/ad64f7
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study aims to assess the accuracy and reliability of two different types of thermometers: contact thermometers (CT) and non-contact thermometers (NCIRT) at Maharaj Nakorn Chiang Mai Hospital. The comparison method was carried out in a hospital laboratory setting and followed a specific calibration procedure. The mean readings, errors, and measurement uncertainties were reported and categorized according to the manufacturer's labeling models. The maximum permissible errors were defined by international standards and clinical acceptances for contact and NCIRTs, respectively. The results showed that thermometers of temperature probes (PTs) provided the highest level of accuracy. When compared to a reference standard thermometer (STD), the mean errors were only 0.04 degrees C at the calibration point of 37.0 degrees C. For digital CTs, the mean error at the calibration points of 37.0 degrees C was 0.24 degrees C. Most CT models exhibited high accuracy and reliability. However, the forehead (IRFTs) and ear (IRETs) readings significantly differed from the STD, with over half of all units not meeting the ISO 80601-2-56/ASTM E1965-98 standard requirements. These results suggest that CTs and PTs are highly accurate and reliable devices for measuring body temperature. However, both IRFTs and IRETs could be useful in certain situations, such as screening many people in a short amount of time, but their accuracy should be interpreted with caution and confirmed using a more precise method if necessary.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Body temperature measurements in the clinic; evaluation of practice in a Norwegian hospital
    Sandsund, M
    Geving, IH
    Reinertsen, RE
    Aadahl, P
    JOURNAL OF THERMAL BIOLOGY, 2004, 29 (7-8) : 877 - 880
  • [32] A reliability study of clinical tooth wear measurements
    Lobbezoo, F
    Naeije, M
    JOURNAL OF PROSTHETIC DENTISTRY, 2001, 86 (06): : 597 - 602
  • [33] Assessment of the Spatial Resolution and Accuracy of Temperature Measurements in Radiothermometry of the Breast
    Polyakov, M. V.
    Khoperskov, A. V.
    Agasieva, S. V.
    Porokhov, I. O.
    BIOMEDICAL ENGINEERING, 2023, 57 (03) : 180 - 184
  • [34] Accuracy and Reliability of Pallor for Detecting Anaemia: A Hospital-Based Diagnostic Accuracy Study
    Kalantri, Ashwini
    Karambelkar, Mandar
    Joshi, Rajnish
    Kalantri, Shriprakash
    Jajoo, Ulhas
    PLOS ONE, 2010, 5 (01):
  • [35] Clinical accuracy and agreement between tympanic and forehead body temperature measurements for screening of patients with COVID-19
    Nia, Hamid Sharif
    Chong, Pei Pei
    Huak, Chan Yiong
    Gorgulu, Ozkan
    Taghipour, Behzad
    Froelicher, Erika Sivarajan
    Sharif, Saeed Pahlevan
    Rahmatpour, Pardis
    JOURNAL OF CLINICAL NURSING, 2022, 31 (21-22) : 3272 - 3285
  • [36] System for Observing the Accuracy and Reliability of Temperature Measurements in VVÉR-440
    S. Shtants
    M. Repa
    Atomic Energy, 2002, 93 : 635 - 641
  • [38] Comparing body temperature measurements using the double sensor method within a wearable device with oral and core body temperature measurements using medical grade thermometers-a short report
    Eisenkraft, Arik
    Goldstein, Nir
    Fons, Meir
    Tabi, Michael
    Sherman, Anna Danielle
    Ben Ishay, Arik
    Merin, Roei
    Nachman, Dean
    FRONTIERS IN PHYSIOLOGY, 2023, 14
  • [39] Effect of Auricular Position on Body Temperature Measurement with Tympanic Thermometers: A Quasi Experimental Study
    Orkun, N.
    Eser, I
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2023, 26 (11) : 1595 - 1601
  • [40] EXTERNAL QUALITY CONTROL ASSESSMENT IN A HOSPITAL CLINICAL CHEMISTRY LABORATORY
    Beutler-Oberholzer, A.
    Zimmermann-Ivol, C. G.
    Hochstrasser, D. F.
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S714 - S714