Novel Matrix Normalization Technique for Accurate LC-ESI-MS/MS Detection and Quantification of Drugs and Their Metabolites in Toxicology Research and Practice

被引:1
|
作者
Rosano, Thomas G. [1 ,2 ]
Rumberger, John M. [1 ]
Scholz, Kiley L. [1 ]
Konetchy Jr, Robert M. [1 ]
Wood, Michelle [3 ]
机构
[1] Natl Toxicol Ctr, Ctr Med Sci, Albany, NY 12208 USA
[2] Albany Med Coll, Dept Lab Med, Albany, NY 12208 USA
[3] Waters Corp, Wilmslow, England
来源
CURRENT PROTOCOLS | 2023年 / 3卷 / 01期
关键词
electrospray ionization; liquid chromatography; matrix effect normalization; tandem mass spectrometry; threshold accurate calibration; TAC; LABELED INTERNAL STANDARDS; PRESSURE CHEMICAL-IONIZATION; TANDEM MASS-SPECTROMETRY; ELECTROSPRAY-IONIZATION; URINE; IMMUNOASSAY; SUPPRESSION; BENZODIAZEPINES; ENHANCEMENT; MANAGEMENT;
D O I
10.1002/cpz1.644
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Accurate identification and quantification of drugs and their metabolites (analytes) in biological matrices is an analytical foundation of clinical and forensic toxicology. For decades, liquid chromatography interfaced by electrospray ionization with tandem mass spectrometry (LC-ESI-MS/MS) has been a widely used technology for analysis in the field of toxicology, as well as in many other fields of bioscience. It is also known that ion response in LC-ESI-MS/MS analysis is influenced by coeluting biological compounds and that preanalytical sample clean-up is often insufficient in removing these interferences. As a result, a normalization technique is commonly used for assessment and compensation of matrix effects encountered in routine analysis. Internal standardization with a stable isotope analog of the analyte is the predominant normalization technique used in LC-ESI-MS/MS analysis. The technique, however, requires commercial availability or costly custom synthesis of an isotopic analog specific for each analyte. Here we describe an alternative technique for matrix normalization for use in high-volume, multianalyte testing without the need for isotope analogs. The technique involves analysis of the original sample (neat analysis) followed by analysis of a second sample aliquot (spike analysis) that has been fortified with a controlled amount of reference analyte. A calibration procedure similar to internal standardization is employed, using an ion response ratio of neat to fortified analyte. As a demonstration of the technique in multianalyte testing, we provide a detailed protocol for simultaneous detection and quantification of 102 drugs and drug metabolites in human urine. We also provide a support protocol for addition of new analytes to the multianalyte panel, allowing convenient collection of the validation data during routine testing. The matrix normalization technique and testing principles may be applicable to a wide range of analytes and biological matrices, not only those encountered in toxicology but also in other fields of bioscience. (c) 2023 Wiley Periodicals LLC.Basic Protocol: Detection and quantification of 102 toxicology analytes in urine by LC-ESI-MS/MS analysis using the threshold accurate calibration techniqueSupport Protocol: Method for addition and validation of new analytes to expand the Basic Protocol
引用
收藏
页数:39
相关论文
共 50 条
  • [31] A sensitive LC-ESI-MS/MS method for the quantification of low concentrations of buspirone in human plasma
    Joshi, Pankaj
    Rao, Shireen
    Singh, Manoj
    Vig, Nitin
    Singh, Abhay Kant
    Rajawat, Narendra Singh
    Gupta, Ankush
    ANALYTICAL METHODS, 2013, 5 (24) : 7014 - 7021
  • [32] A deoxynucleotide derivatization methodology for improving LC-ESI-MS detection
    Flarakos, J
    Xiong, W
    Glick, J
    Vouros, P
    ANALYTICAL CHEMISTRY, 2005, 77 (08) : 2373 - 2380
  • [33] Determination of carbonyl compounds in the atmosphere by DNPH derivatization and LC-ESI-MS/MS detection
    Chi, Yuguang
    Feng, Yanli
    Wen, Sheng
    Lu, Huixiong
    Yu, Zhiqiang
    Zhang, Wenbing
    Sheng, Guoying
    Fu, Jiamo
    TALANTA, 2007, 72 (02) : 539 - 545
  • [34] Simultaneous quantification of the organophosphorus pesticides dimethoate and omethoate in porcine plasma and urine by LC-ESI-MS/MS and flow-injection-ESI-MS/MS
    John, Harald
    Eddleston, Michael
    Clutton, R. Eddie
    Worek, Franz
    Thiermann, Horst
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (17-18): : 1234 - 1245
  • [35] Signal interference between drugs and metabolites in LC-ESI-MS quantitative analysis and its evaluation strategy
    Jiang, Fulin
    Liu, Jingyu
    Li, Yagang
    Lu, Zihan
    Liu, Qian
    Xing, Yunhui
    Zhu, Janshon
    Huang, Min
    Zhong, Guoping
    JOURNAL OF PHARMACEUTICAL ANALYSIS, 2024, 14 (07)
  • [36] A novel approach to high throughput in simultaneously analyzing 80 drugs of abuse in urine by using LC-ESI-MS/MS
    Wang, J. M.
    Xu, B. B.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2019, 125 : 13 - 13
  • [37] Mitigating Matrix Effects in LC-ESI-MS/MS Analysis of a Urinary Biomarker of Xylenes Exposure
    Bowman, Brett A.
    Ejzak, Elizabeth
    Reese, Christopher M.
    Blount, Benjamin C.
    Bhandari, Deepak
    JOURNAL OF ANALYTICAL TOXICOLOGY, 2023, 47 (02) : 129 - 135
  • [38] Matrix interference in LC-ESI-MS/MS analysis of metanephrines in protein precipitated plasma samples
    Rebane, Riin
    Herodes, Koit
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2020, 26 (01) : 46 - 54
  • [39] Influence of ionization source design on matrix effects during LC-ESI-MS/MS analysis
    Ghosh, Chinmoy
    Shinde, Chandrakant P.
    Chakraborty, Bhaswat S.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 893 : 193 - 200
  • [40] Identified synthesis and in-vivo study of oxidized metabolites of Rimonabant via LC-ESI-MS technique
    Tsai, Shuo-En
    Chung, Cheng-Yen
    Li, Sin-Min
    Zeng, Wei-Zheng
    Chou, Jia-Yu
    Jiang, Wen-Ping
    Huang, Guan-Jhong
    Lee, Der-Yen
    Wong, Fung Fuh
    JOURNAL OF MOLECULAR STRUCTURE, 2024, 1298