A comparative analysis of preclinical computed tomography radiomics using cone-beam and micro-computed tomography scanners

被引:0
|
作者
Brown, Kathryn H. [1 ]
Kerr, Brianna N. [1 ]
Pettigrew, Mihaela [1 ]
Connor, Kate [2 ,3 ]
Miller, Ian S. [2 ,3 ,4 ]
Shiels, Liam [2 ,3 ]
Connolly, Colum [2 ,3 ]
Mcgarry, Conor K. [1 ,5 ]
Byrne, Annette T. [2 ,3 ,4 ]
Butterworth, Karl T. [1 ]
机构
[1] Queens Univ Belfast, Patrick G Johnston Ctr Canc Res, Belfast, North Ireland
[2] Royal Coll Surgeons Ireland, Dept Physiol & Med Phys, Dublin, Ireland
[3] Royal Coll Surgeons Ireland, Ctr Syst Med, Dublin, Ireland
[4] Royal Coll Surgeons Ireland, Natl Preclin Imaging Ctr, Dublin, Ireland
[5] Belfast Hlth & Social Care Trust, Northern Ireland Canc Ctr, Belfast, North Ireland
基金
英国国家替代、减少和改良动物研究中心; 爱尔兰科学基金会;
关键词
Radiomics; Preclinical; Computed tomography; Cone-beam; Micro; Phantom; Tissue density; Reliability; FEATURE STABILITY; CT; FEATURES; IMAGES; IMPACT; HEAD;
D O I
10.1016/j.phro.2024.100615
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Radiomics analysis extracts quantitative data (features) from medical images. These features could potentially reflect biological characteristics and act as imaging biomarkers within precision medicine. However, there is a lack of cross-comparison and validation of radiomics outputs which is paramount for clinical implementation. In this study, we compared radiomics outputs across two computed tomography (CT)-based preclinical scanners. Materials and methods: Cone beam CT (CBCT) and mu CT scans were acquired using different preclinical CT imaging platforms. The reproducibility of radiomics features on each scanner was assessed using a phantom across imaging energies (40 & 60 kVp) and segmentation volumes (44-238 mm(3)). Retrospective mouse scans were used to compare feature reliability across varying tissue densities (lung, heart, bone), scanners and after voxel size harmonisation. Reliable features had an intraclass correlation coefficient (ICC) > 0.8. Results: First order and GLCM features were the most reliable on both scanners across different volumes. There was an inverse relationship between tissue density and feature reliability, with the highest number of features in lung (CBCT=580, =580, mu CT=734) =734) and lowest in bone (CBCT=110, =110, mu CT=560). =560). Comparable features for lung and heart tissues increased when voxel sizes were harmonised. We have identified tissue-specific preclinical radiomics signatures in mice for the lung (133), heart (35), and bone (15). Conclusions: Preclinical CBCT and mu CT scans can be used for radiomics analysis to support the development of meaningful radiomics signatures. This study demonstrates the importance of standardisation and emphasises the need for multi-centre studies.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] The root canal morphology in mandibular first premolars: a comparative evaluation of cone-beam computed tomography and micro-computed tomography
    Dan Zhang
    Junhong Chen
    Guihua Lan
    Min Li
    Jianping An
    Xiujie Wen
    Luchuan Liu
    Manjing Deng
    Clinical Oral Investigations, 2017, 21 : 1007 - 1012
  • [2] The root canal morphology in mandibular first premolars: a comparative evaluation of cone-beam computed tomography and micro-computed tomography
    Zhang, Dan
    Chen, Junhong
    Lan, Guihua
    Li, Min
    An, Jianping
    Wen, Xiujie
    Liu, Luchuan
    Deng, Manjing
    CLINICAL ORAL INVESTIGATIONS, 2017, 21 (04) : 1007 - 1012
  • [3] Ex Vivo Detection of Apical Delta in Premolars: A Comparative Study Using Periapical Radiography, Cone-beam Computed Tomography, and Micro-computed Tomography
    Leandro Nascimento, Eduarda Helena
    Abrahao Elias, Marcela Ramos
    Fernandes Vasconcelos, Victor Hugo
    Haiter-Neto, Francisco
    Mendonca, Elismauro Francisco
    Sousa, Thiago Oliveira
    JOURNAL OF ENDODONTICS, 2019, 45 (05) : 549 - 553
  • [4] Imaging of the human cochlea using micro-computed tomography before and after cochlear implantation: comparison with cone-beam computed tomography
    Karkas, Alexandre
    Boureille, Pierre
    Laroche, Norbert
    Vico, Laurence
    Bergandi, Florian
    Marotte, Hubert
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2023, 280 (07) : 3131 - 3140
  • [5] Imaging of the human cochlea using micro-computed tomography before and after cochlear implantation: comparison with cone-beam computed tomography
    Alexandre Karkas
    Pierre Boureille
    Norbert Laroche
    Laurence Vico
    Florian Bergandi
    Hubert Marotte
    European Archives of Oto-Rhino-Laryngology, 2023, 280 : 3131 - 3140
  • [6] Comparative study of jaws with multislice computed tomography and cone-beam computed tomography
    Carrafiello, G.
    Dizonno, M.
    Colli, V.
    Strocchi, S.
    Taubert, S. Pozzi
    Leonardi, A.
    Giorgianni, A.
    Barresi, M.
    Macchi, A.
    Bracchi, E.
    Conte, L.
    Fugazzola, C.
    RADIOLOGIA MEDICA, 2010, 115 (04): : 600 - 611
  • [7] Cone-beam Computed Tomographic-based Assessment of Filled C-shaped Canals: Artifact Expression of Cone-beam Computed Tomography as Opposed to Micro-computed Tomography and Nano-computed Tomography
    Mazzi-Chaves, Jardel Francisco
    Vasconcelos, Karla de Faria
    Pauwels, Ruben
    Jacobs, Reinhilde
    Sousa-Neto, Manoel Damiao
    JOURNAL OF ENDODONTICS, 2020, 46 (11) : 1702 - 1711
  • [8] Micro-Computed Tomography-Guided Artificial Intelligence for Pulp Cavity and Tooth Segmentation on Cone-beam Computed Tomography
    Lin, Xiang
    Fu, Yujie
    Ren, Genqiang
    Yang, Xiaoyu
    Duan, Wei
    Chen, Yufei
    Zhang, Qi
    JOURNAL OF ENDODONTICS, 2021, 47 (12) : 1933 - 1941
  • [9] Scatter Correction Based on Beam Stop Array for Cone-Beam Micro-Computed Tomography
    Tang Tianxu
    Duan Xiaojiao
    Zhou Zhizheng
    Wu Qi
    ACTA OPTICA SINICA, 2019, 39 (08)
  • [10] Comparison of the Accuracy of 3-dimensional Cone-beam Computed Tomography and Micro-Computed Tomography Reconstructions by Using Different Voxel Sizes
    Maret, Delphine
    Peters, Ove A.
    Galibourg, Antoine
    Dumoncel, Jean
    Esclassan, Remi
    Kahn, Jean-Luc
    Sixou, Michel
    Telmon, Norbert
    JOURNAL OF ENDODONTICS, 2014, 40 (09) : 1321 - 1326