Flow Cytometry Method Validation Protocols

被引:8
|
作者
Selliah, Nithianandan [1 ]
Nash, Veronica [1 ]
Eck, Steven [2 ]
Green, Cherie [3 ]
Oldaker, Teri
Stewart, Jennifer [4 ]
Vitaliti, Alessandra [5 ]
Litwin, Virginia [6 ]
机构
[1] Cerba Res, Lake Success, NY USA
[2] AstraZeneca, Gaithersburg, MD USA
[3] Ozette Technol, Seattle, WA USA
[4] Flow Contract Site Lab LLC, Bothell, WA USA
[5] Novartis Pharm AG, Novartis Inst Biomed Res, Basel, Switzerland
[6] Eurofins Biopharm Serv, Montreal, PQ, Canada
来源
CURRENT PROTOCOLS | 2023年 / 3卷 / 08期
关键词
biomarker; fit-for-purpose validation; flow cytometry; precision; sensitivity; RECOMMENDATIONS; GUIDELINES;
D O I
10.1002/cpz1.868
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Analytical method validation provides a means to ensure that data are credible and reproducible. This article will provide a brief introduction to analytical method validation as applied to cellular analysis by flow cytometry, along with practical procedures for four different types of validation. The first, Basic Protocol 1 (the limited validation protocol), is recommended for research and non-regulated laboratories. Next, Basic Protocol 2) presents a reasonable, fit-for-purpose validation approach appropriate for biopharma and research settings. Basic Protocol 3 addresses the type of validation performed in clinical laboratories for moderate-risk tests developed in house. Finally, Basic Protocol 4 describes the process that should be applied whenever a method is being transferred from one facility to another. All four validation plans follow the fit-for-purpose validation approach, in which the validation parameters are selected based on the intended use of the assay. These validation protocols represent the minimal requirement and may not be applicable for every intended use such as high-risk clinical assays or data to be used as a primary endpoint in a clinical trial. The recommendations presented here are consistent with the white papers published by the American Association of Pharmaceutical Scientists and the International Clinical Cytometry Society, as well as with Clinical Laboratory Standards Institute Guideline H62: Validation of Assays Performed by Flow Cytometry (CLSI, 2021). (c) 2023 Wiley Periodicals LLC.Basic Protocol 1: Limited validationBasic Protocol 2: Fit-for-purpose validation for biopharma and research settingsBasic Protocol 3: Validation for moderate clinical risk laboratory developed testsBasic Protocol 4: Transfer validation
引用
收藏
页数:34
相关论文
共 50 条
  • [1] Flow Cytometry Method Validation Protocols (vol 3, e868, 2023)
    Selliah, Nithianandan
    Nash, Veronica
    Eck, Steven
    Green, Cherie
    Oldaker, Teri
    Stewart, Jennifer
    Vitaliti, Alessandra
    Litwin, Virginia
    CURRENT PROTOCOLS, 2024, 4 (01):
  • [2] Flow cytometry as an integrative method for the evaluation of vaccine immunogenicity: A validation approach
    Gianchecchi, Elena
    Torelli, Alessandro
    Piu, Pietro
    Bonifazi, Carolina
    Ganfini, Luisa
    Montomoli, Emanuele
    BIOCHEMISTRY AND BIOPHYSICS REPORTS, 2023, 34
  • [3] An Approach to the Validation of Flow Cytometry Methods
    Cunliffe, Jo
    Derbyshire, Nicola
    Keeler, Sue
    Coldwell, Ruth
    PHARMACEUTICAL RESEARCH, 2009, 26 (12) : 2551 - 2557
  • [4] An Approach to the Validation of Flow Cytometry Methods
    Jo Cunliffe
    Nicola Derbyshire
    Sue Keeler
    Ruth Coldwell
    Pharmaceutical Research, 2009, 26 : 2551 - 2557
  • [5] Relevance of Antibody Validation for Flow Cytometry
    Kalina, Tomas
    Lundsten, Kelly
    Engel, Pablo
    CYTOMETRY PART A, 2020, 97 (02) : 126 - 136
  • [6] Flow Cytometry Protocols, 3rd Edition
    Lacinova, Lubica
    GENERAL PHYSIOLOGY AND BIOPHYSICS, 2012, 31 (03) : 351 - 352
  • [7] Validation of a method for quick concentration assessment of boar semen using flow cytometry
    Martinez-Pastor, F.
    Pardal-Gil, S.
    Justel, V.
    Gonzalez, J. R.
    Dominguez, J. C.
    Gonzalez-Urdiales, R.
    REPRODUCTION IN DOMESTIC ANIMALS, 2015, 50 : 66 - 66
  • [8] Impedance spectroscopy flow cytometry: Model validation
    Gawad, S
    Cheung, K
    Renaud, P
    MICRO TOTAL ANALYSIS SYSTEMS 2004, VOL 1, 2005, (296): : 58 - 60
  • [9] VALIDATION OF IMMUNOPHENOTYPE BY FLOW CYTOMETRY IN MYELODYSPLASTIC SYNDROMES
    Vikentiou, Myrofora
    Psarra, Katherine
    Kapsimali, Violetta
    Papasteriades, Chrysa
    CYTOMETRY PART B-CLINICAL CYTOMETRY, 2010, 78B (06) : 448 - 448
  • [10] Development and validation of a method for the measurement of sperm concentration by fluorescence-activated flow cytometry
    Eustache, F
    Auger, J
    Jouannet, P
    HUMAN REPRODUCTION, 1999, 14 : 150 - 150