Assessment of spectral ghost artifacts in echo-planar spectroscopic micro-imaging with flyback readout

被引:0
|
作者
Weis, Jan [1 ]
Babos, Magor [2 ]
Estrada, Sergio [3 ]
Selvaraju, Ram Kumar [3 ]
机构
[1] Uppsala Univ Hosp, Dept Med Phys, Uppsala, Sweden
[2] Mediso Med Imaging Syst, Budapest, Hungary
[3] Uppsala Univ, Dept Med Chem, Uppsala, Sweden
来源
SCIENTIFIC REPORTS | 2024年 / 14卷 / 01期
关键词
Magnetic resonance; Echo-planar spectroscopic imaging; Flyback readout; Spectral ghost artifacts; High-spatial resolution; Micro-imaging; CHEMICAL-SHIFT; FOURIER-TRANSFORM; HUMAN BRAIN; RESOLUTION; QUANTIFICATION; SENSITIVITY; WATER;
D O I
10.1038/s41598-024-73391-y
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
In this work, echo-planar spectroscopic imaging (EPSI) with flyback readout gradient-echo train was implemented in a preclinical MR scanner. The aim of this study is to visualize and quantify the ghost spectral lines produced by two, three and four interleaved echo trains with different amplitudes of the readout gradients, and to investigate the feasibility of the flyback data acquisition in micro-imaging of small animals. Applied multi-slice EPSI sequence utilizes asymmetric gradient-echo train that combines the shortest possible rewind gradients with readout gradients. It simplifies data processing because all echoes are acquired with the same polarity of the readout gradient. The approach with four interleaved gradient-echo trains and with four echoes in each train provides broad spectral bandwidth in combination with narrow receiver bandwidth and a good water-fat signal separation. It improves signal-to-noise ratio without the undesired consequence of water-fat shift artifacts that are eliminated during data processing. Position, number, and intensity of the ghost spectral lines can be controlled by the suitable choice of spectral bandwidth, number of echo train interleaves, and the number of echoes in each interleave. This study demonstrates that high-spatial resolution EPSI with interleaved flyback readout gradient-echo trains is feasible on standard preclinical scanners.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Design of flyback echo-planar readout gradients for magnetic resonance spectroscopic imaging
    Cunningham, CH
    Vigneron, DB
    Chen, AP
    Xu, D
    Nelson, SJ
    Hurd, RE
    Kelley, DA
    Pauly, JM
    MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (05) : 1286 - 1289
  • [2] Reduction of spectral ghost artifacts in high-resolution echo-planar spectroscopic Imaging of water and fat resonances
    Du, WL
    Du, YP
    Fan, XB
    Zamora, MA
    Karczmar, GS
    MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (06) : 1113 - 1120
  • [3] Reducing flow artifacts in echo-planar imaging
    Pat, GTL
    Meyer, CH
    Pauly, JM
    Nishimura, DG
    MAGNETIC RESONANCE IN MEDICINE, 1997, 37 (03) : 436 - 447
  • [4] ECHO-PLANAR DIFFUSION SPECTROSCOPIC IMAGING
    BITO, Y
    HIRATA, S
    NABESHIMA, T
    YAMAMOTO, E
    MAGNETIC RESONANCE IN MEDICINE, 1995, 33 (01) : 69 - 73
  • [5] A silent echo-planar spectroscopic imaging readout with high spectral bandwidth MRSI using an ultrasonic gradient axis
    Versteeg, Edwin
    Nam, Kyung Min
    Klomp, Dennis W. J.
    Bhogal, Alex A.
    Siero, Jeroen C. W.
    Wijnen, Jannie P.
    MAGNETIC RESONANCE IN MEDICINE, 2024, 91 (06) : 2247 - 2256
  • [6] Phosphorus magnetic resonance spectroscopic imaging using flyback echo planar readout trajectories
    Santos-Diaz, Alejandro
    Obruchkov, Sergei I.
    Schulte, Rolf F.
    Noseworthy, Michael D.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2018, 31 (04) : 553 - 564
  • [7] Phosphorus magnetic resonance spectroscopic imaging using flyback echo planar readout trajectories
    Alejandro Santos-Díaz
    Sergei I. Obruchkov
    Rolf F. Schulte
    Michael D. Noseworthy
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2018, 31 : 553 - 564
  • [8] Ghost artifact suppression for interleaved echo-planar imaging
    Buonocore, MH
    Zhu, D
    Bronstein, JM
    RADIOLOGY, 1998, 209P : 356 - 357
  • [9] Volumetric multishot echo-planar spectroscopic imaging
    Tyszka, JM
    Mamelak, AN
    MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (02) : 219 - 227
  • [10] Dynamic 31P spectroscopic imaging of skeletal muscles combining flyback echo-planar spectroscopic imaging and compressed sensing
    Santos-Diaz, Alejandro
    Harasym, Diana
    Noseworthy, Michael D.
    MAGNETIC RESONANCE IN MEDICINE, 2019, 81 (06) : 3453 - 3461