Preoperative Planning Software Does Not Accurately Predict Range of Motion in Reverse Total Shoulder Arthroplasty

被引:0
|
作者
Thomas, Logan G. [1 ]
Chalmers, Peter N. [1 ]
Henninger, Heath B. [1 ]
Davis, Evan W. [1 ]
Tashjian, Robert Z. [1 ]
机构
[1] Univ Utah, Dept Orthopaed Surg, Salt Lake City, UT 84112 USA
基金
美国国家卫生研究院;
关键词
D O I
10.5435/JAAOS-D-23-00519
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The purpose of this study was to determine whether preoperative planning software (PPS) accurately predicts clinical range of motion (ROM) in patients with reverse total shoulder arthroplasty 1 year postoperatively with preoperative and postoperative computed tomography (CT) scans. Methods: This was a retrospective study of 16 reverse total shoulder arthroplasty patients with preoperative and postoperative (CT) scans obtained at least 1 year postoperatively. Clinical ROM was measured in abduction, external rotation at resting abduction, extension, and flexion at a minimum of 1 year postoperatively. All clinical measurements were obtained before generation of PPS ROM values. Using postoperative CT scans, the achieved implant component positions were quantified and then replicated in PPS on the preoperative CT scans. The preoperative predicted ROM was then recorded, both with and without osteophyte removal. Bland-Altman plots were generated within each motion comparing the differences between clinically measured motion and software-predicted motion. Results: The variation in clinically measured ROM in abduction, external rotation at resting abduction, extension, and flexion were 118 +/- 27 (65 degrees to 180 degrees), 33 +/- 16 (10 degrees to 75 degrees), 56 +/- 8 (50 degrees to 65 degrees), and 137 +/- 25 (80 degrees to 160 degrees), respectively. Clinically measured motion differed greatly from PPS-predicted ROM, with mean differences of 33 +/- 29 (-32 to 93) for abduction, 44 +/- 25 (-38 to 57) for external rotation, 44 +/- 25 (-35 to 65) for extension, and 54 +/- 50 (-51 to 147) for flexion with no significant correlations between clinically measured and PPS-predicted ROM (P > 0.05). With humeral or humeral and glenoid osteophyte resection, correlations for only flexion became significant (P = 0.002 for both). Conclusion: The passive glenohumeral impingement-free ROM generated from PPS incompletely predicts clinically measured active humerothoracic ROM, possibly because of the unmeasured factors of soft-tissue tension, muscular strength, humeral torsion, resting scapular posture, and, most importantly, scapulothoracic motion. Level of Evidence: IV.
引用
收藏
页码:e378 / e386
页数:9
相关论文
共 50 条
  • [31] Reverse total shoulder arthroplasty
    Familiari, Filippo
    Rojas, Jorge
    Doral, Mahmut Nedim
    Huri, Gazi
    McFarland, Edward G.
    EFORT OPEN REVIEWS, 2018, 3 (02): : 58 - 69
  • [32] Reverse total shoulder arthroplasty
    Scalise, J. J.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2009, 60 (01) : 53 - 59
  • [33] The reverse total shoulder arthroplasty
    Matsen, Frederick A., III
    Boileau, Pascal
    Walch, Gilles
    Gerber, Christian
    Bicknell, Ryan T.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (03): : 660 - 667
  • [34] Reverse Total Shoulder Arthroplasty
    Sanchez-Sotelo, Joaquin
    CLINICAL ANATOMY, 2009, 22 (02) : 172 - 182
  • [35] Reverse Total Shoulder Arthroplasty
    Gerber, Christian
    Pennington, Scott D.
    Nyffeler, Richard W.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2009, 17 (05) : 284 - 295
  • [36] Reverse Total Shoulder Arthroplasty
    Lenarz, Christopher J.
    Gobezie, Reuben
    JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2011, (53):
  • [37] Factors that predict postoperative motion in patients treated with reverse shoulder arthroplasty
    Schwartz, Daniel Grant
    Cottrell, Benjamin J.
    Teusink, Matthew J.
    Clark, Rachel E.
    Downes, Katheryne L.
    Tannenbaum, Richard S.
    Frankle, Mark A.
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2014, 23 (09) : 1289 - 1295
  • [38] Evaluation of the accuracy and interobserver agreement of the reverse shoulder angle in the preoperative planning of reverse total shoulder arthroplasties
    Motta, Geraldo
    Amaral, Marcus Vinicius
    Cohen, Marcio
    Schiefer, Marcio
    da Fonseca, Raphael Soares
    Oliveira, Ana Carolina Leal Carolina
    SHOULDER & ELBOW, 2024,
  • [39] Center of rotation affects abduction range of motion of reverse shoulder arthroplasty
    Gutierrez, Sergio
    Levy, Jonathan C.
    Lee, William E., III
    Keller, Tony S.
    Maitland, Murray E.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (458) : 78 - 82
  • [40] A Review of Preoperative Planning and Technical Considerations for Anatomic Total Shoulder Arthroplasty
    Defoor, Mikalyn T.
    Hartzler, Robert U.
    Sheean, Andrew J.
    OPERATIVE TECHNIQUES IN ORTHOPAEDICS, 2024, 34 (04)