Measurement uncertainty in clinical chemistry: ISO 20914 versus nordtest or intermediate precision versus bias

被引:0
|
作者
Nurlu, Nilhan [1 ]
Cat, Abdulkadir [1 ]
Ucar, Kamil Taha [2 ]
机构
[1] Hlth Sci Univ, Istanbul Gaziosmanpasa Training & Res Hosp, Dept Med Biochem, Istanbul, Turkiye
[2] Hlth Sci Univ, Istanbul Basaksehir Cam & Sakura City Hosp, Dept Med Biochem, Istanbul, Turkiye
关键词
Measurement uncertainty; bias; clinical chemistry; quality management; analytical phase; LABORATORY MEDICINE; TOTAL ERROR;
D O I
10.1080/00365513.2024.2338738
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
AimMeasuring uncertainty (MU) is crucial to ensure the accuracy and precision of laboratory results. This study compares the ISO 20914 and Nordtest guidelines to analyze the MU values for 20 clinical chemistry analytes over six months.MethodsThe researchers calculated MU components, including within-laboratory reproducibility (Rw), laboratory analytical performance bias (u(bias)), and combined standard uncertainty (uc), based on internal quality control and external quality assessment data. The final expanded uncertainty (U) values were determined by multiplying the combined uncertainty with a coverage factor (k = 2 for 95% Confidence Interval), following each guideline's respective procedures. Clinical chemistry analytes were analyzed on Roche Cobas 6000 c501 auto analyzer (Roche Diagnostics, Mannheim, Germany) and manufacturer's kits were used analysis.ResultsThe results show that 11 out of 20 clinical chemistry analytes met the targeted maximum allowable measurement uncertainty (MAU) values when calculated according to ISO 20914 guideline. Also, 11 out of 20 clinical chemistry analytes' MU values met the MAU values with the Nordtest guideline's recommended calculations. However, some tests met the MAU in the ISO 20914 approach but not in the Nordtest guideline, and vice versa.ConclusionsThe study found that intermediate precision (uRw) in the ISO 20914 approach and performance bias (u(bias)) in the Nordtest approach significantly impacted MU values. The research highlights the importance of standardization in MU calculation approaches across clinical laboratories. These findings have implications for patient care and clinical decision-making, emphasizing the importance of selecting appropriate laboratory guidelines for routine use.
引用
收藏
页码:147 / 153
页数:7
相关论文
共 50 条
  • [1] Using ISO/TS 20914:2019 to calculate the measurement uncertainty of immunosuppressive drugs in a clinical laboratory
    Grote-Koska, Denis
    Brand, Korbinian
    [J]. SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2020, 80 (04): : 309 - 312
  • [2] A Practical Approach to ISO 15189 and Measurement Uncertainty in Clinical Chemistry: Analytic Bias is Still an Important Challenge
    Ulgen, Yekta
    [J]. 2019 IEEE INTERNATIONAL SYMPOSIUM ON MEDICAL MEASUREMENTS AND APPLICATIONS (MEMEA), 2019,
  • [3] An outline of measurement uncertainty of total protein in urine estimated according to the ISO Technical Specification 20914
    Borrillo, Francesca
    Panteghini, Mauro
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (08) : e175 - e177
  • [4] MEASUREMENT UNCERTAINTY FOR ROUTINE MEASUREMENT PROCEDURES IN CLINICAL CHEMISTRY LABORATORY ACCREDITED ACCORDING TO ISO 15189
    Flegar-Mestric, Z.
    Surina, B.
    Perkov, S.
    Ozvald, I.
    Simonovic, B.
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 : S806 - S806
  • [5] Precision in ROI analysis: Variance versus bias
    Hansen, J.
    Svarer, C.
    Marner, L.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S342 - S343
  • [6] DESCARTES VERSUS SPINOZA: TRUTH, UNCERTAINTY, AND BIAS
    Street, Chris N. H.
    Richardson, Daniel C.
    [J]. SOCIAL COGNITION, 2015, 33 (03) : 227 - 239
  • [7] MEASUREMENT OF ANALYTICAL PRECISION IN CLINICAL CHEMISTRY
    BROUGHTON, PM
    ANNAN, W
    [J]. CLINICA CHIMICA ACTA, 1971, 32 (03) : 433 - +
  • [8] Calculation of measurement uncertainty in clinical chemistry
    Stöckl, D
    Van Uytfanghe, K
    Cabaleiro, DR
    Thienpont, LM
    [J]. CLINICAL CHEMISTRY, 2005, 51 (01) : 276 - 277
  • [9] Calculating uncertainty of measurement for serology assays by use of precision and bias
    Dimech, W
    Francis, B
    Kox, J
    Roberts, G
    [J]. CLINICAL CHEMISTRY, 2006, 52 (03) : 526 - 529
  • [10] Total Analytical Error Versus Measurement Uncertainty
    Westgard, James O.
    Westgard, Sten A.
    [J]. CLINICS IN LABORATORY MEDICINE, 2017, 37 (01) : 1 - +