Comparative study of the clinical performance of two supraglottic airway devices, Laryngeal Mask Airway Supreme and BASKA Balum Mask in adult patients for general anaesthesia

被引:0
|
作者
Singh, Pradeep [1 ]
Bhatia, P. S. [1 ]
Gupta, Arushi [1 ]
Mehta, Khushboo [1 ]
机构
[1] Vardhaman Mahavir Med Coll, Dept Anaesthesia & Intens Care, New Delhi 110029, India
关键词
BASKA mask; Endotracheal tube; General anaesthesia; Laryngeal mask airway supreme; Oropharyngeal leak pressure; Supraglottic airway device; REUSABLE PROSEAL LMA; I-GEL; VENTILATION; PRESSURE; EFFICACY;
D O I
10.1016/j.tacc.2023.101308
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background The oropharyngeal leak pressure (OLP) of supraglottic airway devices (SGADs) is an important determinant of their efficacy to provide positive pressure ventilation, whereas the provision to prevent aspiration determines its safety. This study aims to compare the clinical performance of Laryngeal Mask Airway (LMA) Supreme versus BASKA Balum Mask. Methods Seventy adult patients (18-65 years) posted for elective surgeries general anaesthesia were prospectively randomised into group B (BASKA Mask) and Group S (LMA Supreme). The primary outcome was to measure the OLP at 5 and 30 min after device insertion and secondarily, to compare the ease and time of insertion, number of attempts, leak fraction, anatomical position of the devices, and adverse respiratory event if any. Results The OLP of BASKA Mask was significantly higher than LMA Supreme (31.37 +/- 1.73 cmH2O versus 23.6 +/- 1.73 cmH2O at 5 min and 35.94 +/- 3.84 cmH2O versus 23.71 +/- 1.69 cmH2O at 30 min). The time taken to insert BASKA Mask was also less (11.49 +/- 2.38 s) than LMA Supreme (14.66 +/- 2.79 s). Peak airway pressures were also notably less in group B than group S at both 5 and 30 min after device placement (13.43 +/- 1.44 vs 18.66 +/- 1.43 cmH2O and 14.91 +/- 1.46 vs 21.37 +/- 1.88 cmH2O). The leak fraction ranged between 1.5 and 4.3% in both groups. All other parameters studied were also comparable. Conclusions BASKA Mask had a higher OLP, lower peak pressure, shorter time to insertion than LMA Supreme. The leak fraction, ease of insertion, number of attempts, and anatomical position of the devices were comparable in both groups.
引用
收藏
页数:6
相关论文
共 50 条
  • [41] Comparative study between I-gel, a new supraglottic airway device, and classical laryngeal mask airway in anesthetized spontaneously ventilated patients
    Helmy, Amr M.
    Atef, Hossam M.
    El Taher, Ezzat M.
    Henidak, Ahmed Mosaad
    SAUDI JOURNAL OF ANAESTHESIA, 2010, 4 (03) : 131 - 136
  • [42] A new laryngeal mask supraglottic airway device with integrated balloon line: a descriptive and comparative bench study
    Zhou, YingHai
    Jew, Korinne
    MEDICAL DEVICES-EVIDENCE AND RESEARCH, 2016, 9 (09): : 383 - 388
  • [43] A clinical trial evaluating the laryngeal mask airway-Supreme in obese children during general anesthesia
    Tian, Yue
    Wu, Xiu-ying
    Li, Lu
    Ma, Ling
    Li, Yun-feng
    ARCHIVES OF MEDICAL SCIENCE, 2017, 13 (01) : 183 - 190
  • [44] Time to abandon the 'vintage' laryngeal mask airway and adopt second-generation supraglottic airway devices as first choice
    Cook, T. M.
    Kelly, F. E.
    BRITISH JOURNAL OF ANAESTHESIA, 2015, 115 (04) : 497 - 499
  • [45] Use of ProSeal™ Laryngeal Mask Airway in 2114 Adult Patients: A Prospective Study
    Goldmann, Kai
    Hechtfischer, Carolin
    Malik, Amena
    Kussin, Andrea
    Freisburger, Christian
    ANESTHESIA AND ANALGESIA, 2008, 107 (06): : 1856 - 1861
  • [46] Two oxygen delivery devices compared in a modified laryngeal mask airway
    Adrian K. Dashfield
    Jeremy A. Langton
    Charles G. Johnston
    Carsten Eickmann
    Canadian Journal of Anaesthesia, 1997, 44 : 572 - 573
  • [47] Instrumental and randomised clinical comparison between laryngeal mask airway Proseal and Supreme in pediatric patients
    Santambrogio, Luisa
    Righi, Sabrina
    Pinciroli, Rosa L.
    Piro, Eugenia
    D'alessio, Antonio
    Minuto, Aldo
    TRENDS IN ANAESTHESIA AND CRITICAL CARE, 2020, 30 : 14 - 21
  • [48] Two oxygen delivery devices compared in a modified laryngeal mask airway
    Dashfield, AK
    Langton, JA
    Johnston, CG
    Eickmann, C
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1997, 44 (05): : 572 - 573
  • [49] Intubating laryngeal mask airway in morbidly obese and lean patients - A comparative study
    Combes, X
    Sauvat, S
    Leroux, B
    Dumerat, M
    Sherrer, E
    Motamed, C
    Brain, A
    D'Honneur, G
    ANESTHESIOLOGY, 2005, 102 (06) : 1106 - 1109
  • [50] Clinical Performance of I-gel® and BlockBuster™ Laryngeal Mask Airway in Adult Patients during General Anesthesia: A Randomized Comparison
    Das, Prerna Prabhat
    Tiwari, Tanmay
    Raman, Rajesh
    Gautam, Shefali
    Kushwaha, B. B.
    Kumar, Yatendra
    INTERNATIONAL JOURNAL OF APPLIED AND BASIC MEDICAL RESEARCH, 2024, 14 (02) : 108 - 113