Comparative study of the clinical performance of two supraglottic airway devices, Laryngeal Mask Airway Supreme and BASKA Balum Mask in adult patients for general anaesthesia

被引:0
|
作者
Singh, Pradeep [1 ]
Bhatia, P. S. [1 ]
Gupta, Arushi [1 ]
Mehta, Khushboo [1 ]
机构
[1] Vardhaman Mahavir Med Coll, Dept Anaesthesia & Intens Care, New Delhi 110029, India
关键词
BASKA mask; Endotracheal tube; General anaesthesia; Laryngeal mask airway supreme; Oropharyngeal leak pressure; Supraglottic airway device; REUSABLE PROSEAL LMA; I-GEL; VENTILATION; PRESSURE; EFFICACY;
D O I
10.1016/j.tacc.2023.101308
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background The oropharyngeal leak pressure (OLP) of supraglottic airway devices (SGADs) is an important determinant of their efficacy to provide positive pressure ventilation, whereas the provision to prevent aspiration determines its safety. This study aims to compare the clinical performance of Laryngeal Mask Airway (LMA) Supreme versus BASKA Balum Mask. Methods Seventy adult patients (18-65 years) posted for elective surgeries general anaesthesia were prospectively randomised into group B (BASKA Mask) and Group S (LMA Supreme). The primary outcome was to measure the OLP at 5 and 30 min after device insertion and secondarily, to compare the ease and time of insertion, number of attempts, leak fraction, anatomical position of the devices, and adverse respiratory event if any. Results The OLP of BASKA Mask was significantly higher than LMA Supreme (31.37 +/- 1.73 cmH2O versus 23.6 +/- 1.73 cmH2O at 5 min and 35.94 +/- 3.84 cmH2O versus 23.71 +/- 1.69 cmH2O at 30 min). The time taken to insert BASKA Mask was also less (11.49 +/- 2.38 s) than LMA Supreme (14.66 +/- 2.79 s). Peak airway pressures were also notably less in group B than group S at both 5 and 30 min after device placement (13.43 +/- 1.44 vs 18.66 +/- 1.43 cmH2O and 14.91 +/- 1.46 vs 21.37 +/- 1.88 cmH2O). The leak fraction ranged between 1.5 and 4.3% in both groups. All other parameters studied were also comparable. Conclusions BASKA Mask had a higher OLP, lower peak pressure, shorter time to insertion than LMA Supreme. The leak fraction, ease of insertion, number of attempts, and anatomical position of the devices were comparable in both groups.
引用
收藏
页数:6
相关论文
共 50 条
  • [1] Efficacy of Baska mask and Laryngeal mask airway supreme during positive pressure ventilation - A comparative study
    Jayalekshmi, Sreedevi
    Paul, Cherish
    Thomas, Mariam K.
    JOURNAL OF ANAESTHESIOLOGY CLINICAL PHARMACOLOGY, 2020, 36 (01) : 31 - 36
  • [2] An observational study of the Baska® mask: a novel supraglottic airway
    Alexiev, V.
    Salim, A.
    Kevin, L. G.
    Laffey, J. G.
    ANAESTHESIA, 2012, 67 (06) : 640 - 645
  • [3] A RANDOMIZED CROSSOVER COMPARISON OF THE LARYNGEAL MASK AIRWAY PROSEAL AND LARYNGEAL MASK AIRWAY SUPREME IN ANESTHETIZED ADULT PATIENTS
    Jannu, Vinayaka
    Dhorigol, M. G.
    Sanikop, C. S.
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2015, 4 (85): : 14828 - 14833
  • [4] Comparison of Baska Mask Versus Proseal Laryngeal Mask Airway in Elective Surgeries Under General Anaesthesia: A Randomized Clinical Trial
    Jose, Jophy
    Kagalkar, Nirmala Devi
    Kattimani, Milind M.
    Suntan, Anusha
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (04)
  • [5] A comparison of the Supreme™ laryngeal mask airway with the Proseal™ laryngeal mask airway in anesthetized paralyzed adult patients: a randomized crossover study
    Tham, Huae Min
    Tan, Su Meng
    Woon, Kwee Lian
    Zhao, Yu Dong
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2010, 57 (07): : 672 - 678
  • [6] Laryngeal mask airway and other supraglottic airway devices in paediatric practice
    Patel, Bhavesh
    Bingham, Robert
    BJA EDUCATION, 2009, 9 (01) : 6 - 9
  • [7] Clinical Comparison of I-gel and Laryngeal Mask Airway-Supreme Airway Devices During General Anaesthesia in the Paediatric Population
    Aggarwal, Manish
    Yadav, Rahul
    Singh, Shalendra
    Bansal, Deepali
    TURKISH JOURNAL OF ANAESTHESIOLOGY AND REANIMATION, 2021, 49 (03) : 244 - 249
  • [8] Comparison of laryngeal mask airway Supreme and laryngeal mask airway Pro-Seal for controlled ventilation during general anaesthesia in adult patients Systematic review with meta-analysis
    Maitra, Souvik
    Khanna, Puneet
    Baidya, Dalim Kumar
    EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2014, 31 (05) : 266 - 273
  • [9] SUPRAGLOTTIC AIRWAY DEVICES OTHER THAN LARYNGEAL MASK AIRWAY AND ITS PROTOTYPES
    Sinha, Prabhat Kumar
    Misra, Satyajeet
    INDIAN JOURNAL OF ANAESTHESIA, 2005, 49 (04) : 281 - 292
  • [10] Comparative evaluation of airway dynamics in patients undergoing laparoscopic cholecystectomy under general anaesthesia with controlled ventilation using ProSeal laryngeal mask airway, I-GelT and Baska mask
    Kumar, Brijender
    Gupta, Bhanu
    Thakur, Aman
    Rana, Shelly
    Garg, Mohini
    Nehria, Maninder
    INDIAN JOURNAL OF ANAESTHESIA, 2022, 66 (12) : 842 - 848