Design and development of the DE-SPECT system: a clinical SPECT system for broadband multi-isotope imaging of peripheral vascular disease

被引:0
|
作者
Zannoni, E. M. [1 ]
Sankar, P. [2 ]
Jin, Y. [1 ]
Liu, C. [3 ]
Sinusas, A. J. [3 ,4 ]
Metzler, S. D. [2 ]
Meng, L. J. [1 ,5 ]
机构
[1] Univ Illinois, Dept Nucl Plasma & Radiol Engn, Urbana, IL 61801 USA
[2] Univ Penn, Dept Radiol, Philadelphia, PA USA
[3] Yale Univ, Sch Med, Dept Radiol & Biomed Imaging, New Haven, CT USA
[4] Yale Univ, Sch Med, Dept Med, New Haven, CT USA
[5] Univ Illinois, Beckman Inst Adv Sci & Technol, Urbana, IL USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2024年 / 69卷 / 12期
关键词
organ-dedicated imaging; SPECT; CZT detectors; peripheral vascular disease; RECONSTRUCTION; CAMERA; PERFORMANCE; COLLIMATOR;
D O I
10.1088/1361-6560/ad5266
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Peripheral Vascular Disease (PVD) affects more than 230 million people worldwide and is one of the leading causes of disability among people over age 60. Nowadays, PVD remains largely underdiagnosed and undertreated, and requires the development of tailored diagnostic approaches. We present the full design of the Dynamic Extremity SPECT (DE-SPECT) system, the first organ-dedicated SPECT system for lower extremity imaging, based on 1 cm thick Cadmium Zinc Telluride (CZT) spectrometers and a dynamic dual field-of-view (FOV) synthetic compound-eye (SCE) collimator. Approach. The proposed DE-SPECT detection system consists of 48 1 cm thick 3D-position-sensitive CZT spectrometers arranged in a partial ring of 59 cm in diameter in a checkerboard pattern. The detection system is coupled with a compact dynamic SCE collimator that allows the user to select between two different FOVs at any time during an imaging study: a wide-FOV (28 cm diameter) configuration for dual-leg or scout imaging or a high-resolution and high-sensitivity (HR-HS) FOV (16 cm diameter) for single-leg or focused imaging. Main results. The preliminary experimental data show that the CZT spectrometer achieves a 3D intrinsic spatial resolution of <0.75 mm FWHM and an excellent energy resolution over a broad energy range (2.6 keV FWHM at 218, 3.3 keV at 440 keV). From simulations, the wide-FOV configuration offers a 0.034% averaged sensitivity at 140 keV and <8 mm spatial resolution, whereas the HR-HS configuration presents a peak central sensitivity of 0.07% at 140 keV and a similar to 5 mm spatial resolution. The dynamic SCE collimator enables the capability to perform joint reconstructions that would ensure an overall improvement in imaging performance. Significance. The DE-SPECT system is a stationary and high-performance SPECT system that offers an excellent spectroscopic performance with a unique computer-controlled dual-FOV imaging capability, and a relatively high sensitivity for multi-tracer and multi-functional SPECT imaging of the extremities.
引用
收藏
页数:17
相关论文
共 50 条
  • [1] Design and development of the DE-SPECT system: A clinical SPECT system dedicated for dynamic imaging of peripheral vascular diseases
    Zannoni, Elena Maria
    Sankar, Poopalasingam
    Metzler, Scott
    Liu, Chi
    Sinusas, Albert
    Meng, Ling-Jian
    JOURNAL OF NUCLEAR MEDICINE, 2021, 62
  • [2] Multi-isotope quantification in pre-clinical SPECT
    Iwata, Koji
    Coco, Brent
    Zhang, JoAnn
    Wagenaar, Douglas
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [3] Multi-Isotope Capabilities of a Small-Animal Multi-Pinhole SPECT System
    Lukas, Mathias
    Kluge, Anne
    Beindorff, Nicola
    Brenner, Winfried
    JOURNAL OF NUCLEAR MEDICINE, 2020, 61 (01) : 152 - 161
  • [4] Quantitative Multi-Isotope Imaging of Alpha-Emitter Radiopharmaceutical Therapy with the Alpha-SPECT-Mini System
    Yang, Can
    Zannoni, Elena Maria
    Cai, Ling
    Wilson, Matthew
    Bastiaannet, Remco
    Du, Yong
    Sgouros, George
    Frey, Eric
    Meng, Ling Jian
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [5] Performance evaluation of a preclinical SPECT/CT system for multi-animal and multi-isotope quantitative experiments
    Elena Prieto
    Leticia Irazola
    María Collantes
    Margarita Ecay
    Teresa Cuenca
    Josep Mª Martí-Climent
    Iván Peñuelas
    Scientific Reports, 12
  • [6] Performance evaluation of a preclinical SPECT/CT system for multi-animal and multi-isotope quantitative experiments
    Prieto, Elena
    Irazola, Leticia
    Collantes, Maria
    Ecay, Margarita
    Cuenca, Teresa
    Marti-Climent, Josep Ma
    Penuelas, Ivan
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [7] Multi-isotope SPECT imaging of the 225Ac decay chain: feasibility studies
    Robertson, A. K. H.
    Ramogida, C. F.
    Rodriguez-Rodriguez, C.
    Blinder, Stephan
    Kunz, Peter
    Sossi, Vesna
    Schaffer, Paul
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (11): : 4406 - 4420
  • [8] Experimental Evaluation of DE-SPECT: A Hyperspectral SPECT System for in Vivo 3-D Gamma-Ray Spectrometry of Multifunctional Molecular Imaging and Theragnostic
    Jin, Yifei
    Zannoni, Elena Maria
    Sankar, Poopalasingam
    Zhang, Feng
    Streicher, Michael
    Yang, Hao
    Gura, Drew
    Wu, Ryan
    He, Zhong
    Metzler, Scott
    Liu, Chi
    Sinusas, Albert
    Meng, Ling Jian
    JOURNAL OF NUCLEAR MEDICINE, 2024, 65
  • [9] Design and Development of a Dedicated Robotic SPECT System for Cardiac Imaging
    Ay, M. R.
    Teimourian, B.
    Sajedi, S.
    Akbarzadeh, A.
    Kaviani, S.
    Farahani, M.
    Farzanefar, S.
    Shapouri, Z.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S68 - S68
  • [10] System design and development of a pinhole SPECT system for quantitative functional imaging of small animals
    Aoi, T
    Zeniya, T
    Watabe, H
    Deloar, HM
    Matsuda, T
    Iida, H
    ANNALS OF NUCLEAR MEDICINE, 2006, 20 (03) : 245 - 251