Effect of Skin Pigmentation and Finger Choice on Accuracy of Oxygen Saturation Measurement in an IoT-Based Pulse Oximeter

被引:0
|
作者
Haxha, Shyqyri [1 ]
Nwibor, Chike [1 ]
Ali, Mian [1 ]
Sakel, Mohamed [2 ]
Saunders, Karen [2 ]
Dyo, Vladimir [1 ]
Nabakooza, Shakira [3 ]
机构
[1] Royal Holloway Univ London, Dept Elect Engn, Egham TW20 0EX, England
[2] East Kent Hosp Univ NHS Fdn Trust, Canterbury CT2 7NT, England
[3] Hertfordshire Partnership Univ NHS Fdn Trust, Hatfield AL3 5TQ, St Albans, England
关键词
optical sensors; pulse oximetry; Internet of Things; photoplethysmography; NAIL POLISH; RELIABILITY; PRINCIPLES; AGREEMENT; PRESSURE; ERRORS; SYSTEM; HEART;
D O I
10.3390/s24113301
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Pulse oximeters are widely used in hospitals and homes for measurement of blood oxygen saturation level (SpO2) and heart rate (HR). Concern has been raised regarding a possible bias in obtaining pulse oximeter measurements from different fingertips and the potential effect of skin pigmentation (white, brown, and dark). In this study, we obtained 600 SpO2 measurements from 20 volunteers using three UK NHS-approved commercial pulse oximeters alongside our custom-developed sensor, and used the Munsell colour system (5YR and 7.5YR cards) to classify the participants' skin pigmentation into three distinct categories (white, brown, and dark). The statistical analysis using ANOVA post hoc tests (Bonferroni correction), a Bland-Altman plot, and a correlation test were then carried out to determine if there was clinical significance in measuring the SpO2 from different fingertips and to highlight if skin pigmentation affects the accuracy of SpO2 measurement. The results indicate that although the three commercial pulse oximeters had different means and standard deviations, these differences had no clinical significance.
引用
收藏
页数:17
相关论文
共 32 条
  • [21] Measuring accuracy of plethysmography based respiratory rate measurement using pulse oximeter at a tertiary hospital in India
    Alwadhi, Varun
    Sarin, Enisha
    Kumar, Praveen
    Saboth, Prasant
    Khera, Ajay
    Gupta, Sachin
    Kumar, Harish
    PNEUMONIA, 2020, 12 (01)
  • [22] The effect of Hbs on accuracy of pulse oximetry in blood oxygen saturation level measurement among adult patients with sickle cell disease
    Pour, Esmaeil Sadeghi
    Hayati, Saeed
    Khorasani, Mohammad Javad
    Dezfuli, Seyed Ashkan Tabibzadeh
    EURASIAN CHEMICAL COMMUNICATION, 2020, 2 (03): : 296 - 301
  • [23] Measurement of Multi-Wavelength Pulse Oxygen Saturation Based on Dynamic Spectroscopy
    Wang Xiao-fei
    Zhao Wen-jun
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2014, 34 (05) : 1323 - 1326
  • [24] Wearable ear blood oxygen saturation and pulse measurement system based on PPG
    Ma, Gang
    Zhu, Wenliang
    Zhong, Jun
    Tong, Teng
    Zhang, Jie
    Wang, Lirong
    2018 IEEE SMARTWORLD, UBIQUITOUS INTELLIGENCE & COMPUTING, ADVANCED & TRUSTED COMPUTING, SCALABLE COMPUTING & COMMUNICATIONS, CLOUD & BIG DATA COMPUTING, INTERNET OF PEOPLE AND SMART CITY INNOVATION (SMARTWORLD/SCALCOM/UIC/ATC/CBDCOM/IOP/SCI), 2018, : 111 - 116
  • [25] Repeatability and accuracy of fingertip pulse oximeters for measurement of hemoglobin oxygen saturation in arterial blood and pulse rate in anesthetized dogs breathing 100% oxygen
    Ambrisko, Tamas D.
    Dantino, Stephanie C.
    Keating, Stephanie C. J.
    Strahl-Heldreth, Danielle E.
    Sage, Adrianna M.
    Martins, Felipe Da Costa
    Harper, Tisha A. M.
    Wilkins, Pamela A.
    AMERICAN JOURNAL OF VETERINARY RESEARCH, 2021, 82 (04) : 268 - 273
  • [26] Accuracy of a third (Dolphin Voyager) versus first generation pulse oximeter (Nellcor N-180) in predicting arterial oxygen saturation and pulse rate in the anesthetized dog
    Burns, Patrick M.
    Driessen, Bernd
    Boston, Ray
    Gunther, Robert A.
    VETERINARY ANAESTHESIA AND ANALGESIA, 2006, 33 (05) : 281 - 295
  • [27] Defining the role of skin pigmentation in the accuracy of pulse oximetry in COVID-19 subjects receiving domiciliary oxygen supplementation
    Jolobe, Oscar M. P.
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2023, 68 : 187 - 187
  • [28] Accuracy of Disposable Adhesive Pulse Oximeter Finger Probes Used on Ear Lobes and Fingers in Measuring Oxygen Saturations as Compared to Oxygen Saturations Measured on Arterial Blood Gases
    Bagol, W. S.
    Rafanan, A. L.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2023, 207
  • [29] EFFECT OF FLUORESCEIN, INDOCYANINE GREEN, AND METHYLENE-BLUE ON THE MEASUREMENT OF OXYGEN-SATURATION BY PULSE OXIMETRY
    SIDI, A
    RUSH, WR
    PAULUS, DA
    GRAVENSTEIN, N
    DAVIS, RF
    ANESTHESIOLOGY, 1986, 65 (3A) : A132 - A132
  • [30] Accuracy of pulse oximetry in detection of oxygen saturation in patients admitted to the intensive care unit of heart surgery: comparison of finger, toe, forehead and earlobe probes
    Seifi, Sohila
    Khatony, Alireza
    Moradi, Gholamreza
    Abdi, Alireza
    Najafi, Farid
    BMC NURSING, 2018, 17