Case report: Formation and recurrence of inflammatory pseudotumor after metal-on-metal hip arthroplasty

被引:2
|
作者
Min, Meipeng [1 ]
Xing, Chenyang [1 ]
Xu, Peili [1 ]
Wei, Xincheng [1 ]
Fan, Lei [1 ]
机构
[1] Nanjing Med Univ, Dept Orthoped, Affiliated Hosp 2, Nanjing, Peoples R China
关键词
metal-on-metal (MoM) bearings; inflammatory pseudotumor; prosthesis revision; total hip arthroplasty; adverse reaction to metal debris; RISK-FACTORS; IONS; SURVIVORSHIP; PREVALENCE; BEARINGS; IMPLANTS; FAILURE; SYSTEM; MRI;
D O I
10.3389/fmed.2024.1422230
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The metal-on-metal (MoM) artificial hip joint is a prosthesis used in early hip arthroplasty, particularly for hip resurfacing and total hip arthroplasty. However, abrasion and corrosion of MoM bearings result in the production of metal ions, such as cobalt and chromium, thereby inducing several complications such as inflammatory pseudotumor, aseptic inflammation, and allergy to metal ions (delayed type IV hypersensitivity). In this case report, we present a patient who was hospitalized for recurrence of a mass in the right inguinal area. In 2010, the patient underwent right MoM total hip arthroplasty for right femoral head necrosis and exhibited a good postoperative recovery. In 2019, the patient experienced pain in the right hip with activity limitation without any evident triggers, and a palpable mass was observed in the right inguinal area. A large periprosthetic mass was resected under general anesthesia, and the patient recovered well after the operation. Based on post-surgery imaging and pathological examinations, the mass was diagnosed as a periprosthetic inflammatory pseudotumor. In 2021, the inflammatory pseudotumor recurred at the same site. He then underwent right total hip revision surgery under epidural anesthesia and recovered well after surgery. No recurrence was noted at moderate follow-up. The incidence of inflammatory pseudotumors is high in MoM hip arthroplasty. Early revision is necessary in patients who meet the indications for revision, while regular postoperative follow-up is crucial.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Outcomes After Revision of Metal-on-Metal Hip Resurfacing Arthroplasty
    Gross, Thomas P.
    Liu, Fei
    JOURNAL OF ARTHROPLASTY, 2014, 29 (09): : 219 - 223
  • [22] "Asymptomatic" Pseudotumors After Metal-on-Metal Hip Resurfacing Arthroplasty
    Kwon, Young-Min
    Ostlere, Simon J.
    McLardy-Smith, Peter
    Athanasou, Nicholas A.
    Gill, Harinderjit S.
    Murray, David W.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (04): : 511 - 518
  • [23] Tigecycline sclerotherapy for recurrent pseudotumor in aseptic lymphocyte-dominant vasculitis-associated lesion after metal-on-metal total hip arthroplasty:A case report
    Ⅰ-Hao Lin
    Chun-Hao Tsai
    World Journal of Clinical Cases, 2021, (34) : 10696 - 10701
  • [24] Tigecycline sclerotherapy for recurrent pseudotumor in aseptic lymphocyte-dominant vasculitis-associated lesion after metal-on-metal total hip arthroplasty: A case report
    Lin, I-Hao
    Tsai, Chun-Hao
    WORLD JOURNAL OF CLINICAL CASES, 2021, 9 (34) : 10696 - 10701
  • [25] Noninflammatory Pseudotumor Simulating Venous Thrombosis After Metal-on-Metal Hip Resurfacing
    Maurer-Ertl, Werner
    Friesenbichler, Joerg
    Liegl-Atzwanger, Bernadette
    Kuerzl, Gerda
    Windhager, Reinhard
    Leithner, Andreas
    ORTHOPEDICS, 2011, 34 (10) : E678 - E681
  • [26] Hip resurfacing and metal-on-metal total hip arthroplasty
    Beverland, David E.
    Spencer, Robert F.
    HIP INTERNATIONAL, 2011, 21 (03) : 277 - 278
  • [27] Inflammatory pseudotumor associated with femoral nerve palsy following metal-on-metal resurfacing of the hip
    Clayton, Robert A. E.
    Beggs, Ian
    Salter, Donald M.
    Grant, M. Helen
    Patton, James T.
    Porter, Daniel E.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (09): : 1988 - 1993
  • [28] Metal wear in metal-on-metal total hip arthroplasty is not always caused by the bearing: a case report.
    Grimm, Philemon
    Domman-Scherrer, Corina
    Fretz, Valentin
    Wepf, Alexander
    Meier, Christoph
    Wahl, Peter
    SWISS MEDICAL WEEKLY, 2019, 149 : 54S - 54S
  • [29] Comparison of different pseudotumor grading systems in a single cohort of metal-on-metal hip arthroplasty patients
    W. van der Weegen
    K. Brakel
    R. J. Horn
    J. A. Wullems
    H. P. Das
    P. Pilot
    R. G. Nelissen
    Skeletal Radiology, 2014, 43 : 149 - 155
  • [30] Comparison of different pseudotumor grading systems in a single cohort of metal-on-metal hip arthroplasty patients
    van der Weegen, W.
    Brakel, K.
    Horn, R. J.
    Wullems, J. A.
    Das, H. P.
    Pilot, P.
    Nelissen, R. G.
    SKELETAL RADIOLOGY, 2014, 43 (02) : 149 - 155