Dosimetric characteristics of self-expandable metallic and plastic stents for transpapillary biliary decompression in external beam radiotherapy

被引:0
|
作者
Ueda, Yoshihiro [1 ]
Ikezawa, Kenji [2 ]
Sagawa, Tomohiro [1 ]
Isono, Masaru [1 ]
Ohira, Shingo [1 ]
Miyazaki, Masayoshi [1 ]
Takada, Ryoji [2 ]
Yamai, Takuo [2 ]
Ohkawa, Kazuyoshi [2 ]
Teshima, Teruki [3 ]
Konishi, Koji [1 ]
机构
[1] Osaka Int Canc Inst, Dept Radiat Oncol, 3-1-69 Otemae,Chuo Ku, Osaka 5378567, Japan
[2] Osaka Int Canc Inst, Dept Hepatobiliary & Pancreat Oncol, 3-1-69 Otemae,Chuo Ku, Osaka 5378567, Japan
[3] Osaka Heavy Ion Therapy Ctr, Dept Radiat Oncol, 3-1-10 Otemae,Chuo Ku, Osaka 5400008, Japan
关键词
Pancreatobiliary; Radiotherapy; Self-expandable metallic stents; Plastic stents; Radiopaque markers; DOSE PERTURBATION; NEOADJUVANT CHEMORADIATION; INDUCTION CHEMOTHERAPY; CANCER-PATIENTS; PHASE-II; RADIATION; CHEMORADIOTHERAPY; OBSTRUCTION; PLACEMENT; THERAPY;
D O I
10.1007/s13246-024-01447-4
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
There is little evidence regarding radiation dose perturbation caused by the self-expandable metallic stents (SEMSs) used for transpapillary biliary decompression. We aimed to compare SEMSs with plastic stents (PSs) and clarify their dosimetric characteristics. Fifteen SEMSs (10 braided and 5 lasercut type) and six PSs (diameter: 2.3-3.3 mm) were inserted into a water-equivalent solid phantom. In total, 13 SEMSs had radiopaque markers, whereas the other two did not. Using radiochromic films, the dose difference adjacent to the stents at locations proximal, distal, and arc delivery to the radiation source was evaluated based on comparison to measurement of the dose delivery in phantom without any stent in place. The median values of the dose difference for each stent were used to compare the SEMS and PS groups.Results: The dose difference (median (minimum/maximum)) was as follows: proximal, SEMSs + 2.1% (1.8 / 4.7) / PSs + 5.4% (4.1 / 6.3) (p < 0.001); distal, SEMSs -1.0% (-1.6 /-0.4) / PSs -8.9% (-11.7 / -7.4) (p < 0.001); arc delivery, SEMSs 1.2% (0.9 / 2.3) / PSs 2.2% (1.6 / 3.6) (p = 0.005). These results demonstrated that the dose differences of SEMSs were significantly smaller than those of PSs. On the other hand, the dose difference was large at surface of the radiopaque markers for SEMSs: proximal, 10.3% (7.2 / 20.9); distal, -8.4% (-16.3 / -4.2); arc delivery, 5.5% (4.2 / 9.2). SEMSs for biliary decompression can be safely used in patients undergoing radiotherapy, by focusing on the dose distribution around the stents and by paying attention to local changes in the dose distribution of radiopaque markers.
引用
收藏
页码:1323 / 1335
页数:13
相关论文
共 50 条
  • [21] Self-expandable metal versus plastic stents for malignant biliary obstruction
    Hu, Ya-Fei
    Hu, Hai-Jie
    Jin, Yan-Wen
    Li, Fu-Yu
    ASIAN JOURNAL OF SURGERY, 2024, 47 (10) : 4519 - 4521
  • [22] Esophageal self-expandable metallic stents
    Shields, SJ
    GASTROINTESTINAL ENDOSCOPY, 1997, 45 (05) : 439 - 442
  • [23] Clinical effectiveness of covered self-expandable metallic stents for benign biliary strictures
    Kim, H. J.
    Park, S. W.
    Chung, J. B.
    Song, S. Y.
    Bang, S.
    JOURNAL OF GASTROENTEROLOGY AND HEPATOLOGY, 2008, 23 : A142 - A142
  • [24] Self-expandable metallic stents for malignant biliary obstruction with an anomalous pancreaticobiliary junction
    H. Kogure
    T. Tsujino
    H. Yagioka
    T. Sasaki
    Y. Nakai
    K. Hirano
    N. Sasahira
    H. Isayama
    M. Tada
    T. Kawabe
    M. Omata
    Surgical Endoscopy, 2008, 22 : 787 - 791
  • [25] Self-expandable metallic stents for malignant biliary obstruction with an anomalous pancreaticobiliary junction
    Kogure, H.
    Tsujino, T.
    Yagioka, H.
    Sasaki, T.
    Nakai, Y.
    Hirano, K.
    Sasahira, N.
    Isayama, H.
    Tada, M.
    Kawabe, T.
    Omata, M.
    SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2008, 22 (03): : 787 - 791
  • [26] Revision of Bilateral Self-Expandable Metallic Stents for Malignant Hilar Biliary Obstruction
    Son, Jun Hyuk
    Lee, Sang Hyub
    Lee, Hee Seung
    Bang, Seungmin
    Kang, Jinwoo
    Lee, Jae Woo
    Ryu, Ji Kon
    Kim, Yong-Tae
    GASTROINTESTINAL ENDOSCOPY, 2017, 85 (05) : AB637 - AB637
  • [27] New concept of traction force applied to biliary self-expandable metallic stents
    Hori, Yasuki
    Hayashi, Kazuki
    Yoshida, Michihiro
    Naitoh, Itaru
    Nakazawa, Takahiro
    Miyabe, Katsuyuki
    Shimizu, Shuya
    Kondo, Hiromu
    Nishi, Yuji
    Umemura, Shuichiro
    Kato, Akihisa
    Ohara, Hirotaka
    Joh, Takashi
    ENDOSCOPY, 2016, 48 (05) : 472 - 476
  • [28] Covered and Uncovered Self-Expandable Metallic Stents in the Treatment of Malignant Biliary Obstruction
    Qin, Wen-Yan
    Li, Jing-Zhen
    Peng, Wen-Ping
    Zhang, Ming-Ming
    Lao, Bo
    Hong, Jie-Ming
    Li, Lian-Biao
    IRANIAN RED CRESCENT MEDICAL JOURNAL, 2020, 22 (07)
  • [29] Cost-effectiveness trial of self-expandable metal stents and plastic biliary stents in malignant biliary obstruction
    Daroczi Timea
    Bor Renata
    Fabian Anna
    Szabo Ella
    Farkas Klaudia
    Balint Anita
    Czako Laszlo
    Rutka Mariann
    Szucs Monika
    Milassin Agnes
    Molnar Tamas
    Szepes Zolan
    ORVOSI HETILAP, 2016, 157 (07) : 268 - 274