Comparison of erythrocyte sedimentation rate measurement between Westergren method and automated method among patients attending Jigjiga University Sheik Hassen Yabare Referral Hospital, Jigjiga, Ethiopia

被引:0
|
作者
Walle, Muluken [1 ]
Alemayehu, Ermiyas [2 ]
Tesfaye, Addisu [1 ]
Arkew, Mesay [3 ]
Asmerom, Haftu [3 ]
Agidew, Melaku Mekonnen [4 ]
Getu, Fasil [1 ]
机构
[1] Jigjiga Univ, Coll Med & Hlth Sci, Dept Med Lab Sci, Jigjiga, Ethiopia
[2] Wollo Univ, Coll Med & Hlth Sci, Dept Med Lab Sci, Dessie, Ethiopia
[3] Haramaya Univ, Coll Hlth & Med Sci, Sch Med Lab Sci, Harar, Ethiopia
[4] Debre Tabor Univ, Coll Hlth Sci, Dept Med Biochem, Debre Tabor, Ethiopia
关键词
erythrocyte sedimentation rate (ESR); Westergren; automated; SFRI; Jigjiga; Ethiopia; MATIC CUBE 200; ICSH RECOMMENDATIONS; VALIDATION; ESR;
D O I
10.3389/fmed.2024.1414097
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Erythrocyte sedimentation rate (ESR) is a widely used screening test in clinical practice as an indicator of inflammatory and degenerative malignant diseases. The Westergren method, renowned as the gold standard, is valued for its accuracy and cost-effectiveness but demands considerable time and blood volume. Emerging automated methods offer quicker and more convenient alternatives, aiming to replace manual techniques. Nonetheless, validating these automated methods against the reference Westergren method is essential to ensure reliability. Therefore, this study aimed to evaluate ESR measurement results obtained from both the reference Westergren method and the automated (SFRI ESR 3000) method. Methods: A Hospital-based comparative cross-sectional study was conducted at Jigjiga University Sheik Hassen Yabare Referral Hospital from July 15 to September 16, 2023. Following the acquisition of informed consent, blood samples were obtained from 158 participants, five milliliters of blood from each participant. These samples were then subjected to ESR estimation using both the Westergren (reference) method and the automated (SFRI ESR 3000) method. Subsequently, the collected data were analyzed using SPSS version 20 and MedCalc version 12.3.0.0 statistical Softwares. Statistical analyses such as Paired t-tests, Pearson correlation, linear regression, and the Bland and Altman plot were employed. A p-value of < 0.05 was considered statistically significant. Results: The paired sample t-test analysis revealed no significant difference between the use of the reference Westergren method and the automated method for ESR determination, with a mean difference (MD) of 0.7 +/- 9.2 mm/h (P = 0.36). Additionally, a significant correlation was observed between the two methods, with a remarkable correlation coefficient (r = 0.94, p < 0.001). The Bland-Altman data analysis indicated no evidence of systematic bias and demonstrated good agreement of ESR values between the two methods, with a limit of agreement of -17.3 to +18.7. Moreover, within-run imprecision analysis for the automated method across a range of ESR values showed coefficient of variation of 27.08, 12.65, and 10.32% for low, medium, and high ESR levels, respectively. Conclusions: The SFRI ESR 300 automated method demonstrates the potential for interchangeable use with the Westergren method for determining ESR, given the strong correlation and good agreement. Additionally, the same reference range could be applied during interpretation.
引用
收藏
页数:9
相关论文
共 12 条
  • [1] Second Trimester Spontaneous Fundal Rupture of Unscarred Bicornuate Uterus in Primipara: A Case Report and Literature Review; Jigjiga University Sheik Hassen Yabare Comprehensive Specialized Hospital, Jigjiga, Ethiopia
    Areys, Hassen Mohammed
    Omer, Nour Hies
    Osman, Osman Ali
    INTERNATIONAL MEDICAL CASE REPORTS JOURNAL, 2024, 17 : 181 - 185
  • [2] The Effect of Admission Hypothermia for Neonatal Death Among Neonates Admitted to Neonatal Intensive Care Unit at Sheik Hassan Yabare Jigjiga University Referral Hospital in Jigjiga City, Somali Region, Eastern Ethiopia
    Ibrahim, Ahmed Mohammed
    Farah, Abdirahman Mahamed
    Osman, Mohamed Omar
    Hashi, Abdiwahab
    RESEARCH AND REPORTS IN NEONATOLOGY, 2021, 11 : 43 - 55
  • [3] Failed induction of labor and associated factors among women delivered in Jigjiga University Sheik Hassan Yabare Referral Hospital: a cross-sectional study
    Mebratu, Andualem
    Assefa, Dawit Getachew
    Zeleke, Eden Dagnachew
    Mengistu, Nebiyu
    Woldesenbet, Tigist Tekle
    Aynalem, Amdehiwot
    Sefa, Ahmedin
    Mellesse, Dawit
    Tilahun, Ruth
    Seifu, Wubareg
    Bayisa, Yesuneh
    Meshesha, Meiraf Daniel
    Wudneh, Aregahegn
    Ayele, Getnet Melaku
    Demisse, Biniyam
    Kabthymer, Robel Hussen
    Alemu, Asrat
    Abebe Eshetu, Mesfin
    Mahamed, Abbas Ahmed
    Temesgen, Tesfaye
    Funga, Mengistu Lodebo
    Eritero, Aneleay Cherinet
    Aregawi, Saron
    Wodaynew, Tilahun
    Figa, Zerihun
    Molla, Wondwosen
    CURRENT MEDICAL RESEARCH AND OPINION, 2022, 38 (09) : 1655 - 1662
  • [4] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [5] Comparison of the StaRRsed Interliner device with Westergren method in erythrocyte sedimentation rate measurement
    Orkmez, Mustafa
    Orhan, Seren
    Bozdayi, Mehmet Akif
    Tarakcioglu, Mehmet
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (04) : 616 - 622
  • [6] Measurement of Erythrocyte Sedimentation Rate and its Comparison by Microsed System and Westergren Method
    Rehman, Asma
    Farrukh, Ambreen
    Choudary, Munazza
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (02): : 243 - 245
  • [7] Comparison of erythrocyte sedimentation rate measurement by the automated SEDIsystem™ and conventional Westergren method using the Bland and Altman statistical method
    AlFadhli, SM
    Al-Awadhi, AM
    MEDICAL PRINCIPLES AND PRACTICE, 2005, 14 (04) : 241 - 244
  • [8] Automated measurement of the erythrocyte sedimentation rate: method validation and comparison
    Lapic, Ivana
    Piva, Elisa
    Spolaore, Federica
    Tosato, Francesca
    Pelloso, Michela
    Plebani, Mario
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2019, 57 (09) : 1364 - 1373
  • [9] Prevalence, Antimicrobial Resistance, and Associated Factors of Streptococcus pneumoniae Colonization Rate among Old-Age Patients with Respiratory Tract Infection Attending Sheik Hassan Yebere Referral and Karamara General Hospitals, Jigjiga, Ethiopia
    Mekuria, Surafel
    Seyoum, Ayichew
    Ataro, Zerihun
    Abebe, Tigist
    Urgessa, Kedir
    CANADIAN JOURNAL OF INFECTIOUS DISEASES & MEDICAL MICROBIOLOGY, 2022, 2022
  • [10] ERYTHROCYTE SEDIMENTATION RATE MEASUREMENT BY TEST-1 REFLECTS INFLAMMATION BETTER THAN THOSE BY WESTERGREN METHOD IN PATIENTS WITH MALIGNANCY, AUTOIMMUNE DISEASE AND INFECTION
    Cha, C. H.
    Park, C. J.
    Cha, Y. J.
    Kim, H. K.
    Kim, D. H.
    Hong, H.
    Bae, J. H.
    Jung, J. S.
    Chi, H. S.
    Lee, D. S.
    Cho, H. I.
    HAEMATOLOGICA-THE HEMATOLOGY JOURNAL, 2008, 93 : 476 - 476