Automated measurement of the erythrocyte sedimentation rate: method validation and comparison

被引:31
|
作者
Lapic, Ivana [1 ]
Piva, Elisa [2 ]
Spolaore, Federica [2 ]
Tosato, Francesca [2 ]
Pelloso, Michela [2 ]
Plebani, Mario [2 ]
机构
[1] Univ Hosp Ctr Zagreb, Dept Lab Diagnost, Kispaticeva 12, Zagreb 10000, Croatia
[2] Univ Hosp Padova, Dept Lab Med, Padua, Italy
关键词
erythrocyte sedimentation rate; standardization; TEST1; validation; Ves-Matic Cube 200; Westergren method; MATIC CUBE 200; SYSTEM; BLOOD; ESR; DIAGNOSIS; ANALYZER; LENGTH;
D O I
10.1515/cclm-2019-0204
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Development of automated analyzers for erythrocyte sedimentation rate (ESR) has imposed the need for extensive validation prior to their implementation in routine practice, to ensure comparability with the reference Westergren method. The aim of our study was to perform the analytical validation of two automated ESR analyzers, the Ves-Matic Cube 200 and the TEST1. Methods: Validation was performed according to the recent International Council for Standardization in Hematology recommendations and included determination of intrarun and inter-run precision, assessment of sample carryover, hemolysis interference, sensitivity to fibrinogen, method comparison with the gold standard Westergren method and stability test. Results: The highest intrarun imprecision was obtained for the low ESR range (33.5% for Ves-Matic Cube; 37.3% for TEST1) while inter-run coefficients of variation on three levels were much better for the TEST1 (0%, 2% and 1.2%) compared to the Ves-Matic Cube 200 on two levels (24.9% and 5.8%). Both Ves-Matic Cube 200 and TEST1 showed no statistically significant difference when compared with Westergren. Bland-Altman analysis yielded overall insignificant mean biases for all comparisons, but a wider dispersion of results and 95% limits of agreement for comparisons including the Ves-Matic Cube 200. Carryover was considered insignificant, while hemolysis had a negative effect on all assessed ESR methods. The highest sensitivity to fibrinogen was observed for the Ves-Matic Cube 200, followed by Westergren and the least sensitive was the TEST1. Conclusions: The obtained results proved the analytical validity of the TEST1 and the Ves-Matic Cube 200, and high comparability with the gold standard Westergren method, showing obvious improvements in standardization of ESR methods.
引用
收藏
页码:1364 / 1373
页数:10
相关论文
共 50 条
  • [1] Method comparison of automated systems for the erythrocyte sedimentation rate
    Giavarina, D
    Dall'Olio, G
    Soffiati, G
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1999, 112 (05) : 721 - 722
  • [2] Evaluation of an automated erythrocyte sedimentation rate analyzer as compared to the Westergren manual method in measurement of erythrocyte sedimentation rate
    Subramanian, Arulselvi
    Rangarajan, Kanchana
    Pandey, Ravindra Mohan
    Gandhi, Jatin S.
    Sharma, Vijay
    Bhoi, Sanjeev Kumar
    INDIAN JOURNAL OF PATHOLOGY AND MICROBIOLOGY, 2011, 54 (01) : 70 - 74
  • [3] Comparison of erythrocyte sedimentation rate measurement by the automated SEDIsystem™ and conventional Westergren method using the Bland and Altman statistical method
    AlFadhli, SM
    Al-Awadhi, AM
    MEDICAL PRINCIPLES AND PRACTICE, 2005, 14 (04) : 241 - 244
  • [4] Analytical validation of the iSED automated analyzer for erythrocyte sedimentation rate
    Lapic, Ivana
    Milos, Marija
    Tosato, Francesca
    Piva, Elisa
    Zadro, Renata
    Rogic, Dunja
    Plebani, Mario
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (02) : 109 - 115
  • [5] Measurement of Erythrocyte Sedimentation Rate and its Comparison by Microsed System and Westergren Method
    Rehman, Asma
    Farrukh, Ambreen
    Choudary, Munazza
    PAKISTAN JOURNAL OF MEDICAL & HEALTH SCIENCES, 2019, 13 (02): : 243 - 245
  • [6] Comparison of the StaRRsed Interliner device with Westergren method in erythrocyte sedimentation rate measurement
    Orkmez, Mustafa
    Orhan, Seren
    Bozdayi, Mehmet Akif
    Tarakcioglu, Mehmet
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2021, 43 (04) : 616 - 622
  • [7] Method Verification of the Caretium XC-A30 Automated Erythrocyte Sedimentation Rate Analyser for Erythrocyte Sedimentation Rate
    Khongwichit, Soemwit
    Saelim, Menapha
    Na Songkhla, Yanisa
    Buncherd, Hansuk
    Nopparatana, Chawadee
    Srinoun, Kanitta
    MALAYSIAN JOURNAL OF MEDICAL SCIENCES, 2022, 29 (05): : 39 - 47
  • [8] Erythrocyte sedimentation rate measurements using MIX-RATE® X20 and VISION A automated analyzers: Method validation and comparison study
    Prompetchara, Eakachai
    Nowaratsopon, Sunudda
    Wongkamchai, Suppakorn
    Srieakpanit, Jittra
    Ketloy, Chutitorn
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2022, 44 (05) : 846 - 853
  • [9] VALIDATION OF AN ERYTHROCYTE SEDIMENTATION RATE ANALYZER WITH MODIFIED WESTERGREN METHOD
    Patil, Asha
    Nayak, Deepak M.
    Belurkar, Sushma V.
    Verma, Seemitr
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2013, 2 (03): : 284 - 289
  • [10] Comparison of the erythrocyte sedimentation rate measured in the eye casualty department by the Seditainer method with an automated system
    Browning, AC
    Burbidge, AA
    Hodgkins, PR
    EYE, 1999, 13 (6) : 754 - 757