Comparison of two electronic dynamometers for measuring handgrip strength

被引:0
|
作者
Rolsted, Sebastian Keller [1 ,2 ]
Andersen, Kasper Dyrmose [1 ,2 ]
Dandanell, Gustav [1 ,2 ]
Dall, Christian Have [1 ,3 ,4 ]
Zilmer, Camilla Kampp [1 ]
Bulow, Kasper [2 ]
Kristensen, Morten Tange [1 ,4 ]
机构
[1] Copenhagen Univ Hosp, Dept Phys & Occupat Therapy, Copenhagen, Denmark
[2] Univ Coll Absalon, Ctr Hlth & Rehabil, Slagelse, Denmark
[3] Copenhagen Univ Hosp, Dept Cardiol, Copenhagen, Denmark
[4] Univ Copenhagen, Fac Hlth & Med Sci, Dept Clin Med, Copenhagen, Denmark
来源
HAND SURGERY & REHABILITATION | 2024年 / 43卷 / 03期
关键词
Handgrip; Reliability; Validity; Measurements; GRIP STRENGTH; RELIABILITY; AGREEMENT;
D O I
10.1016/j.hansur.2024.101692
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: Handgrip strength (HGS) is a strong predictor and easily applicable assessment, indicating a person's physical condition and health. However, many dynamometers are available; therefore, it is essential to ensure that the results of HGS testing using different dynamometers can be used interchangeably. The primary purpose of this study was to investigate the inter -instrument agreement and criterion validity of the Baseline BIMS Digital Grip Dynamometer in comparison with the Jamar electronic dynamometer (Jamar+). Methods: Seventy participants, aged between 23-88 (five men and five women in each decade from 20 to 80+), performed three attempts with each dynamometer (30 -sec break between attempts) in a randomized order and separated with a 5-minute break between dynamometers. Intraclass correlation coefficient (3.1), standard error of measurement and minimal detectable change were used for comparison of the strongest and average strength measured with dynamometers. Jamar+ and Baseline BIMS Digital Grip Dynamometer were new dynamometers and considered calibrated by the manufacturer. Results: The overall Intraclass correlation coefficient was excellent (0.98). An average (SD) difference of 0.68 (2.2) kg (p = 0.04) was seen for the comparison of the strongest attempt for Baseline BIMS minus Jamar+, Correspondingly, for the average of three attempts, it was 0.37 (2.29, p = 0.2) kg. The standard error of measurement (%) and minimal detectable change (%) of the strongest attempt was 1.64 kg (4.2%) and 3.55 kg (9.0%), respectively. Conclusions: Findings indicate low measurement error with high agreement and criterion validity for the comparison of Baseline BIMS Digital Grip Dynamometer and Jamar+ and that results of the two dynamometers can be used interchangeably. (c) 2024 SFCM. Published by Elsevier Masson SAS. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/).
引用
收藏
页数:7
相关论文
共 50 条
  • [1] COMPARISON OF JAMAR AND BODYGRIP DYNAMOMETERS FOR HANDGRIP STRENGTH MEASUREMENT
    Guerra, Rita S.
    Amaral, Teresa F.
    Sousa, Ana S.
    Fonseca, Isabel
    Pichel, Fernando
    Restivo, Maria T.
    [J]. JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2017, 31 (07) : 1931 - 1940
  • [2] Comparison of handgrip strength values in young children when using two different types of dynamometers
    Abe, Akemi
    Yamasaki, Sakiya
    Tahara, Ryoji
    Loenneke, Jeremy P.
    Abe, Takashi
    [J]. AMERICAN JOURNAL OF HUMAN BIOLOGY, 2022, 34 (09)
  • [3] Comparison of two handgrip dynamometers in older adults before elective surgery
    Andrade, Maricarmen S.
    Honorato, Macarena P.
    Vargas, Javiera P.
    Galvez, Maria de los Angeles
    Rojas, Mariana R.
    [J]. PERIOPERATIVE MEDICINE, 2023, 12 (01)
  • [4] Comparison of two handgrip dynamometers in older adults before elective surgery
    Maricarmen S. Andrade
    Macarena P. Honorato
    Javiera P. Vargas
    María de los Angeles Galvez
    Mariana R. Rojas
    [J]. Perioperative Medicine, 12
  • [5] Validity and reliability of handgrip dynamometry in older adults: A comparison of two widely used dynamometers
    Benton, Melissa J.
    Spicher, Jefferson M.
    Silva-Smith, Amy L.
    [J]. PLOS ONE, 2022, 17 (06):
  • [6] A comparison of two portable dynamometers in the assessment of shoulder and elbow strength
    Vermeulen, HM
    de Bock, GH
    van Houwelingen, HC
    van der Meer, RL
    Mol, MC
    Plus, BT
    Rozing, PM
    Vlieland, TPMV
    [J]. PHYSIOTHERAPY, 2005, 91 (02) : 101 - 112
  • [7] Comparison of portable handheld versus fixed isokinetic dynamometers in measuring strength of the wrist and forearm
    Lucado, Ann
    Fraher, Lisa
    Patel, Hiral
    Munck, Gina
    [J]. PHYSIOTHERAPY THEORY AND PRACTICE, 2019, 35 (07) : 677 - 685
  • [8] Reproducibility of dynamometers in handrail format in evaluating handgrip strength and traction in young and older adults
    Gil, Andre W. O.
    da Silva, Rubens A.
    Pereira, Camila
    Nascimento, Vitor B.
    Amorim, Cesar F.
    Imaizumi, Mayara
    Teixeira, Denilson C.
    [J]. MEDICAL ENGINEERING & PHYSICS, 2022, 100
  • [9] The Bigger the Hand, the Bigger the Difference? Implications for Testing Strength With 2 Popular Handgrip Dynamometers
    Abe, Takashi
    Loenneke, Jeremy P.
    Thiebaud, Robert S.
    Loftin, Mark
    [J]. JOURNAL OF SPORT REHABILITATION, 2019, 28 (03) : 278 - 282
  • [10] ELBOW POSITION AFFECTS HANDGRIP STRENGTH IN ADOLESCENTS: VALIDITY AND RELIABILITY OF JAMAR, DYNEX, AND TKK DYNAMOMETERS
    Espana-Romero, Vanesa
    Ortega, Francisco B.
    Vicente-Rodriguez, German
    Artero, Enrique G.
    Pablo Rey, J.
    Ruiz, Jonatan R.
    [J]. JOURNAL OF STRENGTH AND CONDITIONING RESEARCH, 2010, 24 (01) : 272 - 277