Impact of Titanium Skull Plate on Transcranial Magnetic Stimulation: Analysis of Induced Electric Fields

被引:0
|
作者
Lu, Mai [1 ]
Ueno, Shoogo [2 ]
机构
[1] Lanzhou Jiaotong Univ, Key Lab Optoelect Technol & Intelligent Control, Minist Educ, Lanzhou 730070, Peoples R China
[2] Univ Tokyo, Grad Sch Med, Dept Biomed Engn, Tokyo 1130033, Japan
来源
LIFE-BASEL | 2024年 / 14卷 / 05期
关键词
transcranial magnetic stimulation; titanium skull plate; realistic head model; figure-of-eight coil; impedance method; DIELECTRIC-PROPERTIES; BIOLOGICAL TISSUES; CRANIOPLASTY; SAFETY; MODELS; BRAIN; HZ;
D O I
10.3390/life14050642
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Background: Implanted titanium skull plates (TSPs) in cranioplasty are used to replace or reconstruct areas of the skull that have been damaged or removed due to trauma, surgery, or other medical conditions. However, the presence of a TSP in the head may influence the distribution of the electric field induced during transcranial magnetic stimulation (TMS) procedures. The purpose of this study was to determine how the presence of TSP would interfere with TMS-induced cortical electric fields. Methods: The TMS with a figure-of-eight coil was applied to a realistic head model with TSPs. The distribution of the induced electric field in head tissues was calculated by employing the impedance method, and the results were compared with that of a normal head without TSP. Results: Simulation results show that the distribution of the induced electric field has changed greatly for the head model with TSP. The maximum value of the induced electric field in head tissues was present under one of the circular coil wings rather than in the tissues beneath the junction of the two wings of the Fo8 coil. Conclusions: The induced electric field in deep brain regions was increased for the head model with TSP, which could potentially lead to deep brain stimulation. Since the presence of metallic TSP can greatly influence the distribution of the induced electric field in TMS applications, it is important to adjust the treatment scheme when considering TMS for individuals with cranial titanium plates.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Calculating the electric field in real human head by transcranial magnetic stimulation with shield plate
    Lu, Mai
    Ueno, Shoogo
    JOURNAL OF APPLIED PHYSICS, 2009, 105 (07)
  • [22] The strength and spread of the electric field induced by transcranial rotating permanent magnet stimulation in comparison with conventional transcranial magnetic stimulation
    Helekar, S. A.
    Convento, S.
    Nguyen, L.
    John, B. S.
    Patel, A.
    Yau, J. M.
    Voss, H. U.
    JOURNAL OF NEUROSCIENCE METHODS, 2018, 309 : 153 - 160
  • [23] Comparison of Induced Fields in Virtual Human and Rat Heads by Transcranial Magnetic Stimulation
    Lu, Ya-Wen
    Lu, Mai
    BIOMED RESEARCH INTERNATIONAL, 2018, 2018
  • [24] Head phantoms for electroencephalography and transcranial electric stimulation: a skull material study
    Hunold, Alexander
    Strohmeier, Daniel
    Fiedler, Patrique
    Haueisen, Jens
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2018, 63 (06): : 683 - 689
  • [25] Distortion of the electric field distribution induced in the brain during transcranial magnetic stimulation
    Kim, D. -H.
    Choi, N. -S.
    Won, C.
    Georghiou, G. E.
    IET SCIENCE MEASUREMENT & TECHNOLOGY, 2010, 4 (01) : 12 - 20
  • [26] 3D Modeling of the Induced Electric Field of Transcranial Magnetic Stimulation
    Rafiroiu, D.
    Vlad, S.
    Cret, L.
    Ciupa, R. V.
    INTERNATIONAL CONFERENCE ON ADVANCEMENTS OF MEDICINE AND HEALTH CARE THROUGH TECHNOLOGY, 2009, 26 : 333 - 338
  • [27] Numerical investigation of the electric field distribution induced in the brain by transcranial magnetic stimulation
    Kim, DH
    Loukaides, N
    Sykulski, JK
    Georghiou, GE
    IEE PROCEEDINGS-SCIENCE MEASUREMENT AND TECHNOLOGY, 2004, 151 (06) : 479 - 483
  • [28] How the brain tissue shapes the electric field induced by transcranial magnetic stimulation
    Opitz, Alexander
    Windhoff, Mirko
    Heidemann, Robin M.
    Turner, Robert
    Thielscher, Axel
    NEUROIMAGE, 2011, 58 (03) : 849 - 859
  • [29] Low risk of heating or displacement of titanium skull plates by low frequency repetitive transcranial magnetic stimulation (rTMS)
    Rotenberg, Alexander
    Harrington, Michael
    Birnbaum, Daniel
    Glass, Iain
    Jensen, Frances
    Riviello, James
    Pascual-Leone, Alvaro
    NEUROLOGY, 2007, 68 (12) : A72 - A72
  • [30] In Vitro Magnetic Stimulation of Pig Phrenic Nerve With Transverse and Longitudinal Induced Electric Fields: Analysis of the Stimulation Site
    Lontis, Eugen R.
    Nielsen, Karsten
    Strulik, Johannes J.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2009, 56 (02) : 500 - 512