Pencil-Lead-Based Quasi-Equilibrium Glucose Biosensors

被引:0
|
作者
Shachneva, Svetlana [1 ,2 ]
Lielpetere, Anna [1 ,3 ]
Schuhmann, Wolfgang [1 ]
机构
[1] Ruhr Univ Bochum, Analyt Chem Ctr Electrochem Sci CES, Univ Str 150, D-44780 Bochum, Germany
[2] Max Planck Inst Intelligence Syst, Robot Mat Dept, Heisenbergstr 3, D-70569 Stuttgart, Germany
[3] Latvian Inst Organ Synth, Aizkraukles St 21, LV-1006 Riga, Latvia
来源
ADVANCED SENSOR RESEARCH | 2024年 / 3卷 / 09期
基金
欧盟地平线“2020”;
关键词
FAD-GDH; glucose biosensors; potentiometry; pulse measurement; DESIGN;
D O I
10.1002/adsr.202400024
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Foreign body response is the main reason for the limited lifetime of implantable glucose biosensors. A new measurement strategy exerting minimal disturbance from the equilibrium glucose concentration in the sensor compartment has been proposed to mitigate its adverse effects on the sensor signal. Here, a new measurement strategy using automatically fabricated and robust pencil-lead-based glucose biosensors is implemented. The sensor response depends on critical parameters such as redox-polymer film thickness, film uniformity, rigidity, polymer composition, and the ratio between the enzyme and the polymer. These parameters are controlled by introducing a short-chain redox polymer, a low crosslinker amount, a short-chain electrografting agent and linker, pulse electrografting, and an automated fabrication procedure.
引用
收藏
页数:7
相关论文
共 50 条