Quantifying neutron scintillator screens with X-ray computed tomography

被引:0
|
作者
Chuirazzi, William [1 ]
Cool, Steven [2 ]
Craft, Aaron [1 ]
机构
[1] Idaho Natl Lab, Postirradiat Examinat Dept, Mat & Fuels Complex, Idaho Falls, ID 83415 USA
[2] DMI Reading Imaging, Reading, MA 01867 USA
关键词
X-ray computed tomography; Neutron scintillator screen; Nondestructive examination; Scintillator screen characterization; DISTANCE TRANSFORMS; TECHNOLOGY; CONCRETE;
D O I
10.1016/j.nima.2024.169248
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
As nondestructive imaging techniques become more commonplace, imaging systems must be improved and optimized to meet the growing demand. One key aspect of neutron imaging systems is the scintillator, which determines the time necessary to acquire an image and can also limit the spatial resolution achievable by a system. In this work, X-ray computed tomography was coupled with image processing to measure parameters of a boron-based neutron scintillator screen. The screen's surface and subsurface were examined for defects and a thickness measurement as a function of position was also successfully implemented. Higher resolution scans of a sub-volume of the scintillator coating enabled visualization of the packing of the converter and phosphor powders while also revealing microscopic porosity within the scintillator material. The converter-to-phosphor ratio was quantified with the examined area showing a ZnS:Ag phosphor volume of approximately 60.70% of the entire scintillator volume, while Na10B5O8 converter accounted for approximately 38.76% of the volume.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Quantifying neutron scintillator screens with X-ray computed tomography
    Chuirazzi, William
    Cool, Steven
    Craft, Aaron
    [J]. Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2024, 1063
  • [2] Data fusion in neutron and X-ray computed tomography
    Schrapp, Michael J.
    Goldammer, Matthias
    Schulz, Michael
    Issani, Siraj
    Bhamidipati, Suryanarayana
    Boeni, Peter
    [J]. JOURNAL OF APPLIED PHYSICS, 2014, 116 (16)
  • [3] X-RAY COMPUTED TOMOGRAPHY
    HORN, E
    [J]. ELECTRONICS AND POWER, 1978, 24 (01): : 36 - 41
  • [4] X-ray computed tomography
    Kalender, Willi A.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (13): : R29 - R43
  • [5] X-ray computed tomography
    Withers, Philip J.
    Bouman, Charles
    Carmignato, Simone
    Cnudde, Veerle
    Grimaldi, David
    Hagen, Charlotte K.
    Maire, Eric
    Manley, Marena
    Du Plessis, Anton
    Stock, Stuart R.
    [J]. NATURE REVIEWS METHODS PRIMERS, 2021, 1 (01):
  • [6] X-RAY COMPUTED TOMOGRAPHY
    HORN, E
    [J]. ELECTRONICS AND POWER, 1978, 24 (03): : 181 - 181
  • [7] X-ray computed tomography
    [J]. Nature Reviews Methods Primers, 1
  • [8] Performance analysis of a neutron and X-ray combined computed tomography system
    Sinha, Vaibhav
    Srivastava, Anjali
    Lee, Hyoung Koo
    Liu, Xin
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2014, 750 : 12 - 18
  • [9] Evaluation of X-ray computed tomography for quantifying macroporosity of loamy pasture soils
    Rab, M. A.
    Haling, R. E.
    Aarons, S. R.
    Hannah, M.
    Young, I. M.
    Gibson, D.
    [J]. GEODERMA, 2014, 213 : 460 - 470
  • [10] Neutron tomography of internal structures of vertebrate remains: A comparison with X-ray computed tomography
    Schwarz, D
    Vontobel, P
    Lehmann, EH
    Meyer, CA
    Bongartz, G
    [J]. PALAEONTOLOGIA ELECTRONICA, 2005, 8 (02)