Determination of a detectability index for image quality evaluation using a PVA-microcapsules phantom

被引:1
|
作者
Songsaeng, Chatsuda [1 ,2 ]
Salomon, Elisabeth [1 ]
Ortner, Markus [1 ]
Zalka, Lukas [1 ,2 ]
Koehrer, Johannes [1 ,2 ]
Homolka, Peter [1 ]
Pospischek, Julia [1 ]
Kollmann, Christian [1 ]
Miloslavsky, Jack [1 ]
Palli, Uli [3 ]
Figl, Michael [1 ,2 ]
Hummel, Johann [1 ,2 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Spitalgasse 23, A-1090 Vienna, Austria
[2] Christian Doppler Lab Math Modelling & Simulat Ne, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[3] Accella AI, Sunnyvale, CA 94087 USA
关键词
Ultrasound phantom; detectability index; image quality;
D O I
10.1117/12.3006046
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The goal of this study was the development of a phantom for the determination of the image quality of ultrasound (US) based on the Linear System Theory. Modular transfer Function (MTF) and noise power spectrum (NPS) were determined on two US phantoms. One contained a cylinder filled with water, which appears as a circle in the US images, the second was completely homogeneous. The base material of the phantom was Poly(vinyl alcohol) which was mixed with water in a 1:9 ratio. Additionally, micro-plastic spheres and starch, respectively, were included to increase echogenicity. An algorithm was developed that calculates a radial MTF from the circular structure representing spatial resolution averaged across all directions. Noise power spectrum was determined as described by Fredenberg et al., image quality was evaluated by means of a detectability index for different diameters. Two transducers with different bandwidths (4-13 MHz and 3-8 MHz) were used to show the dependence of the index on the main frequency of the US wave. In addition, three penetration depths, which also require different frequencies, were used. Detectability was higher with the transducer of higher frequency for all measurements i.e. for all depths and all diameters (paired t-tests, all p < 0.01). There was also a decrease of detectability with increasing depth for both transducers. The dependence of the index on the axial distance of the ROIs was highly significant (two-sided, paired Wilcoxon test, p < 0.00001). With respect to the comparison for the different phantom materials (PVA with starch and PVA with micro-spheres), the null hypothesis (equality of variances; unpaired, two-sided Wilcoxon test) could not be rejected (p = 0.14). The results suggest that the concept of the detectability index can also be applied to US images with some reservations.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Image quality and dose optimisation for infant CT using a paediatric phantom
    Jack W. Lambert
    Andrew S. Phelps
    Jesse L. Courtier
    Robert G. Gould
    John D. MacKenzie
    European Radiology, 2016, 26 : 1387 - 1395
  • [42] Evaluation of a new system for chest tomosynthesis: aspects of image quality of different protocols determined using an anthropomorphic phantom
    Jadidi, M.
    Sundin, A.
    Aspelin, P.
    Bath, M.
    Nyren, S.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1053):
  • [43] Impact of Patient Alignment on Image Quality in C-Arm Computed Tomography - Evaluation Using an ACR Phantom
    Alikhani, Babak
    Renne, Julius
    Maschke, Sabine
    Hinrichs, Jan B.
    Wacker, Frank K.
    Werncke, Thomas
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2021, 193 (04): : 417 - 426
  • [44] WAVE-FORM INDEX EVALUATION USING AN ELECTRONIC INJECTION PHANTOM
    POTTER, DJ
    IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1995, 14 (01): : 91 - &
  • [45] Impact of deep learning reconstructions on image quality and liver lesion detectability in dual-energy CT: An anthropomorphic phantom study
    Pauthe, Aurelie
    Milliner, Milan
    Pasquier, Hugo
    Campagnolo, Lucie
    Mule, Sebastien
    Luciani, Alain
    MEDICAL PHYSICS, 2025,
  • [46] IMAGE QUALITY EVALUATION USING IMAGE QUALITY RULER AND GRAPHICAL MODEL
    Wang, Weibao
    Allebach, Jan
    Guo, Yandong
    2015 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP), 2015, : 2256 - 2259
  • [47] DETECTABILITY INDEX, LIKELIHOOD RATIO AND DIAGNOSTIC OUTPUT AS PSYCHOPHYSICAL MEASURES FOR EVALUATION OF MEDICAL IMAGE-PROCESSING
    DEVALK, JPJ
    EIJKMAN, EGJ
    VINTCENT, J
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1981, 19 (05) : 597 - 604
  • [48] Image quality evaluation using generalized natural image
    Kagitani, Kenji
    IMAGE QUALITY AND SYSTEM PERFORMANCE V, 2008, 6808
  • [49] An Evaluation Index Based on Parameter Weight for Image Inpainting Quality
    Wang, Song
    Li, Hong
    Zhu, Xia
    Li, Ping
    PROCEEDINGS OF THE 9TH INTERNATIONAL CONFERENCE FOR YOUNG COMPUTER SCIENTISTS, VOLS 1-5, 2008, : 786 - 790
  • [50] A new adaptive filter and quality evaluation index for image restoration
    Jiménez, AR
    Ceres, R
    Pons, JL
    PATTERN RECOGNITION, 2001, 34 (02) : 457 - 467