Determination of a detectability index for image quality evaluation using a PVA-microcapsules phantom

被引:1
|
作者
Songsaeng, Chatsuda [1 ,2 ]
Salomon, Elisabeth [1 ]
Ortner, Markus [1 ]
Zalka, Lukas [1 ,2 ]
Koehrer, Johannes [1 ,2 ]
Homolka, Peter [1 ]
Pospischek, Julia [1 ]
Kollmann, Christian [1 ]
Miloslavsky, Jack [1 ]
Palli, Uli [3 ]
Figl, Michael [1 ,2 ]
Hummel, Johann [1 ,2 ]
机构
[1] Med Univ Vienna, Ctr Med Phys & Biomed Engn, Spitalgasse 23, A-1090 Vienna, Austria
[2] Christian Doppler Lab Math Modelling & Simulat Ne, Waehringer Guertel 18-20, A-1090 Vienna, Austria
[3] Accella AI, Sunnyvale, CA 94087 USA
关键词
Ultrasound phantom; detectability index; image quality;
D O I
10.1117/12.3006046
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The goal of this study was the development of a phantom for the determination of the image quality of ultrasound (US) based on the Linear System Theory. Modular transfer Function (MTF) and noise power spectrum (NPS) were determined on two US phantoms. One contained a cylinder filled with water, which appears as a circle in the US images, the second was completely homogeneous. The base material of the phantom was Poly(vinyl alcohol) which was mixed with water in a 1:9 ratio. Additionally, micro-plastic spheres and starch, respectively, were included to increase echogenicity. An algorithm was developed that calculates a radial MTF from the circular structure representing spatial resolution averaged across all directions. Noise power spectrum was determined as described by Fredenberg et al., image quality was evaluated by means of a detectability index for different diameters. Two transducers with different bandwidths (4-13 MHz and 3-8 MHz) were used to show the dependence of the index on the main frequency of the US wave. In addition, three penetration depths, which also require different frequencies, were used. Detectability was higher with the transducer of higher frequency for all measurements i.e. for all depths and all diameters (paired t-tests, all p < 0.01). There was also a decrease of detectability with increasing depth for both transducers. The dependence of the index on the axial distance of the ROIs was highly significant (two-sided, paired Wilcoxon test, p < 0.00001). With respect to the comparison for the different phantom materials (PVA with starch and PVA with micro-spheres), the null hypothesis (equality of variances; unpaired, two-sided Wilcoxon test) could not be rejected (p = 0.14). The results suggest that the concept of the detectability index can also be applied to US images with some reservations.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] EVALUATION OF PORTAL IMAGE QUALITY USING DEDICATED PHANTOM
    Blasiak, B.
    Litoborski, M.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S550 - S550
  • [2] Lesion Detectability of a PEM System with a Proposed Breast Image Quality Phantom
    Luo, W.
    Matthews, C. G.
    Lu, X.
    Landor, H. P.
    Anashkin, E.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [3] Computerized evaluation of mammographic image quality using phantom images
    Dougherty, G
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 1998, 22 (05) : 365 - 373
  • [4] A phantom using titanium and Landolt rings for image quality evaluation in mammography
    Gala, Hugo de las Heras
    Schoefer, Felix
    Tiller, Britta
    Chevalier, Margarita
    Zwettler, Georg
    Semturs, Friedrich
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (08): : L17 - L30
  • [5] Automatic evaluation of the image quality of a mammographic phantom
    Mayo, P
    Rodenas, F
    Verdú, G
    Villaescusa, JI
    Campayo, JM
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2004, 73 (02) : 115 - 128
  • [6] Evaluation of the detectability of breast cancer lesions using a modified anthropomorphic phantom
    Doshi, NK
    Basic, M
    Cherry, SR
    JOURNAL OF NUCLEAR MEDICINE, 1998, 39 (11) : 1951 - 1957
  • [7] A multi-scanner evaluation of PET image quality using phantom studies
    Surti, S
    Badawi, RD
    Holdsworth, CH
    El Fakhri, G
    Kinahan, PE
    Karp, JS
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2425 - 2427
  • [8] Evaluation of Image Quality in Digital Radiography Based On Marginal Detectability
    Scott, A.
    Zhou, Y.
    Zhang, D.
    Binesh, N.
    Lee, C.
    Bosteder, W.
    MEDICAL PHYSICS, 2020, 47 (06) : E534 - E534
  • [9] A quantitative method for visual phantom image quality evaluation
    Chakraborty, D
    Liu, X
    O'Shea, M
    Toto, L
    MEDICAL IMAGING 2000: IMAGE PERCEPTION AND PERFORMANCE, 2000, 3981 : 24 - 33
  • [10] Image quality and detectability in Siemens Biograph PET/MRI and PET/CT systems—a phantom study
    Silje Kjærnes Øen
    Lars Birger Aasheim
    Live Eikenes
    Anna Maria Karlberg
    EJNMMI Physics, 6