A Study on the Efficacy and Safety of Pipeline Shield Stents and Nonmodified Surface Flow-Diverting Stents in Patients with Intracranial Aneurysms

被引:0
|
作者
Garcia-Nunez, Julio [1 ,2 ]
Vera, Daniela D. [1 ]
Ortiz, Andres F. [1 ]
Rodriguez, Adriana [1 ]
Lara, Juan Jose [3 ]
Gomez, Maria Jose [3 ]
Serrano-Gomez, Sergio [4 ]
Reyes, Adriana [4 ]
Ferreira-Prada, Carlos [1 ]
Galvis-Mendez, Melquizidel [1 ]
Vargas-Perez, Oliverio [1 ,2 ]
Mantilla-Garcia, Daniel E. [1 ,2 ]
机构
[1] Fdn Oftalmol Santander, Intervent Radiol Dept, Clin Ardila Lulle, Floridablanca, Colombia
[2] Univ Autonoma Bucaramanga, Intervent Radiol Dept, Bucaramanga, Colombia
[3] Univ Autonoma Bucaramanga, Radiol Dept, Bucaramanga, Colombia
[4] Univ Autonoma Bucaramanga, Res Grp UNAB, Bucaramanga, Colombia
关键词
Embolization; Endovascular procedures; Occlusion rates; Thromboembolic complications;
D O I
10.1016/j.wneu.2024.01.028
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND: Few studies have compared the Pipeline Shield stents with previous generations of flowdiverting stents (FDSs) for the treatment of unruptured intracranial aneurysms. This study aimed to evaluate the efficacy and safety of Pipeline Shield stents and FDSs without modified surfaces. METHODS: The present evaluation is a retrospective cohort study of patients endovascularly treated with Pipeline Shield stents or FDSs without modified surfaces for unruptured intracranial aneurysms between January 2014 and June 2022. The data analyzed were obtained from the anonymized database of our institution 's interventional radiology service. RESULTS: A total of 147 patients with 155 unruptured intracranial aneurysms were included. Of the 155 aneurysms, 96 were treated with Pipeline Shield stents and 59 with FDSs without modified surfaces. The aneurysms treated with Pipeline Shield stents had higher 6 -month (O 'Kelly-Marotta [OKM] D; 87.5% vs. 71.4%; P = 0.025) and 1 -year (OKM D; 82.5% vs. 63.0%; P = 0.047) occlusion rates than the aneurysms treated using FDSs without modified surfaces. No differences between the devices were found at the 1 -year follow-up in the incidence of ischemic stroke ( P = 0.939) or hemorrhagic complications ( P = 0.559). CONCLUSIONS: Pipeline Shield stents demonstrated superior complete occlusion rates (OKM D) at both the 6month and the 1 -year follow-up assessments compared with nonmodified surface FDSs. No significant differences were found in the safety profiles between the 2 types of stents with regard to thromboembolic complications and ischemic events. Further research with larger study popu lations is necessary to validate these findings.
引用
收藏
页码:E781 / E786
页数:6
相关论文
共 50 条
  • [41] INCIDENCE OF TRANSCRANIAL DOPPLER DETECTED MICROEMBOLI AND STROKE IN PATIENTS TREATED WITH FLOW DIVERTING STENTS FOR INTRACRANIAL ANEURYSMS
    Jaleel, N.
    Goren, O.
    Hayward, D.
    Sampath, R.
    Kucera, M.
    Grobe, K.
    Douville, C.
    Loh, Y.
    Patel, A.
    McDougall, C.
    Monteith, S.
    INTERNATIONAL JOURNAL OF STROKE, 2017, 12 : 24 - 24
  • [42] 3T MRI in the Evaluation of Brain Aneurysms Treated with Flow-Diverting Stents: Preliminary Experience
    Toni, F.
    Marliani, A. F.
    Cirillo, L.
    Battaglia, S.
    Princiotta, C.
    Dall'Olio, M.
    Simonetti, L.
    Leonardi, M.
    NEURORADIOLOGY JOURNAL, 2009, 22 (05): : 588 - 599
  • [43] Regarding "Multiple overlapping uncovered stents as an alternative flow-diverting strategy in the management of peripheral and visceral aneurysms"
    Mattiuz, Chiara
    Secchi, Francesco
    Mauri, Giovanni
    JOURNAL OF VASCULAR SURGERY, 2015, 62 (01) : 269 - 269
  • [44] Multiphysics coupling numerical simulation of flow-diverting stents in the treatment of patients with pulsatile tinnitus
    Mu, Zhenxia
    Liu, Li
    Sun, Yufeng
    Gao, Bin
    Lv, Han
    Zhao, Pengfei
    Liu, Youjun
    Wang, Zhenchang
    INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2021, 37 (12)
  • [45] Intracranial Aneurysms Treated by Flow-Diverting Stents: Long-Term Follow-Up with Contrast-Enhanced Magnetic Resonance Angiography
    Maximilian Patzig
    Robert Forbrig
    Lorenz Ertl
    Hartmut Brückmann
    Gunther Fesl
    CardioVascular and Interventional Radiology, 2017, 40 : 1713 - 1722
  • [46] Safety and Efficacy of Endovascular Coils and Non-Flow-Diverting Stents for Management of Unruptured Intracranial Aneurysms: A Location-Specific Outcomes Analysis
    Vera, Daniela
    Cardenas, Sauder A.
    Ortiz, Andres F.
    Rodriguez, Adriana M.
    Ferreira, Carlos A.
    Serrano, Sergio
    Reyes, Adriana
    Galvis, Melquizidel
    Vargas, Oliverio
    Mantilla, Daniel E.
    WORLD NEUROSURGERY, 2024, 182 : E734 - E741
  • [47] Intracranial Aneurysms Treated by Flow-Diverting Stents: Long-Term Follow-Up with Contrast-Enhanced Magnetic Resonance Angiography
    Patzig, Maximilian
    Forbrig, Robert
    Ertl, Lorenz
    Brueckmann, Hartmut
    Fesl, Gunther
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2017, 40 (11) : 1713 - 1722
  • [48] A systematic review and meta-analysis of the performance of flow-diverting stents in the treatment of peripheral and visceral artery aneurysms
    Zhang, Yongxue
    Xiang, Dekun
    Lu, Qingsheng
    Wu, Mengtao
    Cui, Jinguo
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2021, 97 (03) : 461 - 469
  • [49] A Novel Grading Scale for the Angiographic Assessment of Intracranial Aneurysms Treated Using Flow Diverting Stents
    O'Kelly, C. J.
    Krings, T.
    Fiorella, D.
    Marotta, T. R.
    INTERVENTIONAL NEURORADIOLOGY, 2010, 16 (02): : 133 - 137
  • [50] Safety and efficacy of braided stents as stent monotherapy for the treatment of small intracranial aneurysms
    Capirossi, Carolina
    Radu, Razvan A.
    Gascou, Gregory
    Lefevre, Pierre-Henri
    Dargazanli, Cyril
    Machi, Paolo
    Costalat, Vincent
    Cagnazzo, Federico
    INTERVENTIONAL NEURORADIOLOGY, 2023,